ENDOHELIX ENDARTERECTOMY DEVICE

K032105 · Endarx, Inc. · DWX · Oct 30, 2003 · Cardiovascular

Device Facts

Record IDK032105
Device NameENDOHELIX ENDARTERECTOMY DEVICE
ApplicantEndarx, Inc.
Product CodeDWX · Cardiovascular
Decision DateOct 30, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4875
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EndoHelix Endarterectomy Device is intended to remove plaque deposits from arteries.

Device Story

Tubular endarterectomy device featuring an expandable helical element at the distal tip; controlled by a sliding tab on the handle. Used with or without a guidewire to remove arterial plaque. Single-use, sterile, non-reusable. Intended for clinical use.

Clinical Evidence

Bench testing only. Testing assessed device operation, wire actuation, and bond pull strengths. All products met specifications.

Technological Characteristics

Tubular device with expandable helical distal tip; sliding tab control mechanism. Materials are biocompatible and previously cleared. Single-use, sterile, Tyvek/Mylar packaging.

Indications for Use

Indicated for the removal of plaque deposits from arteries.

Regulatory Classification

Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ OCT 3 0 2003 ## 510(k) Summary #### General Information | Classification | Class II | |----------------|-----------------------------------------------------------------------------| | Trade Name | EndoHelix™ Endarterectomy Device | | Submitter | ENDARx Inc.<br>3270 Alpine Road<br>Portola Valley, CA 94028<br>650-926-9335 | Contact Scott Adams President ## Intended Use The EndoHelix Endarterectomy Device is intended to remove plaque deposits from arteries. Predicate Devices - Transluminal Endarterectomy Device K904944 from Baxter Healthcare Corporation - Periscope Device from Vascular Architects K003719 #### Device Description The EndoHelix endarterectomy device is a tubular device with an expanding helical element at the distal tip. The expandable helical element is controlled by the user with a sliding tab on the handle. The device is designed to be used with or without a guidewire. The device is packaged in an industry standard Tyvek/Mylar pouch and is provided sterile. The EndoHelix device is a single use product and is not intended to be resterilized or reused. {1}------------------------------------------------ ## Materials All materials used in the manufacture of the EndoHelix Endarterectomy Device are suitable for this use and have been used in numerous previously cleared products. ## Testing Testing was performed to assess product compliance with the product specification. Testing included device operation, wire actuation, and bond pull strengths. All product met specification. #### Summary of Substantial Equivalence The EndoHelix Endarterectomy Device is equivalent to the predicate devices. The indications for use, basic overall function, methods of manufacturing, and materials used are substantially equivalent. ENDARx Inc. believes the EndoHelix Endarterectomy Device is substantially equivalent to existing legally marketed devices. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. OCT 3 0 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 EndarX, Inc. c/o Mr. J. Scott Adams 3270 Alpine Road Portola Valley, CA 94028 Re: K032105 EndoHelix™ Endarterectomy Device Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II (two) Product Code: DWX Dated: October 20, 2003 Received: October 22, 2003 Dear Mr. Adams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. J. Scott Adams Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html Sincerely vours, Ashley B. Evans \$\int_{0}^{1}\$ Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use | 510(k) Number (if known): | This application | |---------------------------|------------------------------------------------------------------------------------------------| | Device Name: | EndoHelixTM Endarterectomy Device | | Indications for Use: | The EndoHelix Endarterectomy Device is<br>intended to remove plaque deposits from<br>arteries. | PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use / OR Over-The-Counter Use _ (Per 21 CFR 801.109) (Optional Format 1-2-96) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number /C032105 :
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...