ENCLOSE II ANASTOMOSIS ASSIST DEVICE

K032589 · Novare Surgical Systems, Inc. · DXC · Sep 5, 2003 · Cardiovascular

Device Facts

Record IDK032589
Device NameENCLOSE II ANASTOMOSIS ASSIST DEVICE
ApplicantNovare Surgical Systems, Inc.
Product CodeDXC · Cardiovascular
Decision DateSep 5, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ENclose™ II Anastomosis Assist Device is intended for use by cardiac surgeons in place of partial occluding clamps during coronary artery bypass grafting (CABG) procedures requiring one to three proximal anastomoses in ascending aortas free of atheromatous disease.

Device Story

The ENclose™ II Anastomosis Assist Device is a vascular clamp used intraoperatively by cardiac surgeons during coronary artery bypass grafting (CABG). It serves as an alternative to traditional partial occluding clamps for creating one to three proximal anastomosis sites in the ascending aorta. The device is constructed from medical-grade plastic, stainless steel, and silicone. It features an expandable lower jaw and is supplied with an optional fiber-reinforced polymer hex driver to assist in deployment. The device is provided sterile for single-patient use only. By replacing traditional clamps, it aims to facilitate the anastomosis process in patients with aortas free of atheromatous disease.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on physical and functional attributes, including material certification and design comparison to predicate devices.

Technological Characteristics

Materials: medical grade plastic, stainless steel, silicone. Energy source: manual/mechanical. Form factor: vascular clamp with expandable lower jaw. Connectivity: none. Sterilization: radiation (e-beam). Software: none.

Indications for Use

Indicated for cardiac surgeons performing coronary artery bypass grafting (CABG) requiring one to three proximal anastomoses in ascending aortas free of atheromatous disease.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters, which appear to be a combination of letters and numbers. The sequence reads 'KO32589'. The characters are written in a cursive style, with some connections between the letters and numbers. The writing is clear and legible, with a consistent stroke weight. | Summary of Safety Information<br>Premarket Notification, Section 510(k) | | NOVARE SURGICAL SYSTEMS, INC.<br>AUGUST 16, 2003 | |-------------------------------------------------------------------------|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------| | Regulatory Authority: | Safe Medical Devices Act of 1990, 21 CFR 807.92 | | | Device Name: | ENclose™ II Anastomosis Assist Device | | | Common Name(s): | Vascular Clamp | | | Classification Name(s): | Vascular Clamp | | | Manufacturer: | | | | Name: | Novare Surgical Systems, Inc. | | | Reg. Number: | 2954739 | | | Address: | 10231 Bubb Road<br>Cupertino, CA 95014 | | | Classification(s): | | | | | | Sec. 870.4450 Vascular clamp. (a) Identification. A vascular clamp is a surgical instrument used to occlude a blood | Class II for the requested indications Device Class: Cardiovascular Devices Panel Classification Panel: Product Code(s): DXC vessel temporarily. (b) Classification. Class II (performance standards). # Equivalent Predicate Device: The ENclose™ Anastomosis Assist Device is substantially equivalent to the following legally marketed vascular clamp(s): Includer™ Vascular Clamp - Novare Surgical Systems, Inc., K010694 Includer™ (ENclose™) Vascular Clamp - Novare Surgical Systems, Inc., K023682 Sec. 870.4450 Vascular clamp. (a) Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily. (b) Classification. Class II (performance standards). Device Class: Class II Product Code(s): DXC The new anastomosis device is essentially identical to the previous device. Equivalency can be seen with respect to place of manufacture, physical appearance, functional testification, surgical technique and indications for use. A feature comparison chart follows this section. | FEATURE | ENclose™ II Anastomosis Assist Device | Includer™ / ENclose™ | SE? | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----| | Indications for<br>Use: | The ENclose™ II Anastomosis Assist Device is intended<br>for use by cardiac surgeons in place of partial occluding<br>clamps during coronary artery bypass grafting (CABG)<br>procedures requiring one to three proximal anastomoses<br>in ascending aortas free of atheromatous disease. | Identical | YES | | Materials: | Medical grade plastic, stainless steel and silicone | Identical | YES | | Design: | Clamp | Identical | YES | | Sterilization: | Radiation (e-beam) | Identical | YES | | Manufacturer: | Novare Surgical Systems, Inc. | Identical | YES | | Product Code: | DXC | Identical | YES | | K - Number: | Pending | K010694, K023682 | YES | {1}------------------------------------------------ The proposed changes do not represent a significant or deleterious change to the device. Substantial equivalence is based on all physical and clinical attributes of the device being indistinguishable from the cleared device, including, place of manufacture, physical appearance, though slightly different in color and size, has not affected the performance requirements, functional elements, material certification, surgical technique nor indications for use. ## Device Description: General description of the change(s). The proposed changes are limited to manufacturing/processing changes, a change in the colorant used in the existing plastic of the device, a change in the shape of the expandable region of the lower jaw and the addition of a fiber reinforced polymer hex driver to assist with deployment of the device if desired. The current FDA cleared indications granted for single and up to three anastomosis procedures remains unchanged. The modified ENclose™ II Anastomosis Assist Device is an intraoperative device used to assist in the creation of one to three proximal anastomosis sites in the execution of coronary artery bypass grafting (CABG) procedures. Materials for the new device are identical to the previous device with the exception of the colorant. #### Company Contact: Mr. Kerry Pope Novare Surgical Systems, Inc. 10231 Bubb Road Cupertino, CA 95014 408.873.3161 ## Submission Correspondent: Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax #### Special Controls: FDA published special controls. The following special controls are believed to apply to the marketing of class II devices: - Compliance with material standards, (i) - Compliance with biocompatability standard, and (ii) - (iii) Compliance with specified labeling requirements. Novare Surgical Systems, Inc. also meets appropriate general controls authorized under Sections 501, 510, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act. #### Sterility Processing: The ENclose™ II Anastomosis Assist Device is provided sterile and is designed strictly for single patient use. The device may not be cleaned, reprocessed or reused under any circumstances. The product must be handled, stored and placed into use in such a way that it is protected from inadvertent damage or contamination. When used, the product must be placed into use following accepted surgical sterile technique. {2}------------------------------------------------ #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP - 5 2003 Novare Surgical Systems, Inc. c/o David W. Schlerf Buckman Company, Inc. 200 Gregory Lane Suite C-100 Pleasant Hill, CA 94523 Re: K032589 Enclose II Anastomosis Assist Device Regulation Number: 21 CFR 870.4450 Regulation Name: Regulatory Class: Class II Product Code: DXC Dated: August 16, 2003 Received: August 22, 2003 Dear Mr. Schlerf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. David Schlerf Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html Sincerely yours, K. Odel Tule Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ **510(k) Number**: K032589 ENclose™ II Anastomosis Assist Device Device Name(s): # Indications for Use Statement(s): The ENclose™ II Anastomosis Assist Device is intended for use by cardiac surgeons in place of partial occluding clamps during coronary artery bypass grafting (CABG) procedures requiring one to three proximal anastomoses in ascending aortas free of atheromatous disease. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) V. Deaton sion of Cardiovascular Devices **510(k) Number** K032589 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional format 1-2-96) ENCLOSESPECIAL2.DOC
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%