BINAX NOW RSV TEST, MODEL 430-000 AND BINAX NOW NASOPHARYNGEAL SWAB ACCESSORY PACK, MODEL 400-065

K032166 · Binax, Inc. · GQG · Oct 21, 2003 · Microbiology

Device Facts

Record IDK032166
Device NameBINAX NOW RSV TEST, MODEL 430-000 AND BINAX NOW NASOPHARYNGEAL SWAB ACCESSORY PACK, MODEL 400-065
ApplicantBinax, Inc.
Product CodeGQG · Microbiology
Decision DateOct 21, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3480
Device ClassClass 1
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K032166 · Oct 21, 2003BINAX NOW RSV TEST, MODEL 430-000 AND BINAX NOW NASOPHARYNGEAL SWAB ACCESSORY PACK, MODEL 400-065Binax, Inc.Retrospective clinical nasal wash specimens from a university medical centerA retrospective study was conducted to supplement prospective clinical data due to a low number of positive RSV cases in the prospective cohort.Retrospective study; Clinical performance; RSV

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective studyPatients living in the northeastern region of the US; Sample Size: 59 (47 viral culture positive, 12 viral culture negative); Number of Sites: 1 (University medical center)Viral cultureSensitivity and specificity

Indications for Use

The Binax NOW® RSV Test is a rapid immunochromatographic assay for the qualitative detection of respiratory syncytial virus (RSV) fusion protein antigen in nasal wash and nasopharyngeal swab specimens from symptomatic patients. This test is intended for in vitro diagnostic use to aid in the diagnosis of respiratory syncytial virus infections in neonatal and pediatric patients under the age of five. It is recommended that negative test results be confirmed by cell culture.

Device Story

The Binax NOW RSV Test is a rapid, manual immunochromatographic membrane assay for qualitative detection of RSV fusion protein antigen. Input consists of nasopharyngeal swab or nasal wash specimens. Swab samples require elution into transport media or saline; nasal wash samples are used directly. The specimen is applied to a test strip mounted in a cardboard, book-shaped hinged device. The strip contains gold-conjugated anti-RSV antibodies and immobilized anti-RSV antibodies. RSV antigen in the sample binds to the conjugate; these complexes are captured by the immobilized antibody, forming a visible blue sample line. A pink control line forms to verify test function. Results are visually interpreted by a clinician within 15 minutes. The test aids in diagnosing RSV infections in neonatal and pediatric patients. It is intended for professional use in clinical settings.

Clinical Evidence

Clinical performance evaluated via prospective and retrospective studies. Prospective nasal wash study (n=191) showed 89% sensitivity and 98% specificity. Retrospective nasal wash study (n=59) showed 89% sensitivity and 100% specificity. Prospective nasopharyngeal swab study (n=179) showed 93% sensitivity and 93% specificity compared to culture/DFA. Reproducibility study across 3 sites showed 100% agreement.

Technological Characteristics

Immunochromatographic membrane assay; nitrocellulose membrane with immobilized anti-RSV and control antibodies; visualizing particles conjugated to antibodies; cardboard, book-shaped hinged test device; manual visual interpretation; no electronic components or software.

Indications for Use

Indicated for qualitative detection of RSV fusion protein antigen in nasal wash and nasopharyngeal swab specimens from symptomatic neonatal and pediatric patients under age five. Negative results should be confirmed by culture.

