PRECISET DAT PLUS II AND CFAS DAT QUALITATIVE PLUS CALIBRATORS
Applicant
Roche Diagnostics Corp.
Product Code
DKB · Clinical Toxicology
Decision Date
Dec 15, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The Preciset DAT Plus II calibrators are designed for the calibration of the Roche assays for drugs of abuse in human urine on automated clinical chemistry analyzers. The Cfas DAT Qualitative Plus calibrator is designed for the qualitative calibration of the Roche assays for drugs of abuse in human urine on automated clinical chemistry analyzers.
Device Story
Preciset DAT Plus II and Cfas DAT Qualitative Plus Calibrators consist of drug-free human urine spiked with specific drugs or metabolites; used to calibrate automated clinical chemistry analyzers for drugs of abuse testing. Preciset DAT Plus II contains benzodiazepines and opiates at multiple levels; Cfas DAT Qualitative Plus contains barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, and propoxyphene at single levels. Operated by laboratory personnel in clinical settings; calibrators establish reference points for assay curves; output enables accurate quantification or qualitative detection of drugs in patient urine samples; facilitates clinical decision-making regarding substance use.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Calibrators consist of drug-free human urine base with added drug/metabolite analytes and preservatives. Preciset DAT Plus II: 2-drug mixture (benzodiazepines, opiates) at up to 6 levels. Cfas DAT Qualitative Plus: 7-drug mixture (barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, propoxyphene) at single levels. Designed for use on automated clinical chemistry analyzers.
Indications for Use
Indicated for the calibration of Roche drugs of abuse assays in human urine samples using automated clinical chemistry analyzers.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Abuscreen OnLine Preciset DAT I Calibrators (K951595)
Submission Summary (Full Text)
{1}------------------------------------------------
## DEC 1 5 2003
DEC 15 2003
K033306
page 1 of 3
| 510(k) Summary | |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. |
| 1) Submitter name, address, contact | Roche Diagnostics Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 521-7637 |
| | Contact Person: Kerwin Kaufman Date Prepared: October 13, 2003 |
| 2) Device name | Proprietary name: Preciset DAT Plus II and Cfas DAT Qualitative Plus Calibrators |
| | Common name: Calibrators, Drug Mixture |
| | Classification name: Clinical Toxicology Calibrator |
| 3) Predicate devices | We claim substantial equivalence to the currently marketed Roche calibrators: |
| | Abuscreen OnLine Preciset DAT I Calibrators, cleared in 510(k) K951595 (formerly Abuscreen OnLine Calibration Pack). |
Continued on next page
{2}------------------------------------------------
## 510(k) Summary, Continued
K033306
paje 2 of 3
Roche Preciset DAT Plus II calibrators contain a mixture of 2 different drugs, 4) Device propared by the quantitative addition of drug or drug metabolite to drug-free Description human urine. Drugs included are benzodiazepines and opiates. The calibrator set contains up to six levels for each drug contained in bottles 1-6. Bottle 1 is negative (drug free) human urine, followed by bottles 2-6 containing increasing amounts of each drug in a multi-analyte mixture. Drug or drug metabolite and their respective levels included are as follows:
> Benzodiazepines: 0, 50, 100, 200, 400 ng/ml (no 6th level) Opiates: 0, 150, 300, 600, 1000, 2000 ng/ml
Roche Cfas DAT Qualitative Plus calibrator contains a mixture of 7 different drugs, prepared by the quantitative addition of drug or drug metabolite to drug-free human urine. Drugs included are barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, and propoxyphene. The calibrator set contains a single level for each drug mixture. Drugs or drug metabolites and their respective levels included are as follows:
Barbiturates: 200 ng/ml Benzodiazepines: 300 ng/ml Cocaine: 150 ng/ml Methadone: 300 ng/ml Opiates: 2000 ng/ml Phencyclidine: 25 ng/ml Propoxyphene: 300 ng/ml
Continued on next page
{3}------------------------------------------------
K033306
page 3 of 3
510(k) Summary, Continued
| 5.) Intended Use | The Preciset DAT Plus II calibrators are designed for the calibration of the Roche assays for drugs of abuse in human urine on automated clinical chemistry analyzers. |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Cfas DAT Qualitative Plus calibrator is designed for the qualitative calibration of the Roche assays for drugs of abuse in human urine on automated clinical chemistry analyzers. |
| 6.) Comparison to the Predicate Device | The Roche Preciset DAT Plus II and Cfas DAT Qualitative Plus multianalyte calibrators are substantially equivalent to other products cleared for commercial distribution intended for similar use. Most notably, they are substantially equivalent to the currently marketed Roche Abuscreen OnLine Preciset DAT I multianalyte calibrators, manufactured for Roche Diagnostics and cleared in 510(k) submission K951595 by Roche Diagnostics. |
| | The Preciset DAT Plus Il calibrators contain a mixture of 2 different drugs, prepared by the quantitative addition of drug or drug metabolite to drug-free human urine with added preservative. Drugs included are benzodiazepines and opiates. The calibrator set contains up to six levels for each drug. |
| | The Cfas DAT Qualitative Plus calibrator contain a mixture of 7 different drugs, prepared by the quantitative addition of drug or drug metabolite to drug-free human urine with added preservative. Drugs included are barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, and propoxyphene. The calibrator contains only a single level of each drug. |
| | The predicate device, Abuscreen OnLine Preciset DAT I calibrators contain a mixture of 9 different drugs, prepared by the quantitative addition of drug or drug metabolite to drug-free human urine with added preservatives and stabilizers. Drugs included are amphetamines, barbiturates, benzodiazepines, cocaine, methadone, methaqualone, opiates, phencyclidine, and propoxyphene. The calibrator set contains four levels for each drug. |
»
.
1 .
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three horizontal bars above it, representing the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 5 2003
Mr. Kerwin Kaufman Regulatory Affairs Consultant Roche Diagnostics Corporation 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457
Re: k033306
> Trade/Device Name: Roche Preciset DAT Plus II and Cfas DAT Qualitative Plus Calibrators Regulation Number: 21 CFR 862.3200 Regulation Name: Clinical toxicology calibrator Regulatory Class: Class II Product Code: DKB Dated: October 13, 2003 Received: October 14, 2003
Dear Mr. Kaufman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
## Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Steven Putman
Steven 1. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
| 510(k) Number (if known): | K033306 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Roche Preciset DAT Plus II and Cfas DAT Qualitative Plus Calibrators |
| Indications for Use: | The Preciset DAT Plus II calibrators are designed for the calibration of the Roche assays for drugs of abuse in human urine on automated clinical chemistry analyzers. |
| | The Cfas DAT Qualitative Plus calibrator is designed for the qualitative calibration of the Roche assays for drugs of abuse in human urine on automated clinical chemistry analyzers. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <div style="display:inline-block;">X</div> OR Over-the-Counter Use ______ |
|---------------------------------------|---------------------------------------------------------------------------|
|---------------------------------------|---------------------------------------------------------------------------|
(Optional format 1-2-96)
Division Sign-Off for Jean Cooper
Office of In Vitro Diagnostic Device Evaluation and Safety
K033306
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
KO333 06 510(k) _______________________________________________________________________________________________________________________________________________________________________
: 上
:・・・
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.