EPSILON METASUL 32MM ACETABULAR INSERTS, STANDARD AND HOODED
K033634 · Centerpulse Orthopedics, Inc. · KWA · Sep 15, 2004 · Orthopedic
Device Facts
Record ID
K033634
Device Name
EPSILON METASUL 32MM ACETABULAR INSERTS, STANDARD AND HOODED
Applicant
Centerpulse Orthopedics, Inc.
Product Code
KWA · Orthopedic
Decision Date
Sep 15, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The 32mm Epsilon Metasul Acetabular Insert and Metasul Femoral Head are intended for use in total hip arthroplasty for treatment of the following: - patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., . avascular necrosis, osteoarthritis, and inflammatory joint disease (JJD), e.g., rheumatoid arthritis: - those patients with failed previous surgery where pain, deformity, or dysfunction . persists: - . revision of previously failed hip arthroplasty.
Device Story
Total hip replacement system consisting of 32mm acetabular insert and modular femoral head. Acetabular insert features 'polysandwich' design: UHMWPE outer shell thermo-mechanically bonded to mirror-finished CoCr alloy metallic inlay. Femoral head manufactured from wrought forged CoCrMo alloy with 12/14 Morse-type female taper. System functions as metal-on-metal bearing; natural synovial fluid provides lubrication. Hooded insert variants include 20° polyethylene overhang to resist subluxation. Used in orthopedic surgery; implanted by surgeons. Provides stable articulation for hip joint; intended to restore function and reduce pain in patients with degenerative or inflammatory joint disease.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: UHMWPE (ASTM F648) and wrought hot-forged CoCr alloy (ISO 5832-12). Design: Hemispherical acetabular insert with metallic inlay; modular femoral head with 12/14 Morse-type taper. Bearing: Metal-on-metal. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for total hip arthroplasty in patients with noninflammatory degenerative joint disease (e.g., osteoarthritis, post-traumatic arthritis, avascular necrosis), inflammatory joint disease (e.g., rheumatoid arthritis), failed previous surgery with persistent pain/deformity/dysfunction, or revision of failed hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
Predicate Devices
Centerpulse Orthopedics Epsilon Metasul Acetabular Insert, 28mm Standard (K974728)
{1}------------------------------------------------
K033634
510(k) Premarket Notification
32mm Epsilon Metasul Acetabular Insert and 32mm Metasul Femoral Head, as amended
## 510(k) SUMMARY
| SPONSOR NAME: | Centerpulse Orthopedics, Inc., a division of Zimmer<br>9900 Spectrum Drive<br>Austin, TX 78717 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Audrey Swearingen<br>Phone: (512) 432-9255<br>E-Mail: Audrey.Swearingen@Zimmer.com |
| TRADE NAME: | Epsilon™ Metasul® Acetabular Insert and Metasul Modular<br>Femoral Head |
| COMMON NAME: | Total hip replacement system acetabular insert and head |
| CLASSIFICATION: | CFR §888.3330 (KWA) - Hip joint metal/metal<br>semiconstrained, with an uncemented acetabular<br>component, prosthesis, reviewed by the Orthopedic Devices<br>panel. Metal-on-metal hip prostheses are Preamendment<br>Class III devices. |
#### PREDICATE DEVICES:
- Centerpulse Orthopedics Epsilon Metasul Acetabular Insert, 28mm Standard � (K974728) and Hooded (K001526)
- Centerpulse Metasul® Modular Femoral Head (K974728) .
- Biomet M2a™ Ringloc® Acetabular Liner (K002379) .
- Biomet M2a -Taper" Acetabular System (K003363, K993438, unknown) ●
- J&J DePuy Ultamet Femoral Heads (K980513) .
## DEVICE DESCRIPTION:
The Epsilon Metasul 32mm Acetabular Insert is a hemispherically shaped design, composed of an outer component manufactured from polyethylene (UHMWPE) (in compliance with ASTM F648) which is thermo-mechanically bonded to a wrought hotforged CoCr alloy metallic inlay (in compliance with ISO 5832-12). The Epsilon Metasul Acetabular Insert is designed for use only with a Metasul femoral head component, as a metal-on-metal system. The body's natural synovial fluid lubricates the metal surfaces. The Epsilon Metasul 32mm Acetabular Insert, both standard and hooded, is available in sizes designed to mate with Converge® Acetabular Shells, sizes 53mm to 81mm (in 2mm increments).
