← Product Code [LPL](/productcode/LPL) · K032340

# VISTAKON (GALYFILCON A) SOFT CONTACT LENS (K032340)

_Johnson & Johnson Vision Care, Inc. · LPL · Oct 16, 2003 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K032340

## Device Facts

- **Applicant:** Johnson & Johnson Vision Care, Inc.
- **Product Code:** [LPL](/productcode/LPL.md)
- **Decision Date:** Oct 16, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The VISTAKON® (galyfilcon A) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. The VISTAKON® (galyfilcon A) Multifocal Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or not-aphakic persons with non-diseased eyes who may have 0.75 D of astigmatism or less. The VISTAKON® (galyfilcon A) Toric Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or not-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less. The VISTAKON® (galyfilcon A) Toric-Multifocal Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or not-aphakic persons with non-diseased who may have 10.00 D of astigmatism or less. VISTAKON® (galyfilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. Eye Care Practitioners may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement (see "Wearing Schedule"). When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only.

## Device Story

VISTAKON (galyfilcon A) contact lens; hemispherical or hemitoric shell; copolymer of 2-hydroxyethyl methacrylate, (3-methacryloxy-2-hydroxypropyloxy) propylbis(trimethylsiloxy)methylsilane, N,N-dimethylacrylamide, polyvinylpyrrolidone, mPDMS-monomethacrylate, and 2-propenoic acid, 2-methyl-, 1,2-ethandiyl ester. Available as spherical, multifocal, toric, and toric-multifocal designs; includes optional blue visibility tint and benzotriazole UV-absorbing monomer. Supplied sterile in buffered saline. Prescribed by eye care practitioners for daily wear; single-use or frequent/planned replacement. Benefits include vision correction and UV radiation protection. Clinical decision-making based on practitioner assessment of refractive error and ocular health.

## Clinical Evidence

Three-month clinical study; at least 50 myopic patients; 2:1 ratio of subject device to predicate. Evaluated adverse reactions, keratometry, discontinuations, and unscheduled replacements. Results demonstrated similar performance in vision, health, comfort, and fit compared to concurrent controls.

## Technological Characteristics

Material: galyfilcon A (53% polymer, 47% water). Group I (low water, nonionic). UV blocker: benzotriazole monomer (transmittance <1% UVB, <10% UVA). Form factor: hemispherical/hemitoric shell. Sterilization: sterile, buffered saline solution. Connectivity: N/A.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- Focus® NIGHT AND DAY™ (lotrafilcon A)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) PREMARKET NOTIFICATION
VISTAKON®® (galyfilcon A) Contact Lens

# K 032340

# Summary of Safety and Effectiveness

| Submitter<br>Information        | Company:                                                                                                                                                                                              | VISTAKON®<br>Division of Johnson & Johnson Vision Care,<br>Inc.<br>7500 Centurion Parkway<br>Suite 100<br>Jacksonville, FL 32256              |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
|                                 | Contact Person:                                                                                                                                                                                       | Jacqueline Zimovan<br>Project Leader<br>Regulatory & Clinical Affairs                                                                         |
|                                 | Telephone:<br>FAX:                                                                                                                                                                                    | 904-443-1402<br>904-443-1424                                                                                                                  |
|                                 | Date Prepared:                                                                                                                                                                                        | July 28, 2003                                                                                                                                 |
| Identification of<br>the Device | Common Name:<br>Device Name:<br>Classification Name:<br>Device Classification:                                                                                                                        | Soft Contact Lens<br>VISTAKON® (galyfilcon A) Contact Lens<br>Soft Hydrophilic Contact Lens, Daily Wear<br>Class II, 21 CFR 886.5925 (b) (1). |
| Predicate<br>Device             | The predicate device was selected to address both intended use (daily<br>wear) and material type (FDA Group I; low water, nonionic polymer):<br>• Focus® NIGHT AND DAY™ (lotrafilcon A), FDA Group I. |                                                                                                                                               |
|                                 |                                                                                                                                                                                                       | Continued on next page                                                                                                                        |

Page 11

{1}------------------------------------------------

| Description of<br>Device | • The VISTAKON® (galyfilcon A) Contact Lens Clear and<br>Visibility Tint with UV blocker is available as a spherical lens, a<br>multifocal lens, a toric lens and a toric-multifocal lens. |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