Regulatory Classification

Identification

Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: 4K032166 B. Analyte: Respiratory Syncytial virus C. Type of Test: Immunochromatographic membrane assay D. Applicant: Binax Inc. E. Proprietary and Established Names: BINAX NOW® RSV Test F. Regulatory Information: 1. Regulation section: 866.3480 2. Classification: I 3. Product Code: GQG 4. Panel: Microbiology (83) G. Intended Use: 1. Intended use(s): The Binax NOW® RSV Test is a rapid immunochromatographic assay for the qualitative detection of respiratory syncytial virus (RSV) fusion protein antigen in nasal wash and nasopharyngeal swab specimens from symptomatic patients. This test is intended for in vitro diagnostic use to aid in the diagnosis of respiratory syncytial virus infections in neonatal and pediatric patients under the age of five. It is recommended that negative test results be confirmed by cell culture. 2. Indication(s) for use: NA 3. Special condition for use statement(s): NA 4. Special instrument Requirements: NA H. Device Description: The Binax NOW® RSV Test is an immunochromatographic membrane assay used to detect RSV fusion protein antigen in nasal wash and nasopharyngeal swab specimens. Anti-RSV antibody, the Sample Line, is adsorbed onto nitrocellulose membrane. Control antibody is adsorbed onto the same membrane as a second stripe. Both anti-RSV and control antibodies are conjugated to visualizing particles that are dried onto an inert fibrous support. The resulting conjugate pad and the striped membrane are combined to {2} Page 2 of 7 construct the test strip. This test strip is mounted on the right side of a cardboard, book-shaped hinged test device. Swab samples (controls and patients) require a preparation step, in which the sample is eluted off the swab into an appropriate solution. Nasal wash samples do not require any preparation. To perform the test, the sample to be tested is added to the white pad at the top of the test strip, and the test device is closed. RSV antigen present in the sample reacts to bind anti-RSV conjugated antibody. The resulting antigen-conjugate complexes are captured by immobilized anti-RSV antibody, forming the Sample Line. Immobilized Control Line antibody captures a visualizing conjugate, forming a pink Control Line. The Control Line is blue in a device that has not been tested. Test results are interpreted by the presence or absence of visually detectable pink-to-purple colored lines. A positive test result, read at 15 minutes, will include the detection of both a Sample Line and a Control Line. A negative test result, read at 15 minutes, will produce only a Control Line, indicating that RSV antigen was not detected in the sample. Failure of the Control Line to appear, or the Control Line remaining blue, indicates an invalid assay, whether the Sample Line is present or not. The device is provided in kit form, the contents are as follows. Test Devices: A membrane coated with mouse antibody specific for RSV antigen and with control line antibody is combined with mouse anti-RSV and control line antibody conjugates in a hinged test device. The membrane of an untested device contains a blue line at the control line area. Transfer Pipettes: Fixed volume (100 μl) transfer pipettes used to transfer sample to the test devices. Use only pipettes provided by Binax or a calibrated pipette capable of delivering 100 μl sample volume. Positive Control Swab: Inactivated RSV dried onto swab. Negative Control Swab: Inactivated Streptococcus Group A dried onto swab. Elution Ampoules for Control Swabs: Ampoules containing a fixed volume (0.5 ml) of elution solution used to prepare the Control Swabs for testing. Test Vials for Control Swabs: Vials with screw caps used to prepare the Control Swabs for testing. I. Substantial Equivalence Information: 1. Predicate device name(s): BD Directigen™ EZ RSV 2. Predicate K number(s): K022133 {3} Page 3 of 7 3. Comparison with predicate: | Parameter | NOW® Test | | Directigen™ EZ RSV 510(k) #K022133 | | --- | --- | --- | --- | | ANALYTE | RSV antigen | | Same | | DETECTION SYSTEM | Immunochromatographic Membrane Assay | | Same | | TEST MATRIX | Nasopharyngeal specimens | | Same | | INTENDED USE | Detection of RSV antigen in nasopharyngeal specimens | | Same | | ANALYTICAL SENSITIVITY | 1.56 x 10^{-3} to 5.00 x 10^{4} TCID_{50}/ml for 6 RSV A and 5 RSV B strains. | | 4.05 x 10^{2} to 5.95 x 10^{3} TCID_{50}/ml for 2 RSV A and 3 RSV B strains. | | ANALYTICAL SPECIFICITY | Does not cross-react with the 28 bacteria and 20 viruses tested at relevant concentrations in the Binax test. | | Does not