K033307 · Nonin Medical, Inc. · DQA · May 6, 2004 · Cardiovascular
Device Facts
Record ID
K033307
Device Name
NVISION DATA MANAGEMENT SOFTWARE
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 6, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
nVISION® Data Management Software (nVISION) is an optional accessory for compatible Nonin oximeters with memory playback. nVISION® software is indicated for use with the following Nonin oximeters: 9700, 3100, 9600, 2120, 2500, 8500, 8600, 8800, and 9840. It is intended for use by healthcare professionals when (1) transferring data from pulse oximeters to computers in order to maintain individual records of pulse oximetry data, (2) reviewing that data according to user-selected parameters, and (3) generating reports.
Device Story
nVISION Data Management Software functions as an accessory for Nonin pulse oximeters with memory playback. It runs on a personal computer to download, store, and display oximetry data. Healthcare professionals use the software to review data based on user-selected parameters and generate reports for clinical interpretation. The software does not provide automated diagnoses or treatment suggestions; it serves as an adjunct tool for record-keeping and data visualization. The clinician remains responsible for all data analysis and clinical decision-making.
PC-based software for data management and graphical display of pulse oximetry data. Compatible with specific Nonin oximeter models (9700, 3100, 9600, 2120, 2500, 8500, 8600, 8800, 9840).
Indications for Use
Indicated for use by healthcare professionals to manage pulse oximetry data from compatible Nonin oximeters (models 9700, 3100, 9600, 2120, 2500, 8500, 8600, 8800, 9840).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
# MAY - 6 2004
.
KA33307
## SECTION 2. Summary and Certification
#### A. 510(K) Summary
-----
| Submitter: | Nonin Medical, Inc. |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John R. Dalpee<br>Director of Regulatory Affairs<br>Nonin Medical, Inc.<br>2605 Fernbrook Lane N.<br>Plymouth, MN 55447<br>Phone: (763) 577-3166<br>Fax: (763) 553-7807 |
| Date Prepared: | 10 October, 03 |
| Trade Name: | nVISION® Data Management Software (K033307). Class II, 21<br>CRF 870.2700. |
| Classification Name<br>And Number: | Pulse Oximeter Data Management Software |
| Product Code: | DQA |
| Predicate Devices: | nVISION® Data Management Software is substantially equivalent<br>to Respironics' Profox Software (K001708) and Nellcor's Score<br>Software (K961450). |
| Device Description: | nVISION® Data Management Software (nVISION) is an optional<br>accessory for compatible Nonin oximeters with memory playback.<br>nVISION® software is indicated for use with the following Nonin<br>oximeters: 9700, 3100, 9600, 2120, 2500, 8500, 8600, 8800, and<br>9840. nVISION® software runs on a personal computer and<br>provides a graphical display of downloaded oximetry data for<br>review and interpretation by a clinician. This software also allows<br>oximetry data and patient information to be saved in a "library" for<br>future retrieval and analysis. The healthcare professional using<br>nVISION® software is solely responsible for selecting the analysis<br>criteria used to calculate summary statistics included in the reports.<br>nVISION® software is an adjunct system requiring human<br>interpretation of results; it does not suggest a course of treatment or<br>generate a diagnosis. |
{1}------------------------------------------------
| Intended Use: | nVISION® Data Management Software (nVISION) is an optional<br>accessory for compatible Nonin oximeters with memory playback.<br>nVISION® software is indicated for use with the following Nonin<br>oximeters: 9700, 3100, 9600, 2120, 2500, 8500, 8600, 8800, and<br>9840. It is intended for use by healthcare professionals when (1)<br>transferring data from pulse oximeters to computers in order to<br>maintain individual records of pulse oximetry data, (2) reviewing<br>that data according to user-selected parameters, and (3) generating<br>reports. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional and<br>Safety Testing: | Representative samples of nVISION® software successfully<br>underwent performance testing to demonstrate appropriate<br>functional characteristics. |
| Conclusion: | nVISION® Data Management Software is substantially equivalent<br>to Respironics' Profox Software (K001708) and Nellcor's Score<br>Software (K961450).<br><br>This conclusion is based on the fact that nVISION® software is<br>equivalent to the predicate devices in terms of functional design,<br>indications for use, principles of operation, software platform, and<br>hardware requirements. |
:
:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--(
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a bird or a stylized human figure.
MAY - 6 2004
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
John R. Dalpee Director of Regulatory Affairs Nonin Medical, Incoporated 2605 Fernbrook Lane, North Minneapolis, MN 55447-4755
Re: K033307
Trade/Device Name: nVision Data Management Software Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: March 12, 2004 Received: March 15, 2004
Dear Mr. Dalpee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality
{3}------------------------------------------------
Page 2 - John R. Dalpee
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Qul
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use Page
510(K) Number : 033307
Device Name: nVISION® Data Management Software
## Indications for Use:
nVISION® Data Management Software (nVISION) is an optional accessory for compatible Nonin oximeters with memory playback. nVISION Is Indicated for use with the following Nonin oximeters: 9700, 3100, 9600, 2120, 2500, 8500, 8600, 8800, and 9840. It is intended for use by healthcare professionals when (1) transferring data from pulse oximeters to computers in order to maintain Individual records of pulse oximetry data, (2) reviewing that data according to user-selected parameters, and (3) generating reports.
Prescription Use
(Part 21 CFR 801 Subpart D)
And/Or
Over-The Counter Use (21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cain Taylor
(Division Sigh-Off) Division of Anesthealology, General Hospital, Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.