PHOSPHOLIN ES AND CALCIUM CHLORIDE

K033471 · R2 Diagnostics, Inc. · GGW · Feb 4, 2004 · Hematology

Device Facts

Record IDK033471
Device NamePHOSPHOLIN ES AND CALCIUM CHLORIDE
ApplicantR2 Diagnostics, Inc.
Product CodeGGW · Hematology
Decision DateFeb 4, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7925
Device ClassClass 2

Indications for Use

The Phospholin ES Activated Partial Thromboplastin Time reagent is a liquid activated reagent with phospholipids derived from soybean lecithin for use in the determination of Activated Partial Thromboplastin Time (APTT) and related coagulation procedures. Phospholin ES is to be used as an APTT reagent (qualitative assay) on patient plasma for the routine screening in the general patient population for deficiencies involving the intrinsic pathway of coagulation. Phospholin ES is sensitive to lupus-like inhibitors.

Device Story

Phospholin ES is a liquid in vitro diagnostic reagent containing ellagic acid (activator) and soybean lecithin-derived phospholipids. Used in clinical laboratories to perform activated partial thromboplastin time (APTT) and related coagulation factor assays. Requires addition of 0.02-0.025M Calcium Chloride to initiate clotting. Operates by measuring time to clot formation in patient plasma samples; sensitive to lupus anticoagulants and deficiencies in factors II, V, VIII, IX, XI, and XII. Healthcare providers use results to screen for coagulation disorders. Benefits include standardized assessment of intrinsic pathway function and identification of lupus-like inhibitors.

Clinical Evidence

Bench testing only. Method comparison study (n=140) against predicate using NCCLS EP9-A2 guidelines yielded linear regression R=0.92 (photo-optical) and R=0.93 (mechanical). Precision evaluated via within-run and between-run testing (CV <3%). Interference testing performed for hemolyzed, icteric, and lipemic samples. Reference range established using 40 healthy adult samples.

Technological Characteristics

Liquid reagent containing ellagic acid and soybean lecithin-derived phospholipids, buffer, and preservatives. Requires 0.02-0.025M Calcium Chloride. In vitro diagnostic reagent for coagulation testing. No specific hardware or software; functions as a chemical reagent in standard coagulation analyzers.

Indications for Use

Indicated for use in the general patient population for routine screening of patient plasma to detect deficiencies involving the intrinsic pathway of coagulation, monitoring heparin therapy, and detection of Lupus Anticoagulants.