- The lens material, galyfilcon A, is a copolymer of 2-hydroxyethyl . methacrylate and (3-methacryloxy-2-hydroxypropyloxy) propylbis(trimethylsiloxy)methylsilane and N.Ndimethylacrylamide, and polyvinylpyrrolidone and mPDMSmonomethacrylate (n=11), and 2-propenoic acid, 2-methyl-, 1,2ethandiyl ester.
- The VISTAKON® (galyfilcon A) Contact Lens may be tinted blue . using Reactive Blue Dye #4 to make the lens more visible for handling.
- · A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 1% in the UVB range of 280 – 315nm and less than 10% in the UVA range of 316 - 380nm.
- · The VISTAKON® (galyfilcon A) Contact Lens is a hemispherical or hemitoric shell.
- . The lens is supplied in a sterile state, packaged in a buffered saline solution with up to 0.01% methyl ether cellulose.
- . The composition of the lens is 53% galyfilcon A and 47% water by weight when hydrated and stored in the buffered saline solution.

{2}------------------------------------------------

#### Indications for Use

| Lens Design         | Indication                                                                                                                                                                                                                                                    |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Spherical           | The VISTAKON® (galyfilcon A) Contact Lens is indicated<br>for daily wear for the correction of refractive ametropia<br>(myopia and hyperopia) in aphakic or not-aphakic persons<br>with non-diseased eyes who may have 1.00D of<br>astigmatism or less.       |
| Multifocal          | The VISTAKON® (galyfilcon A) Multifocal Contact Lens<br>is indicated for daily wear for the correction of distance and<br>near vision in presbyopic, aphakic or not-aphakic persons<br>with non-diseased eyes who may have 0.75D of<br>astigmatism or less.   |
| Toric               | The VISTAKON® (galyfilcon A) Toric Contact Lens is<br>indicated for daily wear for the correction of visual acuity in<br>aphakic or not-aphakic persons with non-diseased eyes that<br>are hyperopic or myopic and may have 10.00D of<br>astigmatism or less. |
| Toric<br>Multifocal | The VISTAKON® (galyfilcon A) Toric-Multifocal Contact<br>Lens is indicated for daily wear for the correction of<br>distance and near vision in presbyopic aphakic or not-<br>aphakic persons with non-diseased who may have 10.00D<br>of astigmatism or less. |

- · VISTAKON® (galyfilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
- · Eye care practitioners may prescribe the lens for single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement, the lens may be disinfected using a chemical disinfection system only.

{3}------------------------------------------------

Technological Characteristics The characteristics of the VISTAKON® (galyfilcon A) Contact Lens are compared to the characteristics of the predicate device, Focus® NIGHT & DAY™ (lotrafilcon A) Contact Lens, in the following tables.

| Material Comparison |                                                          |                                             |
|---------------------|----------------------------------------------------------|---------------------------------------------|
|                     | Focus® NIGHT AND<br>DAY™ (lotrafilcon A)<br>Contact Lens | VISTAKON®<br>(galyfilcon A)<br>Contact Lens |
| Material            | lotrafilcon A                                            | galyfilcon A                                |
| Type                | Group I                                                  | Group I                                     |

| Parameter Comparison          |                                                           |                                             |
|-------------------------------|-----------------------------------------------------------|---------------------------------------------|
|                               | Focus® NIGHT AND<br>DAYTM (lotrafilcon A)<br>Contact Lens | VISTAKON®<br>(galyfilcon A)<br>Contact Lens |
| Characteristics<br>Comparison | Measured Labeled                                          | Measured Labeled                            |
| Water Content, %              | 23 24                                                     | 48 47                                       |
| Refractive Index @ 20°C       | 1.4263 1.43                                               | 1.4055 1.41                                 |
| Dk*, edge corrected           | 141.3 140                                                 | 55.4 60                                     |
| Dk*, non-edge<br>corrected    | 167.6 NA                                                  | 68.3 70                                     |
| Base Curve, mm                | 8.47 8.4                                                  | 8.35 8.3                                    |
| Diameter, mm                  | 13.8 13.8                                                 | 14.11 14.0                                  |
| Power, D                      | -1.11 -1.00                                               | -1.03 -1.00                                 |

