QUANTEX D-DIMER

K032419 · Instrumentation Laboratory CO · DAP · Oct 24, 2003 · Hematology

Device Facts

Record IDK032419
Device NameQUANTEX D-DIMER
ApplicantInstrumentation Laboratory CO
Product CodeDAP · Hematology
Decision DateOct 24, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7320
Device ClassClass 2

Indications for Use

The Instrumentation Laboratory Co. (IL) quantex D-DIMER is a latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on automated clinical chemistry analyzers.

Device Story

Quantex D-DIMER is a latex-enhanced turbidimetric immunoassay; utilizes polystyrene latex particles coated with monoclonal antibody (MA-8D3) specific for D-Dimer domain in fibrin soluble derivatives. Input: human citrated plasma samples. Operation: mixing plasma with latex reagent and phosphate buffer causes particle agglutination; degree of agglutination is directly proportional to D-Dimer concentration. Output: quantitative D-Dimer concentration measured via decrease in transmitted light. Used in clinical laboratory settings on automated clinical chemistry analyzers (e.g., ILab 600). Healthcare providers use results to aid in clinical decision-making regarding fibrin degradation products. Benefits include automated, quantitative assessment of D-Dimer levels.

Clinical Evidence

Bench testing only. Precision evaluated per NCCLS EP5-T2 (N=60, Total CV 2.83-5.75%). Linearity confirmed over 154–3212 ng/mL (R2=0.9997). Analytical specificity showed no interference from heparin, bilirubin, hemoglobin, or triglycerides. Method comparison (N=137) against predicate yielded slope 1.059, r=0.987. Field study (N=112) across ER, DIC, DVT, and PE patient samples showed correlation (slope 0.752, r=0.984). Normal range established using 125 healthy donors (mean 92 ng/mL, UL 198 ng/mL).

Technological Characteristics

Latex-enhanced turbidimetric immunoassay. Reagents: polystyrene latex particles coated with monoclonal antibody (MA-8D3), phosphate buffer. Calibrators/controls: partially purified D-Dimer from human fibrin digested with human plasmin. Preservative: 0.02% Bronidox. Measurement: turbidimetric (decrease in transmitted light). System: automated clinical chemistry analyzer (ILab 600). Storage: 2-8°C.

