← Product Code [DQO](/productcode/DQO) · K032624

# MICRUS MICROCATHETER (K032624)

_Micrus Corp. · DQO · Sep 10, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K032624

## Device Facts

- **Applicant:** Micrus Corp.
- **Product Code:** [DQO](/productcode/DQO.md)
- **Decision Date:** Sep 10, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Micrus Microcatheter is intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents such as occlusion coils, into peripheral, coronary, and neuro vasculature.

## Device Story

Micrus Microcatheter is an intravascular diagnostic/therapeutic catheter. Components include flexible shaft with Teflon liner, stainless steel wire reinforcement, Pebax outer jacket, atraumatic distal tip with two radiopaque marker bands, and standard luer hub. Distal 100cm features hydrophilic coating. Device used by clinicians to navigate peripheral, coronary, and neuro vasculature to deliver contrast media or embolic coils. Radiopaque markers facilitate visualization of tip position during placement. Device functions as a conduit for interventional procedures.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Flexible shaft with Teflon liner; stainless steel wire reinforcement; Pebax outer jacket; hydrophilic coating on distal 100cm; two radiopaque marker bands; standard luer hub. Intravascular diagnostic catheter (Class II, 870.1200, DQO).

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- Tracker Excel 14 Microcatheter ([K960806](/device/K960806.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 1 0 2003

K032624
p 1/2

## 510(k) Safety and Effectiveness Summary l.

#### Contact Information A.

Margaret Webber Director, Regulatory and Clinical Affairs Micrus Corporation 610 Palomar Avenue Sunnyvale, CA 94085

#### B. Device Name

Micrus Microcatheter Device: Catheter, Intravascular, Diagnostic Regulation Number: 870.1200 Product Code: DQO Device Class: 2

#### C. Predicate Device(s)

| Number  | Description                    | Clearance Date |
|---------|--------------------------------|----------------|
| K960806 | Tracker Excel 14 Microcatheter | 05/02/96       |

#### D. Device Description

The Micrus Microcatheter consists of 4 major components:

- A flexible shaft with a lubricious liner extruded from Teflon. The ■ liner is reinforced with a metal wire made from stainless steel. The outer shaft jacket is made from Pebax,
- 트 An atraumatic distal flexible tip containing two radiopaque marker bands. The marker bands are useful in tracking catheter tip position and during placement of detachable embolic coils.
- A standard luer hub insert, which is molded onto the flexible shaft. ■
- 보 A hydrophilic coating, which covers the distal 100 centimeters of the Microcatheter.

#### E. Intended Use

The Micrus Microcatheter is intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents such as occlusion coils, into peripheral, coronary, and neuro vasculature.

#### Intended Use Predicate Device (per products' Instructions for Use) F.

"The Target Therapeutics' Microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents such as occlusion coils, into peripheral, coronary, and neuro vasculature."

3

{1}------------------------------------------------

### G. 510(k) Summary of Safety and Efficacy

Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-clinical testing performed by Micrus Corporation, it is concluded that the Micrus Microcatheter is substantially equivalent to the Boston Scientific/Target Therapeutics Microcatheter in safety and effectiveness.

4

M. Webb

Margaret Webber Director, Regulatory and Clinical Affairs Micrus Corporation June 27, 2003

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three overlapping human profiles, which are meant to represent the department's mission of protecting the health of all Americans and providing essential human services.

SEP 1 0 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Micrus Corporation c/o Mr. Morten Christensen Underwriters Laboratories, Inc. Office Coordinator 1655 Scott Boulevard Santa Clara, CA 95050-4169

> K032624 Micrus Microcatheter Regulation Number: 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II (two) Product Code: DQO Dated: August 25, 2003 Received: August 26, 2003

Dear Mr. Christensen:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

# Page 2 - Mr. Morten Christensen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices

Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Device Name:

510(k) Number (if known):

Micrus Microcatheter

Indications for Use:

The MicroCatheter is intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents such as occlusion coils, into peripheral, coronary, and neuro vasculature.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Over the Counter Use: or Prescription Use: (Per 21 CFR 801.109)

Veleton

andigation)
 of Cardiovascular Devices 510(k) Number_

---

**Source:** [https://fda.innolitics.com/device/K032624](https://fda.innolitics.com/device/K032624)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
