← Product Code [IMI](/productcode/IMI) · K032793

# US-100 PORTABLE ULTRASOUND THERAPY UNIT (K032793)

_Ito Co., Ltd. · IMI · Oct 9, 2003 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K032793

## Device Facts

- **Applicant:** Ito Co., Ltd.
- **Product Code:** [IMI](/productcode/IMI.md)
- **Decision Date:** Oct 9, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5300
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The US-100 is indicated for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions, including relief of pain, muscle spasms, and joint contractures.

## Device Story

US-100 Ultrasound Therapy Unit provides therapeutic deep heat to body tissues. Device consists of main unit and two applicators; operates in continuous or pulsed modes at 0.8 MHz with 8 W output power. User interface includes LCD screen for parameter selection and status messaging. Operated by healthcare professionals in clinical settings. Output heat facilitates treatment of pain, muscle spasms, and joint contractures, potentially improving patient mobility and comfort.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical safety, electromagnetic emissions, and performance standards per 21 CFR 1050.

## Technological Characteristics

Ultrasonic diathermy device; 0.8 MHz operating frequency; 8 W output power; continuous and pulsed modes; LCD user interface; two applicator options. Complies with 21 CFR 1050 performance standards, electrical safety, and EMC requirements.

## Regulatory Identification

An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.

## Predicate Devices

- Sonicator Plus 992 ([K984142](/device/K984142.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT - 9 2003

## 510(k) Summary

# US-100 Ultrasound Therapy Unit

Common/Classification Name: Ultrasonic Diathermy Device, 21 CFR 890.5300

Ito Company, Ltd. 3-3-3 Toyotama-Minami Nerima-Ku Tokyo 176-8605, Japan

Contact: K. Sunayama, Prepared: September 5, 2003

#### LEGALLY MARKETED PREDICATE DEVICES A.

The US-100 Ultrasound Therapy Unit is substantially equivalent to the ultrasound section of the Sonicator Plus 992 (K984142) by Mettler.

#### DEVICE DESCRIPTION B.

The US-100 Ultrasound Therapy Unit consists of a main unit and two applicators. The device may be operated in continuous or pulsed modes. The US-100 has an output power of 8 W and operates at 0.8 MHz. It is supplied with one applicator, and another smaller applicator is available as an option.

The device has an LCD screen that serves as the interface with the user to specify options, provide messages, and display parameters.

#### ். INTENDED USE

The US-100 is indicated for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions, including relief of pain, muscle spasms, and joint contractures.

#### SUBSTANTIAL EQUIVALENCE SUMMARY D.

The US-100 Ultrasound Therapy Unit is a medical device, and it has very similar indications for use as the legally marketed predicate device. The differences do not change the intended therapeutic effect. The US-100 Ultrasound Therapy Unit has the same technological characteristics as the predicate device. For a few characteristics, performance data were provided, which demonstrate equivalence. This premarket notification 

{1}------------------------------------------------

to assure substantial equivalence except for a few of the characteristics where performance testing was carried out (e.g., electrical safety, EMC).

#### E. TECHNOLOGICAL CHARACTERISTICS

The technological characteristics are the same as those of the predicate device and other ultrasound diathermy devices.

#### F. TESTING

Performance testing carried out by Ito Co, Ltd. on the US-100 addressed the following issues:

- Electrical Safety; (1)
- (2) Electromagnetic Emissions;
- Various Safety and Integrity Issues, and (3)
- Performance Standard of 21 CFR 1050. (4)

#### CONCLUSIONS G.

This pre-market notification has demonstrated Substantial Equivalence as defined and understood in Sections 513(f)(1) and 513(i)(1) of the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.

### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT - 9 2003

Ito Company, Ltd. C/o T. Whit Athey, Ph.D. Senior Consultant The Health Policy Resources Group, LLC 2305 Gold Mine Road, Suite 200 Brookeville, MD 20833-2233

Re: K032793

Trade/Device Name: US-100 Portable Ultrasound Therapy Unit Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic diathermy Regulatory Class: II Product Code: IMI Dated: September 8, 2003 Received: September 9, 2003

Dear Dr. Athey,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Suite 200

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-1308. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

Mark M. Milken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: US-100 Ultrasound Therapy Unit

Indications For Use:

The US-100 is indicated for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions, including relief of pain, muscle spasms, and joint contractures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use -------

Merle N. Millenson
(Division Sign-Off)

Division of General. Restorative and Neurological Devices

510(k) Number K032793

000024

---

**Source:** [https://fda.innolitics.com/device/K032793](https://fda.innolitics.com/device/K032793)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