cross-react with the 58 bacteria, 2 yeasts and 39 viruses tested at relevant concentrations in the BD test. | | INTERFERING SUBSTANCES | Does not cross-react with 2% whole blood or the 17 OTC and prescription medications tested. Palivisumab may interfere with Binax NOW® Test detection of RSV. | | Does not cross-react with 2% whole blood or the 25 OTC and prescription medications tested, including Palivisumab. | | REPRODUCIBILITY | Reproducibility testing conducted at 3 test sites produced an overall test reproducibility of 100%. | | Reproducibility testing conducted at 3 test sites produced an overall test reproducibility of 99.1%. | | NASAL WASH PERFORMANCE | | | | | | Prospective | Retrospective | Prospective | | SENSITIVITY (95% CI) | Insufficient data for claim | 89% (77.3-95.3) | 87% (80.1-90.1) | | SPECIFICITY (95% CI) | 98% (95.4-99.4) | 100% (75.3-99.8) | 86% (80.1-90.1) | | NASOPHARYNGEAL SWAB PERFORMANCE | | | | | | Prospective | | Prospective | | SENSITIVITY (95% CI) | 92% (75.7-97.6) | | 67% (47.2-82.7) | | SPECIFICITY (95% CI) | 95% (90.6-97.5) | | 92% (88.2-95.2) | J. Standard/Guidance Document Referenced (if applicable): NA K. Test Principle: The Binax NOW® RSV Test is an immunochromatographic membrane assay used to detect RSV fusion protein antigen in nasal wash and nasopharyngeal swab specimens. Anti-RSV antibody, the Sample Line, is adsorbed onto nitrocellulose membrane. Control antibody is adsorbed onto the same membrane as a second stripe. Both anti-RSV and control antibodies are conjugated to visualizing particles that are dried onto an inert fibrous support. The resulting conjugate pad and the striped membrane are combined to construct the test strip. This test strip is mounted on the right side of a cardboard, book-shaped hinged test device. Swab samples (controls and patients) require a preparation step, in which the sample is eluted off the swab into an appropriate solution. Nasal wash samples do not require any preparation. {4} To perform the test, the sample to be tested is added to the white pad at the top of the test strip, and the test device is closed. RSV antigen present in the sample reacts to bind anti-RSV conjugated antibody. The resulting antigen-conjugate complexes are captured by immobilized anti-RSV antibody, forming the Sample Line. Immobilized Control Line antibody captures a visualizing conjugate, forming a pink Control Line. The Control Line is blue in a device that has not been tested. Test results are interpreted by the presence or absence of visually detectable pink-topurple colored lines. A positive test result, read at 15 minutes, will include the detection of both a Sample Line and a Control Line. A negative test result, read at 15 minutes, will produce only a Control Line, indicating that RSV antigen was not detected in the sample. Failure of the Control Line to appear, or the Control Line remaining blue, indicates an invalid assay, whether the Sample Line is present or not. # L. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: A blind study of the Binax NOW® test was conducted at 3 separate sites using a panel of coded specimens containing negative, low positive (LOD), and low/moderate positive controls. Participants performed testing on 3 different days. One hundred percent (100%) of the 234 samples tested were correctly interpreted. # b. Linearity/assay reportable range: NA # c. Traceability (controls, calibrators, or method): NA # d. Detection limit: The LOD level of inactivated RSV, detected $95\%$ of the time, is $93.7~\mu \mathrm{g / swab}$ | RSV μg/swab | Number Detected | % Detection | | --- | --- | --- | | Negative | 0/20 | 0% | | 48.4 | 13/19* | 68% | | 56.7 | 32/40 | 80% | | 67.0 | 35/38* | 92% | | 80.3 | 37/40 | 93% | | 93.7 | 38/40 | 95% | *Three invalid tests were reported, including 1 @ 48.4 μg/swab and 2 @ 67.0 μg/swab # e. Analytical specificity: To demonstrate the immunologic specificity of the Binax NOW® test, 48 potential cross-reactants were tested in the Binax NOW® test. The cross-reactant panel included bacteria and viruses that may be present in respiratory specimens. Bacteria tested at concentrations greater than $1 \times 10^{8}$ organisms/ml and viruses tested at concentrations greater than $1 \times 10^{5} \mathrm{TCID}_{50} / \mathrm{ml}$ did not cross-react in the Binax NOW® Test. Metapneumovirus was tested at a concentration of $2 \times 10^{3} \mathrm{TCID}_{50} / \mathrm{ml}$ and did not cross-react. # f. Assay cut-off: NA {5} Page 5 of 7 2. Comparison studies: a. Method comparison with predicate device: NA b. Matrix comparison: NA 3. Clinical studies: a. Clinical sensitivity: # CLINICAL