GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM
K033504 · General Electric Co. · LNH · Nov 25, 2003 · Radiology
Device Facts
| Record ID | K033504 |
| Device Name | GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM |
| Applicant | General Electric Co. |
| Product Code | LNH · Radiology |
| Decision Date | Nov 25, 2003 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 892.1000 |
| Device Class | Class 2 |
Intended Use
The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles
Device Story
The 0.35T Signa Ovation with Excite is an open, whole-body magnetic resonance imaging (MRI) system utilizing a permanent magnet. It acquires 2D and 3D images by detecting the spatial distribution of protons (hydrogen nuclei) based on proton density, T1, T2, and flow properties. The system supports various pulse sequences and imaging planes (sagittal, coronal, axial, oblique, double oblique). It is operated by trained clinicians in a clinical setting. The 'open' architecture facilitates interventional procedures when used with MR-compatible accessories like biopsy needles and in-room displays. The system transforms raw MR signals into anatomical images via Fourier transformation techniques. Physicians interpret these images to assist in clinical diagnosis. The Excite technology and improved gradient performance aim to provide higher resolution and shorter scan times compared to the predicate.
Clinical Evidence
Bench testing only. The device was evaluated against NEMA performance standards and IEC 601-1 (Medical Equipment Safety) and IEC 601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis) standards.
Technological Characteristics
Permanent magnet MRI system operating at 0.35T. Features include four independent receive channels and updated gradient amplifiers. Supports 2D/3D Fourier transformation image reconstruction. Open-bore design. Complies with IEC 601-1 and IEC 601-2-33 standards.
Indications for Use
Indicated for diagnostic imaging of internal structures and organs of the entire body (musculoskeletal, vascular, cardiac, neuro) in patients requiring MR imaging. Can be used for interventional procedures with MR-compatible devices.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Signa® 0.35T MFO/i MR System (K002611)
Related Devices
- K012875 — SIGNA OPENSPEED MAGNETIC RESONANCE SYSTEM · GE Medical Systems · Nov 20, 2001
- K992746 — SIGNA HFO/I MAGNETIC RESONANCE SYSTEM · Ge Medical Systems, Inc. · Nov 9, 1999
- K023442 — EXTRA LARGE BODY COIL · GE Medical Systems · Nov 5, 2002
- K032795 — GE 0.7T SIGNA OPENSPEED WITH EXCITE MR SYSTEM · Ge Medical Systems, LLC · Oct 8, 2003
- K033168 — GE 0.2T SIGNA PROFILE/I MR SYSTEM · General Electric Co. · Oct 10, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
General Electric Company P.O. Box 414, Milwaukee, WI 53201
## 510(k) Summary
## This 510/k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201 |
|-----------------|---------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Manager, Regulatory Programs |
| Telephone: | 262- 544-3894 |
| Fax: | 262- 548-4768 |
| Date Prepared: | November 4, 2003 |
### Device Name:
GE 0.35T Signa ®Ovation with Excite Magnetic Resonance System Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH
#### Marketed Device:
The GE 0.35T Signa ®Ovation with Excite MR System is substantially equivalent to the currently marketed Signa® 0.35T MFO/i MR System (K002611) with the main differences being an addition of four independent receive channels and an update to the performance of the gradient amplifiers.
### Device Description:
The 0.35T Signa ®Ovation with Excite Magnetic Resonance System is a modification to the 0.35T Signa® MFO/i Magnetic Resonance System (K002611) which utilizes a permanent magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 0.35T Signa ®Ovation with Excite Magnetic Resonance System features a permanent magnet operating at 0.35T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.
#### Indications for Use:
The 0.35T Signa ®Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa ®Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular frame. The letters are stylized and connected, giving the logo a distinctive and recognizable appearance. The logo is in black and white.
General Electric Company P.O. Box 414, Milwaukee, WI 53201
relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Due to the 'open' design of the 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles
## Comparison with Predicate Device:
The GE 0.35T Signa® Ovation with Excite MR System is a modification of the Signa® MFO/i MR system (K002611) with the main differences being the addition of Excite Technology and the addition of the improved gradient performance.
### Summary of Studies:
The 0.35T Signa® Ovation with Excite Magnetic Resonance System was evaluated to the appropriate NEMA performance standards as well as the IEC 601-1 International Medical Equipment Safety standard and IEC 601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis. The 0.35T Signa® Ovation with Excite Magnetic Resonance System is comparable to the currently marketed Signa® 0.35T Ovation Magnetic Resonance System.
## Conclusion:
It is the opinion of GE that the 0.35T Signa® Ovation with Excite Magnetic Resonance System is substantially equivalent to the Signa® 0.35T MFO/i Magnetic Resonance System. Usage of the 0.35T Signa® Ovation with Excite Magnetic Resonance System does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three overlapping human profiles facing to the right, with flowing lines beneath them, resembling a stylized representation of people and movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 5 2003
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems P.O. Box 414, W-400 MILWAUKEE WI 53201
Re: K033504 Trade/Device Name: GE 0.35T Signa® Ovation with Excite MR System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance imaging system Regulatory Class: II Product Code: 90 LNH Dated: November 4, 2003 Received: November 5, 2003
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "GE Medical Systems" in a bold, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and professional, suggesting a corporate identity.
General Electric Company P.O. Box 414, Milwaukee, WI 53201
## STATEMENT OF INTENDED USE
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
# Device Name: GE 0.35T Signa® Ovation with Excite MR System
# Indications for Use
The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles