AUTOKIT TOTAL KETONE BODIES KIT (SERUM) AND WAKO KETONE BODY CALIBRATOR
K033860 · Wako Chemicals USA, Inc. · JIT · Apr 23, 2004 · Clinical Chemistry
Device Facts
Record ID
K033860
Device Name
AUTOKIT TOTAL KETONE BODIES KIT (SERUM) AND WAKO KETONE BODY CALIBRATOR
Applicant
Wako Chemicals USA, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Apr 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The Wako Total Ketone Body Calibrators are intended to be used with the Wako Total Ketone Bodies and the Wako Autokit 3-HB test kits to establish points of reference that are used in the determination of values in the measurement of total ketone bodies and 3-HB in human serum or plasma. The calibrators are for Rx in vitro diagnostic use.
Device Story
Wako Total Ketone Body Calibrators; liquid, ready-to-use calibrators; two levels (40 μmol/L and 300 μmol/L 3-Hydroxybutyrate). Used in clinical laboratory settings with Wako Total Ketone Bodies and Wako Autokit 3-HB test kits on Wako 30R® instruments. Provides reference points for assay calibration; enables accurate quantification of ketone bodies in patient serum or plasma. Value assignment derived via molar extinction coefficient (K-factor) of NADH.
Clinical Evidence
No clinical data. Bench testing only. Real-time stability studies demonstrated that both calibrator levels remained within 1% of expected values over 15 months when stored at 10°C.
Technological Characteristics
Liquid, ready-to-use calibrators. Matrix: Goods Buffer, polyethylene glycol, sodium azide, and 3-Hydroxybutyrate. Two levels: 40 μmol/L and 300 μmol/L. Traceability established via molar extinction coefficient (K-factor) of NADH.
Indications for Use
Indicated for use as calibrators for the measurement of total ketone bodies and 3-hydroxybutyrate (3-HB) in human serum or plasma samples.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K042406 — OLYMPUS LACTATE, OSR6193 · Olympus America, Inc. · Dec 3, 2004
K030621 — ATAC CALIBRATOR · Elan Diagnostics · Mar 19, 2003
K042550 — MULTI-LIPID CALIBRATOR · Wako Chemicals USA, Inc. · Nov 8, 2004
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number:
k033860
B. Purpose of the Submission:
New device
C. Analyte:
Hydroxybutyric Acid
D. Type of Test:
Calibrator
E. Applicant:
Wako Chemicals, USA, Inc.
F. Proprietary and Established Names:
Wako Total Ketone Body Calibrators
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1150, Calibrators, Secondary
2. Classification:
II
3. Product Code:
JIT
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
Refer to Indications for use.
2. Indication(s) for use:
The Wako Total Ketone Body Calibrators are intended to be used with the Wako Total Ketone Bodies and the Wako Autokit 3-HB test kits to establish points of reference that are used in the determination of values in the measurement of total ketone bodies and 3-HB in human serum or plasma.
{1}
Page 2 of 3
(Reviewer Note: 3-HB is a common abbreviation for 3-Hydroxybutyrate.)
3. **Special condition for use statement(s):**
The calibrators are for Rx in vitro diagnostic use.
4. **Special instrument requirements:**
Wako 30R®
I. **Device Description:**
The products consist of two levels of liquid ready to use calibrators. They consist of a matrix of Goods Buffer, polyethylene glycol and sodium azide and are prepared by adding 3-Hydroxybutyrate.
J. **Substantial Equivalence Information:**
1. **Predicate device name(s):**
Stanbio Standard, 1 mM Sodium D-3-Hydroxybutyrate (Formerly GDS)
2. **Predicate K number(s):**
k910108
3. **Comparison with predicate:**
Both are liquid based calibrators intended to provide reference points for ketone assays. They are intended to calibrate different assays made by different manufacturers.
K. **Standard/Guidance Document Referenced (if applicable):**
The sponsor did not reference any standards in their submission.
L. **Test Principle:** Not applicable
M. **Performance Characteristics (if/when applicable):**
1. **Analytical performance:**
a. **Precision/Reproducibility:**
Not applicable.
b. **Linearity/assay reportable range:**
Not applicable.
c. **Traceability (controls, calibrators, or method):**
There are two levels of calibrators:
40 μmol/L and 300 μmol/L of 3-Hydroxybutyrate.
Calibrators are prepared by adding 3-Hydroxybutyrate to buffer.
The grade of chemicals utilized was not specified.
The value assignment and traceability is provided through use of the molar extinction coefficient (K-factor) of NADH.
{2}
Page 3 of 3
Results from Real Time Stability studies are summarized. The calibrators were evaluated for up to 15 months when stored at 10 degrees C and were both levels were within 1% of the expected value at all measurement times.
d. Detection limit: Not applicable.
e. Analytical specificity: Not applicable.
f. Assay cut-off: Not applicable.
2. Comparison studies:
a. Method comparison with predicate device: Not applicable.
b. Matrix comparison: Not applicable.
3. Clinical studies:
a. Clinical sensitivity: Not applicable. Clinical studies are not typically submitted for this device type.
b. Clinical specificity: Not applicable. Clinical studies are not typically submitted for this device type.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off: Not applicable.
5. Expected values/Reference range: Not applicable.
N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.