STUDIES ## Nasal Wash - Specificity (Prospective Study): The performance of the Binax RSV test on nasal wash specimens was compared to cell culture in a multi-center study conducted during the 2002 Flu season at physician offices and clinics located throughout the United States. Nasal wash specimens were collected from children and adults presenting with RSV-like symptoms for 3 days or less and evaluated in the Binax test. The population tested was approximately 46% female and 54% male. Patients were not included in the study if they had received an influenza vaccine within 6 months of the enrollment period, or if they had taken either an influenza or RSV medication within 30 days of the enrollment period. There were no invalid tests reported. One hundred ninety-one (191) nasal wash specimens were tested at 4 different test sites. Binax NOW® test overall specificity was 98%, and overall test agreement was 98%. Ninety-five percent (95%) confidence intervals are listed below. | NOW® | + | Nasal Wash Viral Culture | | | --- | --- | --- | --- | | | | + | - | | NOW® | + | 3 | 3 | | Result | - | 1 | 184 | | 95% CI | | | | | Sensitivity | = | 89% (42/47) | (77.3% - 95.3%) | | Specificity | = | 98% (184/187) | (95.4% - 99.4%) | | Overall Agreement = 98% (187/191) (94.7% - 99.1%) | | | | The Binax NOW® test performed similarly at the 4 test sites as shown in the table below. | | POSITIVE POINTS | SPECIFICITY | | | | --- | --- | --- | --- | --- | | | NOW®/CULTURE | NOW®/CULTURE | % | 95% CI | | Site I | 1/1 | 91/94 | 97 | 91.1-98.8 | | Site II | 2/3 | 83/83 | 100 | 95.7-100 | | Site III | 0/0 | 6/6 | 100 | 59.0-99.6 | | Site IV | 0/0 | 4/4 | 100 | 47.8-99.5 | {6} Page 6 of 7 # Nasal Wash - Sensitivity and Specificity (Retrospective Study): Since there were a low number of positive culture confirmed RSV infections generated during the prospective study using nasal wash specimens, a retrospective study was conducted as follows. Nasal wash specimens from 47 viral culture positive RSV patients and 12 viral culture negative RSV patients were evaluated in the Binax NOW® test. All of the samples were obtained from a large university medical center and had been collected from patients living in the northeastern region of the US. The population tested was approximately 49% male and 51% female. NOW® test sensitivity was 89%, while test specificity was 100%. Overall test agreement was 92%. Ninety-five percent (95%) confidence intervals are listed below. | | | Nasal Wash | | | --- | --- | --- | --- | | | | Viral | Culture | | | | + | - | | NOW® | + | 42 | 0 | | Result | - | 5 | 12 | | | | | 95% CI | | Sensitivity | = | 89% | (42/47) | | Specificity | = | 100% | (12/12) | | Overall Agreement | = | 92% | (54/59) | # Nasopharyngeal Swab - Sensitivity and Specificity (Prospective Study): The performance of the Binax RSV test on nasopharyngeal swab specimens was compared to cell culture/DFA in a multi-center US study conducted during the 2002 and 2003 Flu seasons. Nasopharyngeal swab specimens were collected from children presenting with RSV- or flu-like symptoms. All swab samples were placed in 0.5-3 mls of viral transport media prior to evaluation in the Binax test. The population tested was 43% female and 57% male. One hundred and seventy-nine (179) nasopharyngeal swab specimens were tested. There were no invalid tests reported. Binax NOW® test sensitivity, specificity and overall agreement as compared to culture/DFA were 93%. Ninety-five percent (95%) confidence intervals are listed below. {7} Page 7 of 7 Nasopharyngeal Swab Culture / DFA | NOW® | + | + | - | | --- | --- | --- | --- | | 25 | 10 | | Result | - | 2 | 142 | 95% CI | Sensitivity | = 93% | (25/27) | (76.5% - 97.7%) | | --- | --- | --- | --- | | Specificity | = 93% | (142/152) | (88.3% - 96.4%) | | Overall Agreement | = 93% | (167/179) | (88.6%-96.1%) | The Binax NOW® test performed similarly at all test sites as shown in the tables below. | SENSITIVITY | | | | | --- | --- | --- | --- | | Site | # | % | 95% CI | | 1 | 14/15 | 93 | 69.8-98.4 | | 2 | 9/10 | 90 | 58.7-97.7 | | 3 | 0/0 | NA | NA | | 4 | 2/2 | 100 | 29.2-99.2 | | SPECIFICITY | | | | | --- | --- | --- | --- | | Site | # | % | 95% CI | | 1 | 69/74 | 93 | 85.1-97.0 | | 2 | 20/23 | 87 | 67.6-95.3 | | 3 | 16/18 | 89 | 66.9-96.6 | | 4 | 37/37 | 100 | 90.7-99.9 | b. Clinical specificity: see a. c. Other clinical supportive data (when a and b are not applicable): 4. Clinical cut-off: NA 5. Expected values/Reference range: NA M. Conclusion: The assay should be considered SE; labeling is appropriate and has been reviewed.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...