← Product Code FRO · K033453

# KENDALL ULTEC PRO AG+ HYDROCOLLOID DRESSING (K033453)

_Tyco Healthcare · FRO · Apr 29, 2004 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K033453

## Device Facts

- **Applicant:** Tyco Healthcare
- **Product Code:** FRO
- **Decision Date:** Apr 29, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

ULTEC PRO Ag+ Hydrocolloid Dressings are intended for use on venous stasis ulcers, pressure ulcers, arterial ulcers, diabetic ulcers, donor sites, lacerations, post surgical incisions, and other external wounds presenting dry to lightly exuding symptoms.

## Device Story

Sterile, single-use, flexible hydrocolloid wound dressing; designed for dry to lightly exudating wounds. Composition: 65% acrylic hot melt pressure-sensitive adhesive, 25% sodium alginate, 10% inorganic polymer containing silver. Principle of operation: creates moist wound environment; alginate formulation absorbs exudate; silver-containing polymer releases antimicrobial silver to reduce bacterial growth on dressing. Applied by clinicians or patients to external wounds. Benefits: promotes healing via moist environment and antimicrobial barrier properties.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed in accordance with ISO 10993 Part 1 and FDA memorandum G95-1.

## Technological Characteristics

Hydrocolloid dressing; 65% acrylic hot melt pressure-sensitive adhesive, 25% sodium alginate, 10% inorganic silver-containing polymer. Sterile, single-use, flexible form factor. Antimicrobial mechanism via ionic silver release from water-soluble polymer matrix. Biocompatibility per ISO 10993-1.

## Predicate Devices

- Kendall ULTEC PRO Hydrocolloid Dressing ([K943636](/device/K943636.md))
- Arglaes Antimicrobial Film Dressing ([K990810](/device/K990810.md))
- Contreet Hydrocolloid Dressing with silver ([K013525](/device/K013525.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### EXHIBIT # 14

## APR 2 9 2004

510(k) Summary

KO33453

In accordance with section 513(I) of the SMDA and as defined in 21 CFR Part 807,3 final rule dated December 14, 1994, this summary is submitted by:

Kendall, a Division of Tyco Healthcare 15 Hampshire Street Mansfield, MA 02048 Date Prepared: October 23. 2003

- 1. Contact Person
Gail Christie Manager, Regulatory Affairs (508) 261-8440

- 2. Name of Medical Device
Classification Name: Proprietary Name:

Hydrocolloid Dressing ULTEC PRO Ag+ Hydrocolloid Dressing

#### 3. Identification of Legally Marketed Device

The proposed device, Kendall ULTEC PRO Ag+ Hydrocolloid Dressing is substantially equivalent in intended use, function and composition to the Kendall ULTEC PRO Hydrocolloid Dressing (K943636), Arglaes Antimicrobial Film Dressing (K990810), and Contreet Hydrocolloid Dressing with silver (K013525).

#### 4. Device Description

The proposed Kendall ULTEC PRO Ag+ Hydrocolloid Dressing is a sterile, single use, highly flexible wound dressing designed for use on dry to lightly exudating wounds. ULTEC PRO Ag+ Hydrocolloid Dressing was designed to create and maintain a moist wound environment that is optimal for wound healing.

The alginate formulation of ULTEC PRO Ag+ releases anti-microbial silver and absorbs wound exudates. The release of anti-microbial silver enhances the barrier properties of the hydrocolloid by reducing bacterial growth on the dressing.

The proposed wound dressing consists of 65 % acrylic hot melt pressure sensitive adhesive, 25 % sodium alginate, and 10 % inorganic polymer containing silver.

{1}------------------------------------------------

page 2 of 2

#### 5. Device Intended Use

ULTEC PRO Ag+ Hydrocolloid Dressings are intended for use on venous stasis ulcers, pressure ulcers, arterial ulcers, diabetic ulcers, donor sites, lacerations, post surgical incisions, and other external wounds presenting dry to lightly exuding symptoms.

#### ్. Product Comparison

The proposed device has the same technological characteristics as the predicate devices. Both the proposed device and the Contreet Hydrocolloid Dressing with silver consist of a hydrocolloid base with added silver.

Both the proposed ULTEC PRO Aq+ Hydrocolloid Dressing and the Arglaes Dressing utilize a silver based antimicrobial agent that is formulated from ionic silver encapsulated in a water soluble polymer structure. The silver polymer matrix used in the Arglaes product contains the same elemental components as those used in the proposed device. The silver polymer used in these devices is obtained from the same manufacturer.

Both the predicate device, UTEC PRO Hydrocolloid Dressing and the proposed ULTEC PRO Ag+ Hydrocolloid Dressing contain adhesive and alginate.

#### 7. Nonclinical Testing

Biocompatibility testing of the proposed device has demonstrated that it meets the requirements of guidelines presented in the 10993 ISO Standard, Part 1, with the FDA modified matrix presented in memorandum G95-1.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized depiction of an eagle or bird-like figure with three wing-like shapes above a wavy line, possibly representing water.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# APR 2 9 2004

Ms. Gail Christic Manager, Scientific Services/Regulatory Affairs Tyco Healthcare/Kendall 15 Hampshire Street Mansfield, Massachusetts 02048

Re: K033453

Trade/Device Name: Kendall Ultec Pro Ag+ Hydrocolloid Dressing Regulatory Class: Unclassified Product Code: FRO Dated: February 10, 2004 Received: February 13, 2004

Dear Ms. Christie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labcling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Scctions 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Gail Christie

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will and w you'll colling of substantial equivalence of your device to a legally prematication " results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mirrhmational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Fal Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

510(k) Number (if known):

K033453

Device Name: Kendall Ultec Pro Ag+ Hydrocolloid Dressing

Indications for Use:

The Ultec Pro Ag+ Hydrocolloid Dressings are indicated for use on venous stasis ulcers, pressure ulcers, arterial ulcers, diabetic ulcers, donor sites, lacerations, post surgical incisions, and other dry to lightly exudating symptoms.

Prescription Use
Use
(Per 21 CFR 801.109)

OR

Over-The-Counter

Please Do Not Write Below This Line – Continue On Another Page If Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

KO33453 510(k) Number.

---

**Source:** [https://fda.innolitics.com/device/K033453](https://fda.innolitics.com/device/K033453)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
