K032585 · Ron-Tech Medical , Ltd. · HDC · Jan 9, 2004 · Obstetrics/Gynecology
Device Facts
Record ID
K032585
Device Name
TRANSVAGINAL ULTRASOUND PROBE HOLDER SYSTEM
Applicant
Ron-Tech Medical , Ltd.
Product Code
HDC · Obstetrics/Gynecology
Decision Date
Jan 9, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The SafeTChoice™ Transvaginal Ultrasound Probe Holder System is an accessory intended for use in conjunction with a transvaginal ultrasound transducer. The SafeTChoice™ Transvaginal Ultrasound Probe Holder System is intended to seize and hold the cervix and to hold and secure an ultrasound transducer in place, while performing sonographic procedures.
Device Story
System acts as mechanical accessory for transvaginal ultrasound transducers; consists of reusable uterine tenaculum with adaptor bar, sterile disposable adaptor, and disposable probe connector. Device secures transducer to tenaculum, allowing physician to manipulate ultrasound probe with one hand while performing procedure. Used in clinical settings by physicians during sonographic procedures. Benefits include improved procedural efficiency and ease of use by freeing one hand for clinical tasks.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Mechanical accessory system; includes reusable uterine tenaculum, disposable adaptor, and disposable probe connector. No electronic components, software, or energy sources.
Indications for Use
Indicated for use as an accessory to a transvaginal ultrasound transducer to seize and hold the cervix and secure the transducer in place during sonographic procedures.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
KD32585 p. 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Ron-Tech Medical LTD. The words "Ron-Tech" are in large, bold, black letters, with a curved line above them. Below "Ron-Tech" is a black rectangle with the word "Medical" in white letters, and the letters "LTD" in the upper right corner.
9 2004 JAN
### 510(k) Summary:
SafeTChoice™ Transvaginal Ultrasound Probe Holder System
Date submitted: August 15, 2003
Company Name:
Ron-Tech Medical Ltd.
Contact Person: Doron Kilchevsky CEO
> Telephone: +972-9-957-8803 Fax: +972-9-957-8759
Trade Name: SafeTChoice™ Transvaginal Ultrasound Probe Holder System
Classification name: Accessory to an ultrasound transducer
Classification: HDC/ITX
### Predicate Device:
Transvaginal Ultrasound Probe Holder device, Ron-Tech Medical Ltd. Israel, cleared under 510(k) no. K992071
### Description of the device:
The Safe TChoice™ Transvaginal Ultrasound Probe Holder System is an accessory used in conjunction with a transvaginal ultrasound transducer. The device is intended to hold and secure the ultrasound transducer in place on a uterine tenaculum, while performing the sonographic procedure. The configuration of the device allows the physician to manipulate the ultrasound transducer with one hand, thus freeing the other hand to perform the required procedure.
{1}------------------------------------------------
K032585 p. 2 of 2
Image /page/1/Picture/1 description: The image shows the logo for Ron-Tech Medical LTD. The words "Ron-Tech" are in large, bold, black font, with a curved line above them. Below that, the word "Medical" is in a smaller font, and to the right of that, the letters "LTD" are stacked vertically in a very small font.
The SafeTChoice™ Transvaginal Ultrasound Probe Holder System consists of a reusable uterine tenaculum with an adaptor bar, a sterile, disposable adaptor and a disposable probe connector.
#### Indications for Use:
The Safe TChoice™ Transvaginal Ultrasound Probe Holder System is an accessory intended for use in conjunction with a transvaginal ultrasound transducer. The SafeTChoice™ Transvaginal Ultrasound Probe Holder System is intended to seize and hold the cervix and to hold and secure an ultrasound transducer in place, while performing sonographic procedures.
#### Substantial Equivalence:
The SafeTChoice™ Transvaginal Ultrasound Probe Holder System has the same intended use and the same principle of operation as the Transvaginal Ultrasound Probe Holder device, cleared under 510(k) no. K9992071.
#### Conclusion: **
The evaluation of the Safe TChoice™ Transvaginal Ultrasound Probe Holder System does not raise any additional concerns regarding safety and effectivity and may therefore be considered substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circle's perimeter.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### 9 2004 JAN
Ms. Doron Kilchevsky CEO Ron-Tech Medical, Ltd. 26 Hasivim St. Kiryat Matalon Petach-Tikva 49170 ISRAEL
K032585 Re:
Trade/Device Name: Transvaginal Ultrasound Probe Holder System Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Product Code: 85 HDC Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Product Code: 90 ITX Regulatory Class: II Dated: November 26, 2003 Received: December 12, 2003
Dear Ms. Kilchevsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have teviewed your becamed the device is substantially equivalent (for the indications felerenced above and navo uctemined ly marketed predicate devices marketed in interstate for use stated in the electore) to regard to tegars and the Medical Device Amendments, or to commerce prior to may 20, 1978, its ecordance with the provisions of the Federal Food, Ung, devices that nave occh reclassified in assess approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicierore, market are as act include requirements for annual registration, listing of general controls provisions practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is classified (500 acres) and secure) and or regulations affecting your device can be it may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I·DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a determed by other Federal agencies. You must of any I cicrar studies and rogains.
comply with all the Act's requirements, including, but not limited to: registration and listing comply with an the recessfort (21 CFR Part 801); good manufacturing practice requirements as sct (21 CFR Part 807), labornig (21 OFF Part 820); and if applicable, the electronic forul in the quality 375tems (QD) res (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I this letter will and in you ve organ finding of substantial equivalence of your device to a legally premairted predicated ... - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you dosite spoolite Dompliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 10 questions of (301) 594-4639. Also, please note the regulation entitled, "Misbranding Other or Complanes as ( . ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . by recented to promative is not the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
· Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
| 510(k) Number (if known): | K032585 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | SafeTChoice™ Transvaginal Ultrasound Probe Holder System |
| Indications for Use: | The SafeTChoice™ Transvaginal Ultrasound Probe Holder System is an accessory intended for use in conjunction with a transvaginal ultrasound transducer. The SafeTChoice™ Transvaginal Ultrasound Probe Holder System is intended to seize and hold the cervix and to hold and secure an ultrasound transducer in place, while performing sonographic procedures. |
# (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE):
L Prescription Use (Per 21 CFR 801.109) Over the Counter Use_
Nancy C Snogdon
(Division Sign-Off)
11-2
Division of Reproductive, Abdominal, and Radiological Devices
OR
Radiological Devices
510(k) Number K032585
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.