INTERLAB ALKALINE HEMOGLOBIN ELECTROPHORESIS TEST SYSTEM

K032862 · Interlab S.R.L. · JBD · Feb 3, 2004 · Hematology

Device Facts

Record IDK032862
Device NameINTERLAB ALKALINE HEMOGLOBIN ELECTROPHORESIS TEST SYSTEM
ApplicantInterlab S.R.L.
Product CodeJBD · Hematology
Decision DateFeb 3, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7440
Device ClassClass 2

Indications for Use

The Interlab Alkaline hemoglobin Electrophoresis test system is intended for the separation of normal hemoglobins (A1, A2 and F) as well as certain abnormal or variant hemoglobins (S or D and C or E) using cellulose acetate supported on Mylar®. The test is a screening method for in vitro diagnostic use on the Microtech 672 PC and the Microtech 648 ISO fully automated analyzers. To distinguish hemoglobin S from D and C from E an alternate confirmatory test such as acid hemoglobin electrophoresis is necessary.

Device Story

In vitro diagnostic screening system; separates hemoglobin fractions via alkaline electrophoresis on cellulose acetate strips. Input: hemolyzed red blood cell samples. Operation: automated processing on Microtech 672 PC or 648 ISO analyzers; robotic arm moves strips through staining, destaining, and clearing stations. Output: visual banding patterns or relative percentages of hemoglobin fractions. Used in clinical laboratories; results interpreted by laboratory personnel to identify hemoglobin variants. Provides initial screening; requires follow-up acid electrophoresis for definitive differentiation of specific variants (S vs D; C vs E).

Clinical Evidence

No clinical studies performed. Analytical performance established via bench testing. Precision (intra-assay CV 0.9-3.98%; inter-assay CV 1.58-5.34%), linearity, and limit of detection (0.40-0.47 mg/ml) evaluated. Method comparison study (n=93) demonstrated >0.99 correlation and 100% agreement with reference method.

Technological Characteristics

Cellulose acetate supported on Mylar®; electrophoresis-based separation; semi-quantitative visual identification of hemoglobin bands; automated processing on Microtech 672 PC and 648 ISO analyzers; staining/destaining process; in vitro diagnostic use.

Indications for Use

Indicated for screening of normal (A1, A2, F) and abnormal/variant (S, D, C, E) hemoglobins in human blood samples. Requires confirmatory testing for differentiation of S/D and C/E variants.

Regulatory Classification

Identification

An electrophoretic hemoglobin analysis system is a device that electrophoretically separates and identifies normal and abnormal hemoglobin types as an aid in the diagnosis of anemia or erythrocytosis (increased total red cell mass) due to a hemoglobin abnormality.

