The device is intended to be used by health care professionals in the field of echocardiography. It is anticipated that it will be used when echocardiograms must be performed in a location where no interpreting cardiologist is present, and will facilitate prompt interpretation of echocardiograms. The device is capable of transferring ultrasound images over long distances as digital files.
Device Story
Echoencoder is a software-hardware system for converting and transmitting echocardiographic video; inputs include analog video (via VCR/digitizer) or digital video (DICOM). System uses Macintosh computer to convert video to Quicktime format; performs MPEG-4 compression (approx. 25:1 to 50:1 ratio); transfers files to internet server. Optional database (Filemaker Pro-based) tracks interpretations and generates reports. Operated by healthcare professionals in clinical settings; output viewed by cardiologists at remote locations via standard Quicktime player. Facilitates prompt remote interpretation of echocardiograms; improves access to specialist review in rural or underserved areas.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System comprises Macintosh computer, video digitizer, and proprietary software. Inputs: analog video (via firewire) or digital DICOM files. Processing: MPEG-4 video compression (20:1 to 50:1 ratios). Output: 640x480 pixel Quicktime movie files. Connectivity: Internet-based transfer to server. Database: Filemaker Pro-based for report generation.
Indications for Use
Indicated for use by healthcare professionals in echocardiography to facilitate remote interpretation of echocardiograms, particularly in rural areas where an interpreting cardiologist is not present.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K03 23,24
OCT = 2 2003
- 510(k) Summary
| Date prepared: | 7/1/03 |
|-------------------------|------------------------------------------------------|
| Name of contact person: | James P. Loehr, M.D. |
| | Icardiogram, Incorporated |
| | 100 East Six Forks Road |
| | Suite 305 |
| | Raleigh, NC 27609 |
| | 919-534-2150 (telephone) |
| | 919-534-2168 (fax) |
| Device trade name: | Echoencoder |
| Common Name: | Echocardiogram encoding and transfer software system |
| Classification Name: | Class II |
| | 892.2031 Medical image digitizer |
| | 892.2050 Picture archiving and communications system |
Predicate substantially equivalent devices:
| | K990052 Tel-Echo System |
|--|-------------------------|
| | K980060 EchoLINK |
Device description:
Echoencoder is a combined software-hardware system that is composed of three parts: 1) a system for conversion of analog echocardiographic video to a Quicktime movie file format, and subsequent automated transfer of that file to an Internet server; 2) a system for conversion of digital echocardiographic video to a Quicktime movie file format, and subsequent automated transfer of that file to an Internet server; and 3) an optional Server based database system for tracking of interpretations of the video so obtained.
The system for analog video conversion is composed of a Macintosh computer equipped with the echoencoder software, and a video cassette recording device equipped with a video digitizer. The cassette recorder is interfaced with the computer via a firewire cable. The operator inserts a video cassette into the video cassette recorder, and uses the software program to convert the video into a Quicktime movie, perform MPEG-4 based video compression to compress the file by approximately 50:1, and transfer the file to a server where it can be subsequently reviewed by a Cardiologist.
The system for digital video conversion is also composed of a Macintosh computer equipped with the echoencoder software, and is interfaced with a health care provider's network server. The operator selects the file to be converted, generally in DICOM format, and the software subsequently converts the video into a Quicktime movie, performs MPEG-4 based video compression to compress the file by approximately 25:1, and transfers the file to a server where it can be subsequently reviewed by a Cardiologist.
The server based database product is, by the customer's option, interfaced with the above described Echoencoder system. This system that is based on the software product Filemaker Pro, and allows for electronic report generation and review.
{1}------------------------------------------------
## Intended Use
The device is intended to be used by health care professionals in the field of echocardiography. It is anticipated that it will be used when echocardiograms must be performed in a location where no interpreting cardiologist is present, and will facilitate prompt interpretation of echocardiograms. The device is capable of transferring ultrasound images over long distances as digital files.
## Substantial Equivalency Comparison
There are technological differences between Echoencoder and the equivalent devices, but these differences do not alter the effectiveness or safety of the new device.
Echoencoder is substantially equivalent to the EchoLINK device in that both provide for conversion of analog data to a digital format. The EchoLINK device uses MPEG-2 video compression at a ratio of 55:1, and provides full audio sound; depending on the source video, Echoencoder uses a compression ratio of between 20:1. Echoencoder uses a MPEG-4 compression method, which allows for a wide range of final data transmission speeds, resulting in the creation of a Quicktime movie which can be viewed full-screen by a free viewer on any computer platform. Both products provide a method for remote viewing of echocardiograms, and both provide a method of report generation. The Echoencoder method produces an image which is 640 x 480 pixels instead of 325 x 245 for MPACS.
The Tel-Echo system is also designed for remote echocardiography interpretation and uses MPEG compression. It differs from Echoencoder in the type of MPEG used, and sends the video to an Internet server as opposed to transfer directly to an interpreting cardiologist by the Tel-Echo system.
## Conclusion
Echoencoder, EchoLINK, and Tel-Echo are all used for the same purpose, to convert echocardiographic video into digital files for transfer over communications lines. All use compression techniques that are comparable in degree, and all use MPEG, although the Echoencoder technique is unique in that it uses MPEG-4 and can be viewed using a free video player (Quicktime).
Based on the intended use and comparisons with legally marketed devices noted here, Echoencoder is substantially equivalent, safe, and at least as effective as the EchoLINK and Tel-Echo products.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 2 2003
James Loehr. M.D. President . Icardiogram, Inc. 100 East Six Forks Road, Ste. 305 RALEIGH NC 27609
Re: K032324
Trade/Device Name: Echoencoder Regulation Number: 21 CFR §892.2030 Regulation Name: Medical image digitizer Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: July 1, 2003 Received: July 30, 2003
Dear Dr. Loehr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1132324
## INDICATION FOR USE
Icardiogram, Incorporated Applicant:
Applied for 510(k) Number:
Echoencoder Device name:
Indication for Use:
The device is intended to be used by health care professionals in the field of echocardiography. The device is capable of transferring ultrasound images over long distances as digital files. It is used to transmit echocardiograms over the Internet for the purpose of interpretation of echocardiograms at a remote location. It is anticipated that it will be used when echocardiograms must be performed in a location where no interpreting cardiologist is present, and will facilitate prompt interpretation of echocardiograms. It should be particularly useful in rural areas of the United States.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH. Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
OR
Prescription Use Per 21 CFR 801.109
Over-the-Counter (Optional Format 1-2-96)
David G. Ligman
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
K082324
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.