← Product Code [MHX](/productcode/MHX) · K033365

# APEXPRO FH TELEMETRY SYSTEM (K033365)

_Ge Medical Systems Information Technologies · MHX · Nov 6, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K033365

## Device Facts

- **Applicant:** Ge Medical Systems Information Technologies
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Nov 6, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The ApexPro FH Telemetry System is intended for use under the direct supervision of a licensed healthcare practitioner. The system is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The system processes this physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations. The ApexPro FH Telemetry System is intended to be installed in the hospital or clinical environment in order to provide clinicians with patient physiological data, while allowing for patient mobility. These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered. The physiological parameters monitored include ECG, non-invasive blood pressure, non-invasive temperature and SpO2. The ApexPro FH Telemetry System is intended to provide ECG data via Ethernet to the computer platform for processing. The ApexPro FH is also intended to provide physiologic data over the Unity network to clinical information systems for display.

## Device Story

ApexPro FH Telemetry System monitors ambulating patients in clinical/hospital settings; system components include patient-worn acquisition transceivers, transceiver access points with antennas, network infrastructure, and computer platforms for application processing and central station display. Device acquires physiological signals (ECG, NIBP, temperature, SpO2) via patient-worn transceivers; transmits data over RF to access points; processes ECG data via Ethernet to computer platform for arrhythmia detection and parameter limit violation monitoring. Clinicians view processed data on central station displays to inform clinical decision-making. System enables patient mobility during monitoring, potentially enhancing medical procedure effectiveness.

## Clinical Evidence

No clinical data. Bench testing only, including requirements specification review, code inspections, software and hardware testing, safety testing, and environmental testing.

## Technological Characteristics

System comprises patient-worn transceivers, access points, network infrastructure, and computer platforms. Monitors ECG, NIBP, temperature, and SpO2. Utilizes RF transmission for patient data acquisition and Ethernet/Unity network for data processing and display. Software-based arrhythmia detection and parameter limit monitoring.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- ApexPro Telemetry System ([K032369](/device/K032369.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| Section 2 Summary                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Summary of Safety and Effectiveness |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Date:                                      | October 20, 2003                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Submitter:                                 | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Contact Person:                            | Diana M. Thorson<br>Regulatory Affairs Specialist<br>GE Medical Systems Information Technologies<br>Phone: 714-247-4135<br>Fax: 714-247-4155                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device:<br>Trade Name:                     | ApexPro FH Telemetry System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common/Usual Name:                         | Telemetry Monitoring System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Names:                      | 21 CFR 870.1025 Detector and Alarm, Arrhythmia<br>21 CFR 870.2910 Transmitters & Receivers, Physiological Signal, Radio Frequency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device:                          | K032369 ApexPro Telemetry System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Description:                        | The ApexPro FH Telemetry System is composed of six major<br>components:<br>Accessories to the patient worn acquisition transceivers The patient worn data acquisition transceivers The transceiver access points with antenna The network infrastructure A computer platform running the ApexPro Telemetry Application A computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Intended Use:                              | The ApexPro FH Telemetry System is intended for use under the direct<br>supervision of a licensed healthcare practitioner. The system is<br>designed to acquire and monitor physiological data for ambulating<br>patients within a defined coverage area. The system processes this<br>physiological data to detect various ECG arrhythmia events and select<br>physiological parameter limit violations.<br>The ApexPro FH Telemetry System is intended to be installed in the<br>hospital or clinical environment in order to provide clinicians with patient<br>physiological data, while allowing for patient mobility. These systems<br>are typically deployed in sub acute care areas in hospitals or clinical<br>sites where patient mobility can enhance the effectiveness of the<br>medical procedures administered.<br>The physiological parameters monitored include ECG, non-invasive<br>blood pressure, non-invasive temperature and SpO2. The ApexPro FH<br>Telemetry System is intended to provide ECG data via Ethernet to the<br>computer platform for processing. The ApexPro FH is also intended to<br>provide physiologic data over the Unity network to clinical information<br>systems for display. |

{1}------------------------------------------------

- The ApexPro FH employs the same functional technology as the Technology: predicate devices. The system architecture has taken advantage of improvements in signal processing technology as well as advances in RF component technologies to improve performance and level of integration.
- Test Summary: The ApexPro FH complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the system:
	- Requirements specification review .
	- Code inspections .
	- Software and hardware testing .
	- Safety testing .
	- Environmental testing .
	- . Final validation
	- The results of these measurements demonstrated that the ApexPro FH Conclusion: System is as safe, as effective, and performs as well as the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/11 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 6 2003

GE Medical Systems Information Technologies c/o Ms. Diane M. Thorson Corporate Regulatory Affairs 8200 West Tower Avenue Milwaukee, WI 53223

Re: K033365

Trade Name: ApexPro FH Telemetry System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class II (three) Product Code: MHX Dated: October 20, 2003 Received: October 21, 2003

Dear Ms. Thorson:

'We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Diane M. Thorson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Afgar May

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

510(k) filed on October 10, 2003

ApexPro FH Telemetry System Device Name:

Indications For Use:

The ApexPro FH Telemetry System is intended for use under the direct supervision of a licensed healthcare practitioner. The system is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The system processes this physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations.

The ApexPro FH Telemetry System is intended to be installed in the hospital or clinical environment in order to provide clinicians with patient physiological data, while allowing for patient mobility. These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered.

The physiological parameters monitored include ECG, non-invasive blood pressure, non-invasive temperature and SpO2. The ApexPro FH Telemetry System is intended to provide ECG data via Ethernet to the computer platform for processing. The ApexPro FH is also intended to provide physiologic data over the Unity network to clinical information systems for display.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

|                     | Concurrence of CDRH, Office of Device Evaluation (ODE) |  |
|---------------------|--------------------------------------------------------|--|
| (Division Sign-Off) |                                                        |  |
|                     | Division of Cardiovascular Devices                     |  |
| 510(k) Number       | K033365                                                |  |
|                     |                                                        |  |

Prescription Use_X (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K033365](https://fda.innolitics.com/device/K033365)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