Indications for Use

Indicated for the quantitative determination of D-Dimer in human citrated plasma using automated clinical chemistry analyzers.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: #K032419 B. Analyte: Fibrin degradation product (D-Dimer) C. Type of Test: Quantitative D. Applicant: Instrumentation Laboratory Co. E. Proprietary and Established Names: quantex D-DIMER; Fibrin Degradation Products Assay (D-Dimer) F. Regulatory Information: 1. Regulation section: CFR 864.7320 2. Classification: Class II 3. Product Code: DAP, GHH 4. Panel: Hematology (81) G. Intended Use: 1. Intended use(s): The Instrumentation Laboratory Co. (IL) quantex D-DIMER is a latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on automated clinical chemistry analyzers. 2. Indication(s) for use: Quantex D-DIMER is a latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on automated clinical chemistry analyzers. 3. Special condition for use statement(s): N/A 4. Special instrument Requirements: The automated ILab 600 Clinical Chemistry System (#K980757) H. Device Description: The IL quantex D-DIMER Kit consists of Latex Reagent, Phosphate Buffer, Calibrator and Controls I/II. The Latex Reagent is a suspension of polystyrene latex particles of uniform size, coated with a monoclonal antibody highly specific for the D-Dimer domain included in fibrin soluble derivatives. The Calibrator and Controls are solutions of partially purified D-Dimer from human fibrin digested with human plasmin. The quantex D-DIMER turbidimetric immunoassay measures the decrease in transmitted light caused by aggregates formed by latex particle agglutination in the presence of D-Dimer. {2} Page 2 of 6 I. Substantial Equivalence Information: 1. Predicate device name(s): Instrumentation Laboratory Co. IL Test D-Dimer 2. Predicate K number(s): #K972696 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Assay type | Latex enhanced immuno-assay | SAME | | Monoclonal anti-body (MAB) | MA-8D3 | SAME | | Test principle | Turbidimetric agglutination | SAME | | Buffer | Phosphate buffer | SAME | | Calibrator/Controls | Lyophilized solution of partially purified D-Dimer, from human fibrin digested with human plasmin. | SAME | | Storage | 2 - 8° C. | SAME | | Differences | | | | Item | Device | Predicate | | Instrumentation | Automated clinical chemistry analyzers | Automated coagulation analyzers | | Preservatives for Calibrator and Latex Reagent | 0.02% Bronidox | < 0.1% Sodium azide | | Linear range | 150 – 3200 ng/mL | 200 – 1050 ng/mL | | Calibrator target value | 3200 ng/mL | 1000 ng/mL | {3} Page 3 of 6 J. Standard/Guidance Document Referenced (if applicable): N/A K. Test Principle: Polystyrene latex particles are coated with a monoclonal antibody highly specific for the D-Dimer domain included in fibrin soluble derivatives. When plasma containing D-Dimer is mixed with the Latex Reagent and the Reaction Buffer included in the quantex D-DIMER Kit, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of D-Dimer in the sample. The concentration is determined by measuring the decrease in light transmitted when the particles agglutinate in the presence of D-Dimer. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Controls I/II (N=60) were run on the ILab 600 Clinical Chemistry System, according to NCCLS EP5-T2 document. Results are as follows: | | Level I | Level II | | --- | --- | --- | | Mean | 301.6 ng/mL | 636.8 ng/mL | | Within-run | 3.99%CV | 2.17%CV | | Between-run | 2.78%CV | 0.00%CV | | Total | 5.75%CV | 2.83%CV | b. Linearity/assay reportable range: Calibrator/saline dilutions (12 levels) were tested in triplicate, over (2) separate runs. They generated values over a range of 154 – 3212 ng/mL. The regression study of expected vs reported values yielded this equation: $y = 1.01x + 0.1663$ ; $R2 = 0.9997$ c. Traceability (controls, calibrators, or method): d. Detection limit: 77 ng/mL A saline study was performed, using (3) assay reagent lots (20 replicates). The mean +3SD (2 calibrations/lot) was calculated. The maximum value obtained was determined to be the detection limit. e. Analytical specificity: Interference testing demonstrated no significant interference as follows: - Unfractionated heparin (UFH) - &lt; 1.5 IU/mL - Low molecular weight heparin (LMWH) - &lt; 1.5 IU/mL - Bilirubin - &lt; 18mg/dL - Hemoglobin - &lt; 500 mg/dL - Triglycerides - &lt; 1280 mg/dL {4} Page 4 of 6 f. Assay cut-off: 198 ng/mL The upper limit (UL) of normal (mean of 92 ng/mL + 2SD of 53 ng/mL) was determined from a study performed on citrated plasmas from normal blood bank donors (N=125). 2. Comparison studies: a. Method comparison with predicate device: An in-house study was performed on duplicate citrated plasma samples (N=137), obtained from the emergency room, Hospital de San Paul in Spain. The samples ranged in value 17 – 24149 ng/mL. The quantex D-Dimer/ILab 600 combination was compared to the IL Test D-Dimer/ACL Futura (#K951891) combination. Regression results on the mean of duplicates were: Slope – 1.059, Intercept – 50.2, r – 0.987 Singlet results were statistically similar: Slope – 1.053, Intercept – 51.6, r – 0.979 b. Matrix comparison: N/A 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a and b are not applicable): A field site study was performed at Hospital St. Joan de Reus, also in Spain, on patient plasma samples (N=112), comparing the quantex D-Dimer to the IL Test D-Dimer on the ACL 7000 (#K961991). Single determinations were made on these samples: ER (N=44); DIC (N=30); DVT (N=20); PE (N=18). The study yielded these regression statistics: Slope – 0.752, Intercept – 52.5, r – 0.984 4. Clinical cut-off: N/A 5. Expected values/Reference range: {5} Page 5 of 6 Mean = 92 ng/mL The normal range study was performed on citrated plasma from normal blood bank donors, males and females (N = 125), ranging in age 18-66 years. Test results yielded a mean of 92 ng/ml; and a UL of normal of 198 ng/mL (mean + 2 SD). Five outliers were donor samples that tested above the UL; and (3) donor samples had insufficient quantities and could not be tested. M. Conclusion: The company provided performance data that was within acceptable analytical limits for this type of device. Based upon the data presented, this device may be found substantially equivalent to a legally marketed device. {6} Page 6 of 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...