← Product Code [BZG](/productcode/BZG) · K033518

# LOTUS PULMONARY FUNCTION FILTER, MODELS CPF 101, CPF 102, CPF 103 (K033518)

_Custom Plastic Products · BZG · Mar 10, 2004 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K033518

## Device Facts

- **Applicant:** Custom Plastic Products
- **Product Code:** [BZG](/productcode/BZG.md)
- **Decision Date:** Mar 10, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1840
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The Custom Plastic Products Pulmonary Function Filter is intended to prevent cross-contamination between the patient and the spirometers and pulmonary function testing instruments.

## Device Story

The Lotus Pulmonary Function Filter is a disposable, barrier-type device designed for use with spirometers and pulmonary function testing instruments. It consists of a plastic housing containing a filter medium with >99% bacterial filtration efficiency. The device is placed in the breathing circuit to facilitate both inspiratory and expiratory airflow while preventing the transmission of bacteria between the patient and the testing equipment. It is intended for use in clinical settings where pulmonary function testing is performed. By acting as a physical barrier, the filter reduces the risk of cross-contamination, thereby protecting both the patient and the diagnostic instrumentation.

## Clinical Evidence

Bench testing only. Testing included First Article Inspections, material verification, cytotoxicity, and bacterial filtration efficiency. Standard Force Vital Capacity tests were performed comparing the Lotus Filter to the predicate device to ensure accuracy and equivalence in clinical measurement outcomes.

## Technological Characteristics

Disposable barrier filter; plastic housing; >99% bacterial filtration efficiency; designed for bidirectional (inspiratory/expiratory) airflow. No electronic components, software, or energy sources.

## Regulatory Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

## Predicate Devices

- Clear Advantage Pulmonary Function Filter ([K951410](/device/K951410.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 033518

| 510(k) Summary                    |
|-----------------------------------|
| (As required by 21 CFR 807.92(a)) |

Submitter Information A.

> Custom Plastic Products, Inc. 4519 Hastings Court Chino, CA 91710

| Phone Number: | 949-364-9349 |
|---------------|--------------|
| Fax Number:   | 949-347-1204 |

Contact: Jim Barley Regulatory Affairs Date:

October 24, 2003

- Device Information B.

|    | Trade/Proprietary Name: | Lotus Pulmonary Function Filter                                       |
|----|-------------------------|-----------------------------------------------------------------------|
|    | Common name of device:  | Pulmonary Function Filter                                             |
|    | Classification Name:    | Filter, Bacterial, Breathing-Circuit                                  |
| C: | Predicate Device:       | Creative Biomedics, Inc.<br>Clear Advantage Pulmonary Function Filter |
|    | Predicate 510(k) #:     | K951410                                                               |

Device Description: D.

> The Custom Plastic Products Pulmonary Function Filter is a disposable barrier type filter intended to prevent contamination cross-contamination when using spirometers and pulmonary function testing instruments. The filter also prevents the transmission of harmful bacteria from the instruments to the patient.

The device consists of a plastic housing with a filtering efficiency greater than The filter is designed for inspiratory and expiratory air flow. 99%.

{1}------------------------------------------------

## Intended Use: E.

The Custom Plastic Products Pulmonary Function Filter is intended to prevent cross-contamination between the patient and the spirometers and pulmonary function testing instruments.

- Comparison of Required Technological Characteristics: F.
Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Custom Plastic Products Lotus Filter and the cited predicate device.

- Summary and Conclusion of Nonclinical and Clinical Tests: G.
Prior to testing, First Article Inspections were conducted on all components. In addition, material verification was performed on all components.

Cytotoxicity and Bacterial Filtration Efficiency Testing were performed on the device's filter.

To shown that test results using the Lotus Filter were accurate, Standard Force Vital Capacity Tests were performed using both the Lotus Filter and the predicate device. The results of testing showed that test results using the Lotus Filter were substantially equivalent to the Creative Biomedics Inc. Clear Advantage Filter.

Conclusion:

The Custom Plastic Products Lotus Filter is substantially equivalent to the Creative Biomedics Inc. Clear Advantage Filter in indications for use and technological characteristics.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 0 2004

Mr. Jim Barley President, JB and Associates 28481 La Falda Laguna Niguel, CA 92677

Re: K033518

Trade/Device Name: Lotus Pulmonary Function Filter Regulation Number: 21 CFR 868.1840 Regulation Name: Spirometer, Diagnostic (Accessory to) Regulatory Class: II Product Code: BZG Dated: February 23, 2004 Received: February 24, 2004

Dear Mr. Barley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Jim Barley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Chris

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K033518

Lotus Pulmonary Function Filter Device Name:

Indications For Use:

The Custom Plastic Products Pulmonary Function Filter is intended to reduce cross-contamination between the patient and the spirometers and pulmonary function testing instruments.

Prescription Use (Part 21 CFR 801 Subpert D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of Device Evaluation (ODE)

Ktin

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number: K0335

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K033518](https://fda.innolitics.com/device/K033518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).