← Product Code [CGA](/productcode/CGA) · K033627

# Q.STEPS BIOMETER G BLOOD GLUCOSE MONITORING SYSTEM (K033627)

_Biomedix, Inc. · CGA · May 21, 2004 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K033627

## Device Facts

- **Applicant:** Biomedix, Inc.
- **Product Code:** [CGA](/productcode/CGA.md)
- **Decision Date:** May 21, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The Q.STEPS™ Biometer G Blood Glucose Monitoring System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood from the fingertip for all ages (excluding neonates.) It is intended for use outside the body (for in vitro diagnostic use) by health care professionals in settings such as clinical laboratories and physician offices laboratories (POLs) as an aid to monitor the effectiveness of diabetes control.

## Device Story

System consists of meter, glucose test strips, control solution, and lancet device. Operates as electrochemical biosensor; uses glucose oxidase enzyme to catalyze oxidation of D-glucose in capillary blood samples. Electron transfer at electrode surfaces generates current proportional to glucose concentration. Used by healthcare professionals in clinical settings to monitor diabetes control. Provides quantitative glucose results to assist clinical decision-making regarding patient glycemic management.

## Clinical Evidence

Clinical and internal studies compared Q.STEPS system results against YSI 2300 STST reference analyzer. Accuracy: >98% of clinical data fell into zones A & B of Clarke Error Grid; bias within 20mg/dL for glucose <100mg/dL. Precision: CV <8% across 20-day testing. Linearity: 50-400 mg/dL. Hematocrit interference: acceptable within 30-60% range. Interference testing: no significant interference observed for 23 common substances at physiological/therapeutic levels, except Uric Acid and high-dose compounds.

## Technological Characteristics

Enzymatic electrochemical biosensor; glucose oxidase catalyst. Amperometric sensing principle. Measures current proportional to glucose concentration. Sample volume 5 µL; test time 15 seconds. Measurement range 50–400 mg/dL. Hematocrit range 30–60%.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Lifescan One Touch Basic Blood Glucose Monitoring System ([K031472](/device/K031472.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
K033627

B. Purpose for Submission:
New device

C. Analyte:
Glucose

D. Type of Test:
Quantitative, electrochemical biosensor

E. Applicant:
Biomedix, Inc.

F. Proprietary and Established Names:
Q. Steps Biometer G Blood Glucose Monitoring System

G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345, Glucose Test System
2. Classification:
Class II
3. Product Code:
CGA, Glucose Oxidase, Glucose
JJX, Single (Specified) Analyte Controls (Assayed and Unassayed)
4. Panel:
Chemistry 75

H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
The Q. Steps Biometer G Blood Glucose Monitoring System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood from the fingertip for all ages (excluding neonates). It is intended for use outside the body (for in vitro diagnostic use) by healthcare professionals in settings such as clinical laboratories and physician office laboratories as an aid to monitor the effectiveness of diabetes control.
3. Special condition for use statement(s):
For use by healthcare professionals only

{1}

Page 2 of 4

4. Special instrument Requirements: N/A

I. Device Description:
The Q. Steps Biometer G Blood Glucose Monitoring System is comprised of the Q. Steps Biometer G meter, Q. Steps Glucose Test Strips, Q. Steps Control Solution and BD Lancet device.

J. Substantial Equivalence Information:
1. Predicate device name(s):
Lifescan One Touch Basic Blood Glucose Monitoring System
2. Predicate K number(s):
K031472
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Test principle | Based on glucose oxidase | same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Intended use | For use by healthcare professionals use | For use by healthcare professionals as well as lay persons with diabetes  |
|  Technology | amperometric | photometric  |
|  Test memory | 99 result capacity | 75 result capacity  |
|  Sample volume | 5 ul | 10 ul  |
|  Measurement range | 50 - 400 mg/dL | 0 – 600 mg/dL  |
|  Test time | 15 seconds | 45 seconds  |
|  Hematocrit range | 30 – 60% | 25 – 60%  |

K. Standard/Guidance Document Referenced (if applicable):
None referenced

L. Test Principle:
The test system uses enzymatic electrochemical biosensor technology. When a drop of blood is applied to the test strip, the oxidation of D-glucose catalyzed by Aspergillus niger causes electron transfer at the electrode surfaces. A current is generated and detected by the meter. The magnitude of the current generated is proportional to the glucose in the sample.

{2}

Page 3 of 4

# M. Performance Characteristics (if/when applicable):

## 1. Analytical performance:

a. Precision/Reproducibility:

Using commercial glucose reference solutions, 5 concentrations were measured for each level. Two measurements were performed twice daily for 20 days.

|  Glucose level (mg/dL) | N | Mean | SD | %CV  |
| --- | --- | --- | --- | --- |
|  50 | 80 | 54.09 | 2.38 | 4.40  |
|  80 | 80 | 83.20 | 2.01 | 2.42  |
|  120 | 80 | 121.19 | 4.89 | 4.03  |
|  200 | 80 | 201.23 | 5.92 | 2.94  |
|  375 | 80 | 369.34 | 6.95 | 1.88  |

b. Linearity/assay reportable range:

Testing was conducted using venous whole blood spiked to five levels of glucose concentration across the measuring range of the device. Four replicates of each level were run on 3 lots of test strips. The bias of the Q. Steps Biometer G vs YSI were calculated. The mean bias at 45 mg/dL was 8.2% and at 400 mg/dL was 5.2%.

c. Traceability (controls, calibrators, or method):

N/A

d. Detection limit:

N/A

e. Analytical specificity:

Interference testing was conducted to determine the effect of 23 select endogenous and exogenous substances. Stock solutions of the substances were prepared and added to venous blood samples spiked to two glucose concentrations and standardized to 45% hematocrit. Control samples were aliquots of the spiked venous samples with no added interferants. Paired differences were analyzed to determine statistically significant differences between two lots of test strips. A p value of &lt; 0.5 was considered to be statistically significant. Interference was observed only in higher than therapeutic dosages of acetaminophen, ascorbic acid, L-Dopa, Dopamine, Methyldopa, and Tolazamide.

f. Assay cut-off:

N/A

## 2. Comparison studies:

a. Method comparison with predicate device:

Two hundred capillary blood samples were tested at 4 sites by healthcare professionals using the Q Steps Biometer G, One Touch Basic, and YSI. The range of samples tested was 55 – 465 mg/dL. The linear regressions were as follows:

$$
Q. Steps vs One Touch \quad y = 0.92x + 9.67, r = 0.97
$$

{3}

Page 4 of 4

Q. Steps vs YSI  $y = 0.96x + 0.94, r = 0.98$

b. Matrix comparison: N/A/

3. Clinical studies:

a. Clinical sensitivity: N/A
b. Clinical specificity: N/A
c. Other clinical supportive data (when a and b are not applicable): see section 2.a.

4. Clinical cut-off: N/A

5. Expected values/Reference range: 70 – 110 mg/dL, fasting, non-diabetic less than 160 mg/dL, one hour after meal less than 140 mg/dL, 2 hours after meal - referenced from Joslin Diabetes Manual

N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K033627](https://fda.innolitics.com/device/K033627)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
