PORTEX 24G PEDIATRIC EPIDURAL AND PERIPHERAL BLOCK ANESTHESIA CATHETER
K033080 · Portex, Inc. · CAZ · Mar 10, 2004 · Anesthesiology
Device Facts
Record ID
K033080
Device Name
PORTEX 24G PEDIATRIC EPIDURAL AND PERIPHERAL BLOCK ANESTHESIA CATHETER
Applicant
Portex, Inc.
Product Code
CAZ · Anesthesiology
Decision Date
Mar 10, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Portex® 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter is designed for use primarily in pediatric applications (children under 12 years, infants and neonates) for continuous or intermittent regional anesthesia, e.g. epidural, caudal and peripheral blocks. The duration of use should not exceed 72 hours.
Device Story
The Portex 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter is a flexible nylon tube used for the administration of anesthetics or analgesics. It features an open-ended, marked tip (1 cm increments up to 20 cm) and includes a stylet and catheter connector. The device is intended for use in clinical settings by healthcare professionals to provide regional anesthesia in pediatric patients. By facilitating continuous or intermittent drug delivery, the catheter enables pain management or anesthesia for up to 72 hours, potentially improving patient comfort and surgical outcomes.
Clinical Evidence
No clinical data. Bench testing only. Performance was validated through testing of dimensional characteristics, flow rate, compression resistance, hub/catheter detachment, deflection resistance, elasticity, tensile strength, and elongation.
Technological Characteristics
Flexible nylon catheter; 24g size (0.022" OD / 0.012" ID); 36-inch length. Features marked tip for depth indication. Includes stylet and connector. Sterile, single-use component of anesthesia conduction mini tray. Materials and ink are identical to previously cleared predicate devices.
Indications for Use
Indicated for pediatric patients (neonates, infants, children <12 years) requiring continuous or intermittent regional anesthesia, including epidural, caudal, and peripheral blocks. Duration of use limited to 72 hours.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
Predicate Devices
B. Braun Periflex Continuous Epidural Pediatric Tray (K962696)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K033080" written in a handwritten style. The text is oriented diagonally, starting from the bottom left and moving towards the upper right. The characters are slightly slanted, giving the impression of cursive writing, and the numbers are clearly distinguishable.
MAR 1 0 2004
# smiths
Smiths Medical ASD, Inc.
Anesthesia and Safety Devices Division
10 Buwman Drive Keene NI103431-0724 USA [et: +] 403 352 3812 Fax: +1 603 352 3703 www.smiths-modical.com
## H: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
#### 510(K) SUMMARY:
## COMPANY INFORMATION:
Smiths Medical ASD, Inc. 10 Bowman Drive Keene, NH 03431 (603) 352-3812 Contact: Cynthia Engelhardt Technical Writer, Regulatory Affairs
## PREPARATION DATE OF SUMMARY:
March 94, 2004
#### TRADE NAME:
Portex® 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter
## COMMON NAME:
Anesthesia Conduction Catheter
# PRODUCT CLASS/CLASSIFICATION:
Class II, 73 BSO, 21 CFR 868.5120
{2}------------------------------------------------
#### PREDICATE DEVICE(S):
Predicate 1: The B. Braun Periflex Continuous Epidural Pediatric Tray, K962696. This is the main predicate device cited for catheter size, design characteristics and indication for use.
Predicate 2: Our current 16g and 18g Epidural Catheter kits, K992471. This predicate device is cited, in conjunction with Predicate 1 above for materials, biocompatibility, sterilization and packaging.
Predicate 3: Our 23g Pediatric Epidural Minipack, K924541. This predicate device is cited, in conjunction with Predicate 1 above for tensile strength and elongation.
Predicate 4: Our 21g High Durometer Epidural; Catheter, K935927. This predicate device is cited, in conjunction with Predicate 2 above for biocompatibility for the ink.
#### DESCRIPTION:
The Portex® 24g Pediatric Epidural and Peripheral Block Anesthesia Catherer is made of flexible, nylon tubing. The catheter is open-ended catheter with finished tip. The tipe of the catheter is marked. The catheter has a single mark at 5 cm from the tip with 1 cm increments, up to 20 cm. The 10 cm mark is indicated by two marks, 15 cm by three marks, and 20 cm by four marks.
The catherer is available in 24g (O.I). 0.022"/I.D. 0.012") size. The catheters have a nominal length of 36 inches. The catheter includes a stylet.
The catheters are provided with a catheter connector to provide a means of administration of anesthetics and/or analgesics. They are provided sterile as a component of a continuous anesthesia conduction catheter mini tray.
#### INDICATIONS FOR USE:
The Portex® 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter is designed for use primarily in pediatric applications (children under 12 years, infants and necures) for continuous or intermittent regional anesthesia, e.g. epidural, caudal and periodenal blocks. The duration of use should not exceed 72 hours.
{3}------------------------------------------------
## TECHNICAL CHARACTERISTICS:
The design of the proposed catheter is similar to predicate 1 marketed under premarket notification K962696 by B. Braun and predicate 3 that we have authorization to market under premarket notification K924541 . The catherer material is identical to predicate 2 that we have authorization to market under premarket notification K965017. The loak is identical to predicate 4 that we have authorization to market under premarket not fination K935927. The technical characteristics of the characteristics fall within the values of the predicate Portex catheters and the B. Braun catheter.
## NON-CLINICAL DATA:
Data submitted demonstrates that the anesthesia conduction catheter performs equivalently to the predicate devices. Data submitted covers; dimensional characteristics, flow rate, compression resistance, hub/catheter detachment, deflection resistance mounts of elasticity, tensile strength, and elongation.
#### CLINICAL DATA:
Not applicable
#### CONCLUSION:
The comparison to the predicate devices demonstrate that the proposed device is safe and effective and is substantially equivalent to the predicate devices.
Very truly yours.
SMITHS MEDICAL ASD, INC.
Cynthia Errigott
Cynthia Engelhardt Technical Writer, Regulatory Affairs
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird with three stylized wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 0 2004
Ms. Cindy Engelhardt Regulatory Affairs/Technical Writer Portex Incorporated 10 Bowman Drive Keene, NH 03431
Re: K033080
Trade Name: Portex 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthetic Conduction Kit Regulatory Class: II Product Code: CA7. Dated: February 18, 2004 Received: February 19, 2004
Dear Ms. Engelhardt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classificd (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 – Ms. Cindy Engelhardt
CFR Part 807); labeling (21 ClFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033080
Device Name: Portex® 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter
Indications for Use:
The Portex® 24g Pediatric Epidural and Peripheral Block Anesthesia Catheter is designed for use primarily in pediatic applications (children under 12 years, infants and neonates) for continuous or intermittent regional anesthesia, e.g. epidural, caudal and peripheral blocks. The duration of use should not exceed 72 hours.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
tAill
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.