← Product Code [LLZ](/productcode/LLZ) · K033825

# ORTHOPACS (K033825)

_Etrauma.Com Corp. · LLZ · Feb 2, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K033825

## Device Facts

- **Applicant:** Etrauma.Com Corp.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Feb 2, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

OrthoPACS is intended for the manipulation and displaying of medical images by a suitable licensed and qualified healthcare professional. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. OrthoPACS also uses such images, in conjunction with templates for prosthetic devices, for the purposes of choosing the nature and characteristics of the prosthetic device to be used when planning a potential surgical procedure.

## Device Story

OrthoPACS is a digital info-imaging system for orthopedic applications in offices, private practices, surgery, and clinics. It ingests medical images from various modalities via DICOM or similar standards; provides tools for image viewing, processing, and measurement; and overlays templates for prosthetic device selection during surgical planning. Operated by physicians, nurses, technicians, and office managers, the system facilitates rapid retrieval of patient historicals and clinical images. Output is displayed on viewing stations to assist clinicians in surgical decision-making and prosthetic sizing. The system supports transmission to remote viewing stations over medical imaging networks.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Standalone software package for digital image processing and measurement. Complies with ISO 9001:1994 and ISO 13485:1996. Supports DICOM and similar interface standards for image acquisition, transmission, storage, and printing. Architecture includes acquisition/transmission, viewing station, and storage modules with remote accessibility.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Sectra Orthopedic Package ([K031590](/device/K031590.md))
- Cedara I-SoftView ([K022881](/device/K022881.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K033825

OrthoPACS Device

eTrauma.com Corporation 510(k) Premarket Notification

FEB - 2 2004

## 510(K) SUMMARY OrthoPACS December 10, 2003

#### 1. SUBMITTER:

eTrauma.com Corporation 1425 East Newport Center Drive Deerfield Beach, FL 33442 Telephone: 877-387-2862 Contact: Dan Hodgeman, President Date Prepared: December 10, 2003

#### 2. DEVICE:

| Trade name:            | OrthoPACS                                   |
|------------------------|---------------------------------------------|
| Classification Name:   | Picture Archiving and Communications System |
| Classification Number: | 892.2050                                    |
| Product Code           | LLZ                                         |
| Common Name:           | Medical Image Management System             |
| Classification:        | Class II                                    |
| Classification Panel:  | Radiology                                   |

#### 3. PREDICATE DEVICE:

The predicate devices used to determine substantial equivalence for the eTrauma OrthoPACS System were the Sectra Orthopedic Package (K031590) and the Cedara I-SoftView (K022881).

#### 4. DEVICE DESCRIPTION:

eTrauma's OrthoPACS is a Digital Info-Imaging System for orthopaedic specific applications in office, private practice, surgery and clinics. The system allows orthopaedic surgcons to navigate quickly through a typical office day by using an efficient and fast system for viewing clinical images and retrieval of information regarding patient historicals whether tocal or residing on other digital information systems such as electronic medical record and/or practice management software. The four general classes of users for OrthoPACS include x-ray technicians, office managers, physicians and nurses.

{1}------------------------------------------------

OrthoPACS Device

K033825
Page 2 of 2

OrthoPACS consists of the following components:

- Acquisition and Transmission Modules ●
- Viewing Station Module .
- Storage Module .
- RemoteImage Accessibility .

## 5. INDICATIONS FOR USE:

OrthoPACS is intended for the manipulation and displaying of medical images by a suitable licensed and qualified healthcare professional. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. OrthoPACS also uses such images, in conjunction with templates for prosthetic devices, for the purposes of choosing the nature and characteristics of the prosthetic device to be used when planning a potential surgical procedure.

# 6. COMPARISON OF CHARACTERISTICS:

All of the systems have the same intended use and are stand alone software packages that allow digital image processing and measurement capability. All systems transmit to remote viewing stations over a medical imaging network. None of the systems contact or require interaction with, the patient. All are used by health care professionals who interpret images and information being displayed or printed.

#### 7. PERFORMANCE DATA:

The subject device is developed according to ISO 9001:1994 and complies with ISO 13485:1996.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 2 2004

eTrauma.com Corporation % Ms. Debbie Iampictro President QRC Consulting 7 Tiffany Trail HOPKINTON MA 01748 Re: K033825 Trade/Device Name: OrthoPACS

Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communication system Regulatory Class: II Product Code: 90 LLZ Dated: December 10, 2003 Received: December 10, 2003

Dear Ms. Iampietro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good nanufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrfv/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brydon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

eTrauma.com Corporation 510(k) Premarket Notification OrthoPACS Device

Page 1 of 1

KO33825 510(k) Number (if known):

Device Name: OrthoPACS

Indications for Use:

OrthoPACS is intended for the manipulation and displaying of medical images by a suitable licensed and qualified healthcare professional. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. OrthoPACS also uses such images, in conjunction with templates for prosthetic devices, for the purposes of choosing the nature and characteristics of the prosthetic device to be used when planning a potential surgical procedure.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Posted July 1, 1998)

Qurich. derme

(Division Sign-Off) Division of Reproductive, Abdominal, Lad Radiological Devices 510(k) Number .

(Optional Format 3-10-98)

Prescription Use
(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/device/K033825](https://fda.innolitics.com/device/K033825)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
