ACUITY WITH CONEBEAM COMPUTED TOMOGRAPHY

K033339 · Varian Medical Systems, Inc. · KPQ · Dec 30, 2003 · Radiology

Device Facts

Record IDK033339
Device NameACUITY WITH CONEBEAM COMPUTED TOMOGRAPHY
ApplicantVarian Medical Systems, Inc.
Product CodeKPQ · Radiology
Decision DateDec 30, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5840
Device ClassClass 2

Indications for Use

The Acuity Radiation Therapy Simulator is to be used in radiation therapy simulation, using a fluoroscopic and/or radiographic x-ray system for visualizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irridation field to be applied. The Acuity ConeBeam Computed Tomography (CBCT) is intended to assist the Radiation Oncologist in acquiring 3D "multislice" planning data in patient set-ups for the purpose of radiation therapy treatment planning and patient positioning.

Device Story

Acuity with CBCT is a radiation therapy simulator; integrates conventional fluoroscopic/radiographic simulation with CBCT imaging. Device acquires 3D multislice planning data (approx. 340 0.5mm slices) via single gantry rotation (15-17cm volume). Operated by radiation oncologists/technicians in clinical settings. Input: X-ray projections during gantry rotation. Transformation: Reconstruction of volumetric CT data from projections. Output: Digital 3D images displayed on computer screen; integrated into Vision Image database. Clinical utility: Enables treatment planning and patient positioning without moving patient from treatment position; provides CT-like planning data when conventional CT is unavailable.

Clinical Evidence

Bench testing only. Device tested for compliance with 21 CFR 1020.30 (Diagnostic x-ray systems) and 21 CFR 1020.33 (Computed tomography equipment). No clinical data provided.

Technological Characteristics

Stationary X-ray system; integrates fluoroscopic/radiographic and CBCT modes. CBCT reconstruction capability: 15-17cm volumetric data per rotation. Connectivity: Integrates with Vision Image database. Standards: Tested per 21 CFR 1020.30 and 1020.33.

Indications for Use

Indicated for patients undergoing radiation therapy simulation requiring visualization of treatment volumes and confirmation of therapeutic irradiation field size and position, and for patients requiring 3D multislice planning data for treatment planning and positioning where conventional CT is not possible.

Regulatory Classification

Identification

A radiation therapy simulation system is a fluoroscopic or radiographic x-ray system intended for use in localizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irradiation field produced. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 3 0 2003 Image /page/0/Picture/1 description: The image shows the logo for Varian Medical Systems. The logo consists of the word "VARIAN" in a bold, sans-serif font, with a stylized "i" that resembles a medical symbol. Below the word "VARIAN" is the phrase "medical systems" in a smaller, sans-serif font. Varian Medical Systems, Inc. 3100 Hansen Way Palo Alto, CA 94304-1038 115A tel +1 650 493 4000 www.varian.com 510(k) Summary \$K033339\$ P.142 The following information is provided following the format of 21 CFR 807.92 for the Acuity with Conebeam Computed Tomography 1. Submitter: Varian Medical Systems 3100 Hansen Way M/S H055 Palo Alto, CA 94304-1129 Contact Name: Vy Tran Phone: (650) 424-5731 Fax: (650) 842-5040 Email: vy.tran@varian.com Date summary was prepared: October 16, 2003 - 2. Name of the Device: Trade/Proprietary Name: Common or Usual Name: Classification Name: Acuity with Conebeam Computed Tomography Acuity with Conebeam Computed Tomography Acuity CBCT Stationary x-ray system 21 CFR §892.1680 Class II Product Code: KPR ## 3. Predicate Devices to claim substantial equivalence: - a. Acuity Radiation Therapy Simulator, K023052 - b. Ximatron Scanvision, K981056 - 4. Description of the Device: Acuity with ConeBeam Computed Tomography (CBCT) is intended to assist the Radiation Oncologist in acquiring 3D multislice planning data in patient set-ups where conventional (Computerized Tomography) CT imaging is not possible. CBCT is a new feature to the Varian Medical Systems Radiation Therapy Simulator, "Acuity" (K023052), and acquires CT slice information as digital images from the Acuity simulator. The images can be viewed and manipulated directly on a computer screen and placed into the Vision Image database for the purpose of radiation therapy treatment planning and patient positioning. CBCT provides Acuity the capability of acquiring and reconstructing 15-17cm of volumetric data in one gantry rotation. This represents approximately 340 0.5mm slices. Acuity combines conventional and CT simulation into one product. Acuity can change from radiographic and fluoroscopic mode for isocenter localization to CT acquisition mode for treatment planning. This is accomplished with no movement of the patient from the actual treatment position. {1}------------------------------------------------ - 5. Intended Use Statement: The Acuity Radiation Therapy Simulator is to be used in radiation therapy simulation, using a fluoroscopic and/or radiographic x-ray system for visualizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irridation field to be applied. The Acuity ConeBeam Computed Tomography (CBCT) is intended to assist the Radiation Oncologist in acquiring 3D "multislice" planning data in patient set-ups for the purpose of radiation therapy treatment planning and patient positioning. - 6. Summary of the Technological Characteristics: the Acuity Radiation Therapy Simulator, K023052 has been modified to include Conebeam Computerized Tomography as a new feature. This new feature will allow the Acuity Radiation Therapy Simulator to acquire 3D multi-slice planning data in patient set-ups for the purpose of radiation therapy treatment planning and patient positioning which is substantially equivalent to the Varian Medical Systems Scanvision device, K981056. The substantial equivalence comparison chart provides a comparison of the technological charactcristics to those of the predicate devices. The chart is located in Tab G. - 7. Performance Standards Acuity with Conebeam Computed Tomography was tested in accordance with 21 CFR&1020 - Performance standards for ionizing radiation emitting products. Specifically §1020.30 - Diagnostic x-ray systems and their major components and §1020.33 - Computed tomography (CT) equipment. The results of the tests demonstrated that the Acuity CBCT met the performance standards requirements. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure embracing a bird, symbolizing the department's mission to protect and promote the health and well-being of Americans. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## DEC 3 0 2003 Mr. Vy Tran Corporate Director, Regulatory Affairs Varian Medical Systems, Inc. 3100 Hansen Way PALO ALTO CA 94304-1038 Re: K033339 Trade/Device Name: Acuity with ConeBeam Computerized Tomography Regulation Number: 21 CFR 892.5840 Regulation Name: Radiation therapy simulation system Regulatory Class: II Product Code: 90 KPO Dated: October 16, 2003 Received: October 17, 2003 Dear Mr. Tran: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Driner I statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brygdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ 510(k) Number (if known): Device Name: Acuity with Conebeam Computed Tomography Indications For Use: The Acuity Radiation Therapy Simulator is to be used in radiation therapy simulation, using a fluoroscopic and/or radiographic x-ray system for visualizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irridation field to be applied. The Acuity ConeBeam Computed Tomography (CBCT) is intended to assist the Radiation Oncologist in acquiring 3D "multislice" planning data in patient set-ups for the purpose of radiation therapy treatment planning and patient positioning. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) *Prescription Use* Daniel A. Leppanen (Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...