K033173 · FHC, Inc. · GZL · Mar 24, 2004 · Neurology
Device Facts
Record ID
K033173
Device Name
MICRO TARGETING ELECTRODE
Applicant
FHC, Inc.
Product Code
GZL · Neurology
Decision Date
Mar 24, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1330
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The FHC microTargeting® Electrode is intended for use in intra-operative recording of single unit neuronal activity or intra-operative stimulation of neural elements in the brain.
Device Story
The microTargeting® Electrode is a depth electrode used during functional neurosurgery. It functions as an input device for intra-operative single unit neuronal activity recording and as an output device for neural stimulation. The device consists of electrodes and a protective tube, with an optional sterilization tray. It is operated by a neurosurgeon in an intra-operative setting. The recorded signals are processed by external systems to assist the surgeon in identifying neural targets within the brain. By providing real-time electrophysiological feedback, the device aids in precise electrode placement, potentially improving surgical outcomes and reducing risks associated with functional neurosurgical procedures.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use.
Technological Characteristics
Depth electrode for intra-operative neural recording and stimulation. Components include electrodes and a protective tube. Device is a passive electrical conductor used in conjunction with external recording/stimulation systems. No active software or internal power source.
Indications for Use
Indicated for patients undergoing functional neurosurgery requiring intra-operative recording of single unit neuronal activity or stimulation of neural elements in the brain.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
Microrecording Systems Consultants µEEG Pro System 5000 (K991077)
Related Devices
K991522 — MICROTARGETING ELECTRODE · FHC, Inc. · Aug 4, 2000
K140680 — LEADPOINT FOCUS WORKSTSTION, LEADPOINT FOCUS NOTEBOOK SYSTEM · Alpine Biomed Aps · Dec 12, 2014
K082404 — NAVIGATION PROBES · We Sense, Ltd. · Feb 25, 2009
K042208 — ISIS MER SYSTEM/ELEKTA MER SYSTEM · Inomed Medizintechnik GmbH · Mar 18, 2005
K202087 — DIXI Medical Microdeep Micro-Macro Depth Electrodes · Dixi Medical · Nov 18, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Ko 33/13
## 510(k) SUMMARY of SAFETY and EFFECTIVENESS
### A. General Information
| 1. Submitter's Name: | FHC, Inc. |
|-------------------------|---------------------------------------|
| 2. Address: | 9 Main Street<br>Bowdoinham, ME 04008 |
| 3. Telephone Number: | 207-666-8190 |
| 4. Contact Person: | Frederick Haer |
| 5. Date Prepared: | September 15, 2003 |
| 6. Registration Number: | 1226598 |
| B. Device | |
| 1. Name: | microTargeting® Electrode |
| 2. Trade Name: | microTargeting® Electrode |
| 3. Common Name: | Depth Electrode |
| 4. Classification Name: | Depth Electrode |
| 5. Product Code: | 84 GZL |
| 6. Class: | II |
| 7. Regulation Number: | 882.1330 |
{1}------------------------------------------------
K033/73
## C. Identification of Leqally Marketed Devices
| Name | K Number | Date Cleared |
|-----------------------------------------|----------------------|--------------|
| FHC microTargeting® Electrodes | K991522 | Aug. 4, 2000 |
| Radionics Semi-Microelectrode (SME) Kit | K961858 | Aug. 7, 1996 |
| Microrecording Systems Consultants | µEEG Pro System 5000 | |
### D. Description of Device
The FHC microTargeting® Electrode is for intra-operative single unit recording and stimulation during functional neurosurgery.
# microTargeting® Electrode Components
- . Electrodes
- Protective Tube .
# microTargeting® Electrode Accessories
- Sterilization Tray .
## E. Intended Use Statement
The FHC microTargeting® Electrode is intended for use in intra-operative recording of single unit neuronal activity or intra-operative stimulation of neural elements in the brain.
## F. Technological Characteristics Summary
The FHC microTargeting® Electrodes are substantially equivalent to the legally marketed FHC microTargeting® Electrodes (K991522), Radionics Semi-Microelectrode (SME) Kit (K961858), and Microrecording Systems Consultants uEEG Pro System 5000 (K991077).
Differences that exist between these devices, relating to technical specifications, physical appearance, and design do not affect the relative safety and effectiveness of the microTargeting® Electrodes.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird with three curved lines representing its wings or feathers.
Public Health Service
MAR 2 4 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Frederick Haer President/CEO FHC, Inc. 9 Main Street Bowdoinham, Maine 04008
Re: K033173
:
Trade/Device Name: microTargeting® Electrode Regulation Number: 21 CFR 882.1330 Regulation Name: Depth electrode Regulatory Class: II Product Code: GZL Dated: March 2, 2004 Received: March 3, 2004
Dear Mr. Haer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Frederick Haer
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
Ld.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use {PRIVATE }
510(k) Number (if known): K033173
Device Name: microTargeting® Electrodes
Indications For Use:
The FHC microTargeting® Electrodes are intended for use in intraoperative recording of single unit neuronal activity or intra-operative stimulation of neural elements in the brain.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Parrest
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of ___1__
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.