The Epsilon Metasul 32mm Insert has what is commonly referred to as a "polysandwich" design. The inner diameter, which forms the bearing surface of the insert, features a metallic Metasul inlay that is polished to a mirror-finish and thermomechanically bonded into the polyethylene liner, which is then locked into the Converge acetabular shell via the proven snap mechanism. On the hooded inserts, the face of the polyethylene outer diameter incorporates a 20° overhang of polyethylene extending superiorly from the midpoint of the insert face. This hood feature is designed to provide additional resistance to subluxation and instability.
The 32mm Metasul® Modular Femoral Head is manufactured from Protasul-21WF
{2}------------------------------------------------
(wrought forged CoCrMo, in compliance with ISO 5832-12). The design incorporates a 12/14 Morse-type female taper and a beveled face that allows for easier reduction of the hip intraoperatively. This femoral head component is offered in both a standard and an eccentric version, and is designed specifically to articulate with Centerpulse Orthopedics acetabular inserts having a Metasul® Inlav.
# INTENDED USE:
The 32mm Epsilon Metasul Acetabular Insert and Metasul Femoral Head are intended for use in total hip arthroplasty for treatment of the following:
- patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., . avascular necrosis, osteoarthritis, and inflammatory joint disease (JJD), e.g., rheumatoid arthritis:
- those patients with failed previous surgery where pain, deformity, or dysfunction . persists:
- . revision of previously failed hip arthroplasty.
# BASIS OF SUBSTANTIAL EQUIVALENCE:
Performance tests, design comparisons, indications for use and labeling of the 32mm Epsion Metasul Acetabular Insert and Metasul Femoral Head demonstrate that they are substantially equivalent in terms of design features, materials, and indications for use to the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol that resembles a stylized caduceus, with three lines representing snakes wrapped around a staff.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 5 2004
Ms. Audrey Swearingen Manager, Regulatory Affairs Zimmer Inc. 9900 Spectrum Drive Austin, Texas 78717
Re: K033634
KU35034
Trade/Device Name: Epsilon™ Metasul® Acetabular 32mm Insert/Femoral Head Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: III Product Code: KWA Dated: August 23, 2004 Received: August 24, 2004
Dear Ms. Swearingen:
We have reviewed your Section 510(k) premarket notification of intent to market the devices in we nave reviewed your bector 9 rQst, presidentially equivalent (for the indications felerenced above and nave determined two marketed predicate devices marketed in interstate for use stated in the encrosury manatiment date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Red Ford, Free L. Days commerce prior to May 20, 1976, the enaounce with the provisions of the Federal Food, Drug, devices that have been recults in access approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Art. The You may, merelove, market the dovice, adoject to an any and registration, listing of t gencial controls provided or wactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 above) ins existing major regulations affecting your device can
may be subject to such additional controls. Existing major and contribution may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Ood on reacts concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issualled of wouldevice complies with other requirements of the Act that IDA has made a determinations administered by other Federal agencies. You must of ally reachar statutes and regulations and adding, but not limited to: registration and listing (21 comply with an the Ace 3 requirements) news and manufacturing practice requirements as set CITY art 6077, labeling (21 CFR Part 820); and if applicable, the electronic (200); and if applicable, the electronic (200 form in the quality bybelins (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Ms. Audrey Swearingen
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and in your a comming of substantial equivalence of your device to a legally premated predicated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mitormational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033634
Device Name:
Epsilon™ Metasul® 32mm Acetabular Insert
Indications for Use:
The Epsilon™ Metasul® Acetabular Insert is intended for use in total hip arthroplasty for treatment of the following:
- patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., . osteoarthritis, post-traumatic arthritis or avascular necrosis and inflammatory joint disease (IJD), e.g., rheumatoid arthritis;
- those patients with failed previous surgery where pain, deformity, or dysfunction . persists;
- revision of previously failed hip arthroplasty. .
Prescription Use
X
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
KO33634 510(k) Number_
Page __ of ___________________________________________________________________________________________________________________________________________________________________
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.