Regulatory Classification

Identification

A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K033471 B. Analyte: Activated Partial Thromboplastin Time C. Type of Test: Qualitative, partial thromboplastin time D. Applicant: R2 Diagnostics, Inc. E. Proprietary and Established Names: Phospholin ES and Calcium Chloride F. Regulatory Information: 1. Regulation section: 21 CFR 864.7925 2. Classification: Class II 3. Product Code: GGW 4. Panel: Hematology (81) G. Intended Use: 1. Intended Use: Phospholin ES APTT reagent is an ellagic acid based reagent with soybean phospholipids, buffers, stabilizers and preservatives. Phospholin ES is intended for use as an activated partial thromboplastin time (APTT) reagent. The APTT test is a qualitative assay used in the routine coagulation screening of patient plasma to detect deficiencies in the intrinsic pathway. It is also used to monitor heparin therapy and in the detection of Lupus Anticoagulants. Phospholin ES APTT reagent should be used in an approved clinical setting by trained and qualified laboratory personnel. 2. Indication for use: The Phospholin ES Activated Partial Thromboplastin Time reagent is a liquid activated reagent with phospholipids derived from soybean lecithin for use in the determination of activated partial thromboplastin time (APTT) and related coagulation procedures. Phospholin ES is to be used as an APTT reagent (qualitative assay) on patient plasma for the routine screening in the general {2} Page 2 of 5 patient population for deficiencies involving the intrinsic pathway of coagulation. Phospholin ES is sensitive to lupus-like inhibitors. 3. Special condition for use statement(s): Not applicable 4. Special instrument Requirements: Not applicable ## H. Device Description: The Phospholin ES Activated Partial Thromboplastin Time reagent is a liquid activated reagent with phospholipids derived from soybean lecithin. The reagent also contains buffer, stabilizers, and preservatives. ## I. Substantial Equivalence Information: 1. Predicate device name(s): Dade Behring Actin FSL 2. Predicate K number(s): K863594 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | Phospholin ES | Actin FSL | | Intended Use | APTT reagent with increased sensitivity to lupus anticoagulants | same | | Performance | Normal range: Photo-optical: 23.3-36.4 Mechanical; 27.1-3 36.5 Total precision: %CV <3% Lupus sensitivity: yes Interference substances; Hemolyzed, Lipemic, icteric | Normal range: Photo optical 24.8-32.4 Mechanical 27.7-3 34.1 Total precision: same Lupus sensitivity: yes Interference substances: same | | Differences | | | | Item | Device | Predicate | | Materials | Activator: ellagic acid Phospholipid: Soy | Activator: ellagic acid Phospholipid: soy, rabbit brain | ## J. Standard/Guidance Document Referenced (if applicable): EP9-A2 Method Comparison and Bias Estimation Using Patient Samples, Approved Standard-Second Edition, NCCLS EP15A User Demonstration of Performance for Precision and Accuracy, NCCLS {3} EP7-P Interference Testing in Clinical Chemistry, NCCLS H47-A One Stage Prothrombin Time (PT) Test and Activated Partial Activated Thromboplastin (APTT) Test; Approved Guideline, NCCLS Exner, T. et al (1991) Guidelines for testing revised criteria for lupus anticoagulation's, SSC Subcommittee for Standardization of Lupus Anticoagulants. Throb. Heamost. Mar 4:65(3) 320-322 ## K. Test Principle: In the basic screening test, the activated partial thromboplastin time indirectly measures the formation of thrombin by its action on fibrinogen resulting in a fibrin clot. In the test, citrated plasma is mixed with APTT reagent for a specified period of time (3-5 minutes) at 37°C followed by the addition of prewarmed (37°C) calcium chloride (0.02-0.025M). The clotting time is initiated by the addition of the calcium chloride and the time required for clot formation is the APTT time. Clot detection can be made by manual, electro-mechanical, or photo-optical measurement. ## L. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: The within-run precision was measured using three levels of lyophilized control, 10 vials of each control, tested in triplicate. **Photo-optical** | Sample | Number | Mean (secs.) | SD | %CV | | --- | --- | --- | --- | --- | | Normal | 30 | 29.3 | 0.2 | 0.8 | | Mid Range Abnormal | 30 | 61.7 | 0.4 | 0.7 | | High Range Abnormal | 30 | 88.6 | 0.5 | 0.6 | **Mechanical** | Sample | Number | Mean (secs.) | SD | %CV | | --- | --- | --- | --- | --- | | Normal | 30 | 32.7 | 0.3 | 0.9 | | Mid Range Abnormal | 30 | 66.3 | 0.8 | 1.2 | | High Range Abnormal | 30 | 91.5 | 0.9 | 1.0 | The between-run precision was measured using three levels of lyophilized control, 2 vials of each control, tested in triplicate for 5 days. {4} Page 4 of 5 ## Photo-optical | Sample | Number | Mean (secs.) | SD | %CV | | --- | --- | --- | --- | --- | | Normal | 15 | 30.6 | 0.68 | 2.24 | | Mid Range Abnormal | 15 | 65.2 | 1.81 | 2.78 | | High Range Abnormal | 15 | 91.4 | 2.61 | 2.85 | ## Mechanical | Sample | Number | Mean (secs.) | SD | %CV | | --- | --- | --- | --- | --- | | Normal | 15 | 33.4 | 0.80 | 2.4 | | Mid Range Abnormal | 15 | 66.2 | 0.50 | 0.75 | | High Range Abnormal | 15 | 90.0 | 1.23 | 1.37 | b. Linearity/assay reportable range: Not provided c. Traceability (controls, calibrators, or method): Not provided d. Detection limit: Not provided e. Analytical specificity: Data was collected from 10 hemolyzed, 10 Icteric and 10 Lipemic samples run in triplicate and analyzed according to NCCLS EP7-A Approved Guidelines. The results showed possible interference from hemolyzed, icteric and lipemic samples. f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Data was collected in triplicate from 140 normal and abnormal samples and analyzed using NCCLS EP9-A2 Approved Guidelines. Linear regression analysis yielded: Photo-optical $y = 0.864x + 6.8433$ R=0.92 Mechanical $y = 1.0929x - 2.8217$ R=0.93 b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Not Provided b. Clinical specificity: Not Provided c. Other clinical supportive data (when a and b are not applicable): {5} Page 5 of 5 4. Clinical cut-off: Not provided 5. Expected values/Reference range: A reference study was conducted using frozen plasma samples form 40 normal healthy adult males and females. The APTT results were as follows: | | Mean (secs.) | Reference Range | | --- | --- | --- | | Photo-optical | 29.8 | 23.2-36.4 | | Mechanical | 31.8 | 27.1-36.5 | These values should serve only as guidelines. It is recommended that each laboratory establish its own expected APTT values. M. Conclusion: Based on the review of the information provided, Phospholin ES and Calcium Chloride reagent appears to be substantially equivalent (SE) to the other devices regulated under 21 CFR 864.7925, partial thromboplastin time test, product code GGW.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...