K033590 · Erbe USA, Inc. · FQH · Jan 30, 2004 · General Hospital
Device Facts
Record ID
K033590
Device Name
ERBE HELIX HYDRO-JET
Applicant
Erbe USA, Inc.
Product Code
FQH · General Hospital
Decision Date
Jan 30, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5475
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K033590 · Jan 30, 2004
ERBE HELIX HYDRO-JET
Erbe USA, Inc.
Published clinical literature
The sponsor utilized published clinical literature to support the safety and effectiveness of the device for the expanded indication of neurosurgical soft tissue dissection.
Neurosurgery; Soft tissue dissection; Published literature
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published literature on neurosurgical patients; Retrospective clinical observation/literature review
Patients with various brain tumors (e.g., gliomas, metastatic lesions, meningiomas) and epilepsy; Sample Size: 75 (largest patient population cited)
Not applicable for this study
Ease of handling, accuracy of dissection, vessel sparing, thermal damage assessment
Comparison of Helix Hydro-Jet vs CUSA in neurosurgery; Comparative clinical observation
102 human neurosurgical patients; Sample Size: 102
The Helix Hydro-Jet is intended for the cutting and dissection of soft tissue in neurosurgery and soft tissue such as the liver, kidney, etc. within the abdomen, including Total Mesorectal Excision (TME), in open as well as endoscopic surgery.
Device Story
Helix Hydro-Jet is a hydraulic pressure delivery system using physiological saline to cut/dissect soft tissue. Device utilizes a 120 um nozzle to create a helical water jet stream; non-thermal, selective modality preserves high-fibrin structures (vessels, nerves, ducts, arachnoid membranes). Used in neurosurgery and abdominal surgery; operated by surgeons in OR. Input: physiological saline; Output: high-pressure water jet for tissue dissection. Clinical benefit: precise dissection, reduced tissue necrosis/edema, minimized blood loss, and sparing of vital structures compared to thermal or ultrasonic devices. Modifications include updated suction connector/tubing and increased capacity (1,000 ml to 2,000 ml) suction container/bag.
Clinical Evidence
Evidence includes animal studies (55 rabbits) and human clinical data (102 patients). Animal studies compared Helix Hydro-Jet to CUSA, showing less edema and reduced microglial/astroglial reaction with Helix. Human studies (n=102) demonstrated successful brain parenchyma dissection and vessel sparing without thermal damage. Literature review (largest population 75) confirmed efficacy in soft tissue dissection near vital structures. No significant increase in infection, tumor recurrence, or spreading observed.
Technological Characteristics
Hydraulic pressure delivery system; uses physiological saline. Nozzle size: 120 um. Operating pressure: 3 to 12 bars (43.5 to 174 psi) for neurosurgery; up to 25 bars (362.5 psi) for other applications. Non-thermal, selective cutting mechanism based on tissue density/solidity. Suction container capacity: 2,000 ml. Mechanical device; no software-based algorithms.
Indications for Use
Indicated for cutting and dissection of soft tissue in neurosurgery and abdominal soft tissue (e.g., liver, kidney) including Total Mesorectal Excision (TME) in open or endoscopic surgery.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
{0}------------------------------------------------
JAN 3 0 2004
## 510K SUMMARY
033590
ERBE USA, Inc. Submitted By: 2225 Northwest Parkway Marietta, GA 30067 Tel: 770-955-4400 Fax: 770-955-2577
John Tartal Contact Person:
11/11/03 Date Prepared:
Water Jet Dissector Common Name:
Helix Hydro-Jet™ Trade/Proprietary Name:
Classification Name: Jet Lavage (21CFR880.5475)
FQH Product Code:
Legally Marketed Device: Helix Hydro-Jet
Note: This 510(k) is being submitted to expand the intended use for the Helix Hydro-Jet.
## Device Description:
The Helix Hydro-Jet is a hydraulic pressure delivery system that uses physiological saline to cut and dissect soft tissue. A very detail description of the system can be found in the initial 510(k), Number K012464, Device Description 1-1.
Note: A summary of design control activities, risk analysis, verifications/validations, and biocompatibility can also be found respectively in Sections 4-1, 5-1, 6-1, 9-1, and 8-1 of the initial 510(k). Also, certifications and the declaration of conformity are in the first 510(k) Appendices A and B respectively.
## Intended Use:
The Helix Hydro-Jet is intended for the cutting and dissection of soft tissue in neurosurgery and soft tissue such as the liver, kidney, etc. within the abdomen, including Total Mesorectal Excision (TME), in open as well as endoscopic surgery.
Similarities and Differences of the Modified Device to the Current Device (Predicate Comparison/Substantial Equivalence):
The modified device is the same as the predicate except for slight modifications to the unit's suction connector/tubing as well as the suction container and bag since the last 510(k) submission (K022613). The suction connector/tubing was change so that only
{1}------------------------------------------------
# 510K SUMMARY
the tubing for the suction container/bag could be connected. Also, the confainer and bag were made larger from a 1,000 ml to 2,000 ml capacity (dimensional/size change). Changes to the label and labeling are apart of this submission (See IV-4 to IV-48). The suction connector/tubing, container, and bag were verified/validated in design control. Risk analysis also has been performed with regards to changing the part, component, and accessory.
The significant change involves the expansion of the indications as follows.
