ACON OXY ONE STEP OXYCODONE TEST STRIP AND ACON OXY ONE STEP OXYCODONE TEST DEVICE
K033047 · ACON Laboratories, Inc. · DJG · Feb 17, 2004 · Clinical Toxicology
Device Facts
Record ID
K033047
Device Name
ACON OXY ONE STEP OXYCODONE TEST STRIP AND ACON OXY ONE STEP OXYCODONE TEST DEVICE
Applicant
ACON Laboratories, Inc.
Product Code
DJG · Clinical Toxicology
Decision Date
Feb 17, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Intended Use
The ACON® OXY One Step Oxycodone Test Strip and ACON OXY One Step Oxycodone Test Device are rapid chromatographic immunoassays for the qualitative detection of Oxycodone in urine at a cutoff concentration of 100 ng/mL. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis. They are intended for healthcare professionals including professionals at point-of-care sites.
Device Story
ACON OXY One Step Oxycodone Test Strip/Device are lateral flow immunochromatographic assays; used for qualitative screening of Oxycodone in urine. Principle: competitive binding antigen-antibody immunochemistry; mouse monoclonal antibody selectively detects Oxycodone/metabolites at 100 ng/mL cutoff. Operation: manual, instrument-free; urine sample applied to test strip/device. Output: visual colored line in test region (negative) or absence of line (positive); control line indicates proper wicking/volume. Used by healthcare professionals at point-of-care sites. Clinical utility: provides preliminary screening result; positive results require confirmation via GC/MS. Benefit: rapid, point-of-care identification of potential Oxycodone presence to guide further clinical diagnostic or confirmatory testing.
Clinical Evidence
Clinical evaluation of 300 urine specimens compared ACON OXY tests against an FDA-cleared predicate and GC/MS. Results: 97% positive agreement, 100% negative agreement, and 99% overall agreement versus predicate. Comparison against GC/MS at 100 ng/mL cutoff showed 96% positive agreement and 98% negative agreement. Study confirms performance robustness and suitability for point-of-care use.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding format. Mouse monoclonal antibody-based detection. Form factor: test strip or test device. Qualitative visual readout. Instrument-free operation. No specific materials or standards cited.
Indications for Use
Indicated for healthcare professionals at point-of-care sites for the qualitative detection of Oxycodone in human urine at a 100 ng/mL cutoff concentration. Provides preliminary analytical results requiring confirmation by GC/MS.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K043051 — QUICKSCREEN OXYCODONE TEST, MODELS 9120, 9120T, 9120X AND 9121 · Phamatech, Inc. · Feb 7, 2005
K060351 — MEDTOX OXYCODONE · Medtox Diagnostics, Inc. · May 12, 2006
K043507 — ACON OXY II ONE STEP OXYCODONE TEST STRIP; TEST DEVICE · ACON Laboratories, Inc. · Feb 25, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 1 7 2004
page 1 of 4
#### 8. 510(k) SUMMARY
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) number is K033041.
### Submitter:
ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121
Tel.: 858-535-2030 Fax: 858-535-2038
### Date:
September 22, 2003
## Contact Person:
Edward Tung, Ph.D.
### Product Names:
ACON® OXY One Step Oxycodone Test Strip ACON® OXY One Step Oxycodone Test Device
## Common Name:
Immunochromatographic test for the qualitative detection of Oxycodone in urine.
### Regulation Name:
Opiates test system.
### Product Code:
DIG
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the handwritten text "page 2 of 4". The text indicates that this is page 2 of a document that has a total of 4 pages. The handwriting is cursive and slightly slanted.
## Classification Number:
21 CFR. 862.3650
### Device Classification:
The Oxycodone test systems have been classified as Class II devices with moderate complexity. The ACON OXY One Step Oxycodone Test Strip and ACON OXY One Step Oxycodone Test Device are similar to another FDA-cleared device for the qualitative detection of Oxycodone in urine specimens. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis.
## Intended Use:
The ACON® OXY One Step Oxycodone Test Strip and ACON OXY One Step Oxycodone Test Device are rapid chromatographic immunoassays for the qualitative detection of Oxycodone in urine at a cutoff concentration of 100 ng/mL. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis. They are intended for healthcare professionals including professionals at point-of-care sites.
### Description:
The ACON OXY One Step Oxycodone Test Strip and ACON OXY One Step Oxycodone Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of Oxycodone in a urine sample. The test is based on the principle of antigen-antibody immunochemistry. It utilizes the mouse monoclonal antibody to selectively detect elevated levels of Oxycodone and its metabolite in urine at a cutoff concentration of 100 ng/mL. These tests can be performed without the use of an instrument.
A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing Oxycodone at the concentration below the cutoff level will generate a colored-line in the test region. To serve as a procedural control, a coloredline should always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
## Comparison to a Predicate Device:
A comparison of the features of the ACON OXY One Step Oxycodone Test Strip and ACON OXY One Step Oxycodone Test Device versus a FDA-cleared Oxycodone test with 100 ng/mL Oxycodone cutoff is shown below:
{2}------------------------------------------------
page 3 of 4
- Both tests are assays intended for the qualitative detection of Oxycodone in urine samples. .
- Both tests are intended as a screening method that provides a preliminary analytical test . result.
- Both tests are immunochromatographic, lateral flow assays for the rapid detection of . Oxycodone with a visual, qualitative end result.
- . Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody interactions to indicate a positive or negative result.
- . Both tests have a cutoff Oxycodone concentration of 100 ng/mL.
### Safety and Effectiveness Data:
## Accuracy
A clinical evaluation was conducted using 300 clinical urine specimens including approximately 10% of the specimens containing Oxycodone concentration fell between -25% cutoff to +25% cutoff range. This evaluation compared the test results between ACON® OXY One Step Oxycodone Test Strip and ACON® OXY One Step Oxycodone Test Device with a FDA-cleared Oxycodone test; as well as compared against data obtained from the customary Gas Chromatography/Mass Spectrometry analysis. These comparisons of data vielded the following results:
## ACON OXY One Step Oxycodone Test Strip versus FDA-cleared Oxycodone Test:
Positive Agreement: 142 / 146 = 97% (93 % - 99 %*) Negative Agreement: 154 / 154 = 100% (98 % - 100%**) Overall Agreement: 296 / 300 = 99% (97 % - 99 %*) * 95% confidence intervals
- ** Since the proportion can not go above 100%, this is really a 97,5% confidence interval
## ACON OXY One Step Oxycodone Test Device versus FDA-cleared Oxycodone Test:
Positive Agreement: 142 / 146 = 97% (93 % - 99 %*) Negative Agreement: 154 / 154 = 100% (98 % - 100 % **) Overall Agreement: 296 / 300 = 99% (97 % - 99 %*) * 95% confidence intervals
** Since the proportion can not go above 100%, this is really a 97.5% confidence interval
{3}------------------------------------------------
page 4 74.
ACON OXY One Step Oxycodone Test Strip versus data obtained with GC/MS at the cutoff concentration of 100 ng/mL:
| | | Specimen Cut off Range by GC/MS Data | | | | % Agreement with<br>GC/MS Data | |
|---------------------------|----------|--------------------------------------|-----------------|-------------------|-------------------|--------------------------------|---------------------|
| | | Negative | <-25%<br>Cutoff | -25% to<br>Cutoff | Cutoff to<br>+25% | > +25%<br>Cutoff | |
| ACON<br>OXY Test<br>Strip | Positive | 0 | 1 | 4 | 2 | 135 | 96%<br>(92% - 98%)* |
| | Negative | 147 | 2 | 6 | 1 | 2 | 98%<br>(95% - 99%)* |
* Denotes 95% Confidence Interval.
## ACON OXY One Step Oxycodone Test Device versus data obtained with GC/MS at the cutoff concentration of 100 ng/mL
| | | Specimen Cutoff Range by GC/MS Data | | | | % Agreement with GC/MS Data |
|----------------------------|----------|-------------------------------------|---------------|----------------|-------------------------|-----------------------------|
| | | Negative | < -25% Cutoff | -25% to Cutoff | Cutoff to > +25% Cutoff | |
| ACON<br>OXY Test<br>Device | Positive | 0 | 1 | 4 | 135 | 96%<br>(92% - 98%)* |
| | Negative | 147 | 2 | 6 | 2 | 98%<br>(95% - 99%)* |
* Denotes 95% Confidence Interval.
### Performance Characteristics and Other information:
The performance characteristics of ACON OXY One Step Oxycodone Test Strip, ACON OXY One Step Oxycodone Test Device were verified by analytical sensitivity study, specificity and cross reactivity study, interference studies, precision study, read time flex study, temperature flex study, specimen storage and stability study. Study results indicate that these test devices are robust and can perform satisfactorily when used according to the "Indication for Use" statement specified in their package inserts.
### Conclusion:
These clinical studies demonstrated substantial equivalency on performance between the ACON OXY One Step Oxycodone Test Strip, ACON OXY One Step Oxycodone Test Device and a FDA-cleared Oxycodone test with the same Oxycodone cutoff concentration. It is also demonstrated that these tests are safe and effective in qualitatively detecting Oxycodone at a concentration of 100 ng/mL. The POL study demonstrated that these tests are suitable for healthcare professionals including professionals at point-of-care sites.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows a circular logo with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an eagle or bird in flight, composed of several curved lines. The logo is in black and white.
Public Health Service
Food and Drug Administration 2098 Gaither Hoad Rockville MD 20850
FEB 1 7 2004
Edward Tung, Ph.D. Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121
Re: k033047
> Trade/Device Name: ACON® OXY One Step Oxycodone Test Strip ACON® OXY One Step Oxycodone Test Device Regulation Number: 21 CFR 862.3650 Regulation Name: Opiates test system Regulatory Class: Class II Product Code: DJG Dated: December 12, 2003 Received: December 18, 2003
Dear Dr. Tung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Jean M. Cooper, MS, DVM.
Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### 11. INDICATIONS FOR USE
510(k) Number:
ACON® OXY One Step Oxycodone Test Strip Device Name:
ACON® OXY One Step Oxycodone Test Device
<033047
Indications for Use:
The ACON OXY One Step Oxycodone Test Strip and ACON OXY One Step Oxycodone Test Device are rapid chromatographic immunoassays for the qualitative detection of Oxycodone levels in urine at a designated cutoff concentration of 100 ng/mL. They are intended for healthcare professionals including professionals at point-of-care sites.
Ahets
Division Sign Off
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k033047
(Please do not write below this point) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Or
Over-The-Counter Use
(Per 21 CFR 801.109)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.