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K032862 B. Analyte: Hemoglobin Fractions C. Type of Test: Qualitative, Hemoglobin Electrophoresis D. Applicant: InterLab Scientific Instruments, srl Via Rina Monti N° 26 C.A.P. 00155 Rome, ITALY E. Proprietary and Established Names: InterLab Alkaline Electrophoresis Test System F. Regulatory Information: 1. Regulation section: 21 CFR 864.7440 2. Classification: Class II 3. Product Code: JBD 4. Panel: Hematology (81) G. Intended Use: 1. Intended use: The Interlab Alkaline hemoglobin Electrophoresis test system is intended for the separation of normal hemoglobins (A1, A2 and F) as well as certain abnormal or variant hemoglobins (S or D and C or E) using cellulose acetate supported on Mylar®. The test is a screening method for in vitro diagnostic use on the Microtech 672 PC and the Microtech 648 ISO fully automated analyzers. To distinguish hemoglobin S from D and C from E an alternate confirmatory test such as acid hemoglobin electrophoresis is necessary. {2} Page 2 of 5 2. **Indication for use:** The Interlab Alkaline hemoglobin Electrophoresis test system is intended for the separation of normal hemoglobins (A1, A2 and F) as well as certain abnormal or variant hemoglobins (S or D and C or E) using cellulose acetate supported on Mylar®. The test is a screening method for in vitro diagnostic use on the Microtech 672 PC and the Microtech 648 ISO fully automated analyzers. To distinguish hemoglobin S from D and C from E an alternate confirmatory test such as acid hemoglobin electrophoresis is necessary. 3. **Special condition for use statement(s):** Not applicable 4. **Special instrument Requirements:** The Cellulose Acetate supported on Mylar® strips are for use on the Microtech 672 PC and the Microtech 648 ISO instruments. These instruments employ the use of a robotic arm that moves the strip to the different stations. The instruments are offered as “open systems” and are considered Class I Exempt based on 21 CFR 862.2485, product code JJN. H. **Device Description:** The InterLab Alkaline Electrophoresis Test System provides identification of normal and certain abnormal and variant hemoglobins visually by staining of the separate fractions. The kit contains materials for 24 runs. There are two kits available (SRE205K and SRE157K) providing the ability to run 192 or 288 tests. Each kit contains the following: strips, running and soaking buffer, staining solution, destaining solution, and clearing solution. All reagents are ready to use. I. **Substantial Equivalence Information:** 1. **Predicate device name(s):** Sebia Hydragel 15 Hemoglobin 2. **Predicate K number(s):** K991362 3. **Comparison with predicate:** {3} Page 3 of 5 | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | *InterLab Alkaline Electrophoresis Test System* | *Hydragel 15 Hemoglobin* | | Intended Use | Separation of normal hemoglobins as well as certain abnormal or variant hemoglobins. | Same | | Method | Electrophoresis | Same | | Sample | Hemolyzed separated red blood cells using distilled water | Hemolyzed separated red blood cells using hemolyzing solution | | Sample application | Pipetting Station | Pipetting Station or manual | | Results | Visual or relative percentages of fractions | Same | | Normal Values | Hemoglobin A1: 95.5-98.5% Hemoglobin A2: 1.5-3.5% Hemoglobin F: <2.0% | Hemoglobin A1: ≥ 96.5% Hemoglobin A2: ≤ 3.5% Hemoglobin F: <2.0% | | Differences | | | | Item | Device | Predicate | | Support Medium | Cellulose Acetate on Mylar | Agarose Gel | | Reagents | Running Buffer Staining Solution (*Ponceau red*) Destaining Solution Clearing Solution | Buffered Strips **Amidoblack** Stain Solution Hemolyzing Solution Destaining Solution Wash Solution Saline | | Equipment | Microtech 684 ISO System Microtech 672 PC System | Hydrasys System Micropipettor | J. Standard/Guidance Document Referenced (if applicable): EP5A Evaluation of Precision Performance of Clinical Chemistry Device Approved Guideline, NCCLS K. Test Principle: The principle of hemoglobin electrophoresis is based on the fact hemoglobin molecules in an alkaline solution have a net negative charge and move toward the anode in an electrophoretic system. The pattern of a normal adult displays predominantly Hemoglobin (Hb) A1, along with HbF, and HbA2, when the concentrations fall within normal ranges. The presence of abnormal bands is indicative of variant hemoglobins in the blood sample. {4} Page 4 of 5 # L. Performance Characteristics (if/when applicable): ## 1. Analytical performance: ### a. Precision/Reproducibility: **Intra-assay** The intra-assay precision was measured using an abnormal control and three different patient samples run in replicate (4-6) on 8 cellulose acetate strips. The patterns were visually inspected and found to be qualitatively identical. Relative percentages of the fractions resulted in CV ranges from 0.9-3.98%. **Inter-assay** Inter-assay precision was measured using one abnormal control and three different patients run in replicates (4-6) on one cellulose acetate strip over ten days. The patterns were visually inspected and found qualitatively identical. Relative percentages of the fractions resulted in CV ranges from 1.58 – 5.34% ### b. Linearity/assay reportable range: An abnormal patient sample was serially diluted to 1:22. Hemoglobin A1 and A2 were evaluated assess the lowest concentration that can be detected. Linearity ranges for HbA2 in hemolyzed sample for both size kits were: Microtech 672 PC .44-2.1 g/L Microtech 548 ISO 0.5-1.7 g/L ### c. Traceability (controls, calibrators, or method): Not applicable ### d. Detection limit: The limit of detection was determined to be concentrations greater than 0.40 mg/ml with the exception of Hemoglobin S (HbS). The system will detect HbS concentrations greater than 0.47 mg/ml. ### e. Analytical specificity: Ineffective centrifugation and/or RBC washing with saline solution will not yield a clear hemolysate. This may result in the presence of a red line at the application point, which is indicative of a poor quality electrophoretic pattern. Frozen samples may produce an artifact band near HbA1. The point of sample application should not be visible in the pattern. If the application point is present as a line strongly stained, expect a smeared pattern, with poor resolution and substandard focusing of the bands. Deviation from the recommended test procedure may affect the results. ### f. Assay cut-off: Not applicable {5} Page 5 of 5 2. Comparison studies: a. Method comparison with predicate device: A comparison study was performed using 93 samples from both normal and suspected pathological patients. The samples were run using both sample size kits. This study resulted in correlation coefficients greater than 0.99 and yielded a 100% agreement to the reference method for observed bands. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a and b are not applicable): Not Applicable 4. Clinical cut-off: NA 5. Expected values/Reference range: Reference Ranges HbA1 96-98 HbA2 1.5-3.5 HbF ≤2 These ranges are based on ranges published in Clinical Laboratory Methods, John D. Bauer, MD (9th Edition, the C.V. Mosby Company (1982)). M. Conclusion: Based on the review of the information provided, the InterLab Alkaline Electrophoresis Test System appears to be substantially equivalent (SE) to devices regulated under 21 CFR 864.7440, electrophoretic hemoglobin analysis system, product code JBD.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...