1. Addition of using the device to cut and dissect soft tissue in neurosurgery.
The change to the intended use is supported by scientific documentation provided in this Section as a part of Clinical Evaluations (See III-9).
## Conclusion:
All available/known studies involving the Helix Hydro-Jet and like water jet equipment in conjunction with the expanded intended use has been provided in this submission (Note: An exhaustive literature search was performed.). The water jet is a non-thermal, selective instrument which preserves structures with higher fibrin content such as vessels, nerves, and ducts (including arachnoid membranes). With its accuracy and non-thermal modality damage to surrounding tissues is minimized. As a result, in neurosurgery less tissue necrosis and edema can reduce post-operative neurological symptoms. Furthermore, sparing of vessels can result in less blood loss. The clinical documentation provided demonstrates that the Helix Hydro-Jet may be used safely and effectively to cut/dissect soft tissue in neurosurgery. Summary of the studies are as follows.
## Animal Studies
In experimental trials with porcine brain cadaveric tissue, scientists investigated nozzle sizes, pressures, penetration depths (based upon cutting distance and speed), and effects on various brain tissues. As a result, the attributes of the hand piece were better defined. The optimal size of the nozzle/helical water jet stream was found to be 120 um (Note: Current attribute of Helix Hydro-Jet Applicators) and with better technique as well as use of suction, clogging of an applicator and foaming at the surgical site obscuring visibility was virtually eliminated. Also with this jet stream size, linearity in regard to pressure and penetration depths was established. The studies showed brain parenchyma was well dissected using 3 to 12 bars (43.5 to 174 psi). Cuts were extremely sharp and precise. In surrounding tissue no disruption or vacuolization was seen. Depth of penetration was determined to be depended upon the density/consistency of the tissue dissected and ranged from 3 to 19 mm. Vessels as small as 0.3 mm were found to be spared between 3 and 7 bars (43.5 to 101.5 psi).
## Human Studies
The Helix Hydro-Jet was used on patients with various brain tumors (e.g., gliomas, metastastic lesions, meningiomas, etc.) and epilepsy (Note: Published literature largest patient population was 75.). In general; the applicators were easy to handle,
{2}------------------------------------------------
#### 510K SUMMARY
dissections (separation and removal) were accurate/successful on malleable soft tissues (brain parenchyma or tumors), vessels in some cases were spared up to 25 bars (362.5 psi), and there was no thermal damage to surrounding tissue. In particular, the device was found to be efficacious in soft tissue dissection near vital structures where precision and less tissue trauma is paramount. However, due to the advantage of being selective based upon solidity/density, the device becomes harder to use on firmer masses or structures (e.g., to dissect tumor capsules, separate/remove dense tumors, cut fibrous appendages like trabecular structures, etc.). As a result, this issue was addressed as a precaution in the User Manual. Nonetheless, the Helix Hydro-Jet showed to be an effective tool and should be included in the neurosurgical armamentarium (Note: The higher risks of infection or tumor spreading due to the almost permanent water rinsing has been raised as a possible potential disadvantages of the device. However, current results do not support the complications of a higher risk of infection, an increase of turnor recurrence, or tumor spreading.).
#### Investigations Comparing Helix versus CUSA
Tissue trauma after neurosurgical procedures on 55 rabbits was investigated respectively with the Helix Hydro-Jet and Cavitational Ultrasonic Surgical Aspirator (CUSA). Neurosurgery was performed on the animals, and then they were sacrificed at various time points for tissue examination. At all time points, there was less edema formation as well as less intense microglial and astroglial reaction in neurological tissues of rabbits operated on with the Helix Hydro-Jet in comparison to the CUSA. As a result, the water jet has to be considered as less traumatic than the ultrasonic aspirator. Furthermore the Helix Hydro-Jet was compared to CUSA in neurosurgery on 102 humans. The Helix Hydro-Jet was used in combination with conventional methods in tissue dissection and aspiration. In comparison with CUSA, there were two specific advantages with the Helix Hydro-Jet (1) The waterjet enabled brain parenchyma tissue dissection and vessel coagulation without thermal damage to the remaining tissue as well as (2) dissection of tissue without damage to the arachnoid membranes. CUSA could not achieve these accolades concluding that in certain conditions the waterjet is more suitable.
In conclusion, the use of the Helix Hydro-Jet on soft tissue in neurosurgery was found to be safe and effective.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 5 0 2004
Mr. John Tartal Manager, Quality Assurance/Regulatory Affairs ERBE USA, Inc. 2225 Northwest Parkway Marietta, Georgia 30067
Re: K033590
Trade/Device Name: ERBE Helix Hydro-JetTM Regulation Number: 21 CFR 880.5475 Regulation Name: Jet lavage Regulatory Class: II Product Code: FQII Dated: November 11, 2003 Received: November 10, 2003
Dear Mr. Tartal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CHR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. John Tartal
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Sincerely yours,
Mark N Melkersen
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (If Known): K033590
Device Name: ERBE Helix Hydro-Jet™
Indications for Use:
The Helix Hydro-Jet is intended for the cutting and dissection of soft tissue in neurosurgery and soft tissue such as the liver, kidney, etc. within the abdomen, including Total Mesorectal Excision (TME), in open as well as endoscopic surgery.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
fo Mark A. Milliken
Rossuration 3
K033590
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.