{4}------------------------------------------------

| Non-clinical<br>Testing | A series of in vitro and in vivo preclinical toxicology and<br>biocompatibility tests were performed to assess the safety and<br>effectiveness of the contact lens. All non-clinical toxicology tests<br>were conducted in accordance with the GLP regulation (21 CFR Part<br>56). All other testing was conducted according to valid scientific<br>protocols.<br><br>The results of the non-clinical testing on the VISTAKON® (galyfilcon<br>A) Contact Lens demonstrate that:<br>The lens material and extracts are not toxic and not irritating Lens physical and material properties are consistent with currently<br>marketed lenses.                                                                                                                                                                                                   |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Testing        | A three-month clinical study was completed to evaluate the safety and<br>efficacy of the VISTAKON® (galyfilcon A) Contact Lens when worn<br>by myopic patients on a daily wear basis. The clinical study was<br>conducted in accordance with FDA's 1994 Guidance Document for<br>Daily Wear Contact Lenses.<br><br>The study evaluated at least 50 patients with a 2:1 ratio of subject<br>device to predicate device for three (3) months. Parameters measured<br>included adverse reactions, keratometry changes, reasons for<br>discontinuations, and the number of reasons for unscheduled lens<br>replacements.<br><br>Clinical evaluation demonstrated similar overall performance in the<br>clinically relevant areas of vision, health, comfort and fit as compared<br>to concurrent controls when used under daily wear conditions. |

{5}------------------------------------------------

### Conclusions Drawn from Studies

| Validity of<br>Scientific Data | Toxicology studies were conducted by a contract<br>laboratory under Good Laboratory Practice Regulations.<br>Microbiology, chemistry, shelf-life stability, and<br>leachability studies were conducted by VISTAKON®<br>laboratories and followed scientific protocols. The data<br>were determined to be scientifically valid under 21 CFR<br>860.7 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence     | Information presented in this Premarket Notification<br>establishes that the VISTAKON® (galyfilcon A) Contact<br>Lens is as safe and effective as the predicate device when<br>used in accordance with the labeled directions for use and<br>for the requested indication.                                                                          |
| Risk and<br>Benefits           | The risks of the subject device as the same as those<br>normally attributed to the wearing of soft (hydrophilic)<br>contact lenses on a daily wear basis. The benefits to the<br>patient are the same as those for other soft (hydrophilic)<br>contact lenses.                                                                                      |

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

OCT 1 6 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vistakon c/o Ms. Jacqueline Zimovan Project Leader Regulatory and Clinical Affairs P.O. Box 17700 Jacksonville, FL 32245-7700

Re: K032340.

Trade/Device Name: VISTAKON® (galyfilcon A) Contact Lens (Spherical, Toric, Multifocal and Toric-Multifocal) (Clear and tinted with UV Blocker) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: July 28, 2003 Received: July 29, 2003

Dear Ms. Zimovan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

### Page 2 - Ms. Jacqueline Zimovan

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Reizi Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

### Statement of Indications for Use

510(k) Number:

K032340

Device Name:

VISTAKON® (galyfilcon A) Contact Lens Clear and Visibility Tint, with UV blocker

Indications for Use:

The VISTAKON® (galyfilcon A) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.

The VISTAKON® (galyfilcon A) Multifocal Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or not-aphakic persons with non-diseased eyes who may have 0.75 D of astigmatism or less.

The VISTAKON® (galyfilcon A) Toric Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or not-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less.

The VISTAKON® (galyfilcon A) Toric-Multifocal Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or not-aphakic persons with non-diseased who may have 10.00 D of astigmatism or less.

VISTAKON® (galyfilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

Eye Care Practitioners may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement (see "Wearing Schedule"). When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use | <div> <span> </span> </div>                                                                                      | OR | Over the Counter |
|------------------|------------------------------------------------------------------------------------------------------------------|----|------------------|
|                  | <div> <span> </span> </div>                                                                                      |    |                  |
|                  | <div> <span>(Division Sign-Off)</span> </div> <div>Division of Ophthalmic Ear,<br/>Nose and Throat Devises</div> |    |                  |
|                  |                                                                                                                  |    |                  |
|                  | 510(k) Number .                                                                                                  |    |                  |

---

**Source:** [https://fda.innolitics.com/device/K032340](https://fda.innolitics.com/device/K032340)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
