CHOLESTECH LDX ASPARTATE AMINOTRANSFERASE (AST)

K032027 · Cholestech Corp. · CIS · Sep 5, 2003 · Clinical Chemistry

Device Facts

Record IDK032027
Device NameCHOLESTECH LDX ASPARTATE AMINOTRANSFERASE (AST)
ApplicantCholestech Corp.
Product CodeCIS · Clinical Chemistry
Decision DateSep 5, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1100
Device ClassClass 2

Intended Use

The Cholestech LDX aspartate aminotransferase (AST) test is for the in vitro quantitative determination of aspartate aminotransferase (AST) in whole blood or serum on the Cholestech LDX Analyzer. AST measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis), and heart diseases.

Device Story

Device uses single-use test cassette with Cholestech LDX Analyzer; inputs include venous whole blood, fingerstick whole blood, or serum. Cassette filters cells; plasma reacts with enzymatic reagents on pad; color reaction measured via reflectance photometry. Analyzer reads calibration data from magnetic stripe on cassette. Used in professional point-of-care settings; operated by healthcare professionals. Output is quantitative AST concentration (U/L). Results assist clinicians in diagnosing and managing liver and heart conditions.

Clinical Evidence

Bench testing only. Precision study (n=80 per level) showed total CV 4.4-8.8%. Linearity verified (R²=0.9886). Method comparison against IFCC reference method (n=109) showed slope 0.97, correlation 0.983. Matrix comparison (n=46 venous, n=21 fingerstick) confirmed performance across sample types. Interference testing showed <10% interference for listed substances.

Technological Characteristics

Enzymatic solid-phase assay; reflectance photometry. Measures AST via L-Aspartic acid/alpha-Ketoglutarate reaction cascade. Cassette-based with magnetic stripe for calibration. Assay range 10-400 U/L. Sample types: whole blood (capillary/venous) or serum. Operates at 37°C. Point-of-care analyzer platform.

Indications for Use

Indicated for professional, point-of-care quantitative determination of AST in whole blood or serum for patients requiring diagnosis and treatment monitoring of liver diseases (e.g., viral hepatitis, cirrhosis) and heart disease.

Regulatory Classification

Identification

An aspartate amino transferase (AST/SGOT) test system is a device intended to measure the activity of the enzyme aspartate amino transferase (AST) (also known as a serum glutamic oxaloacetic transferase or SGOT) in serum and plasma. Aspartate amino transferase measurements are used in the diagnosis and treatment of certain types of liver and heart disease.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K032021 Submitter Information (21 CFR 807.92(a)(1)) | Submitter: | Cholestech Corporation | |------------|---------------------------| | | 3347 Investment Boulevard | | | Hayward, CA 94545 | | | phone: (510) 732-7200 | | | fax: (510) 732-7227 | Thomas E. Worthy, PhD. Contact: Vice President, Research and Regulatory Affairs Cholestech Corporation. Summary Date: June 30, 2003 Name of Device and Classification (21 CFR 807.92(a)(2)) Name (trade): Cholestech LDX aspartate aminotransferase (AST) Name (usual): Colorimetric assay for the determination of aspartate amino transferase (SGOT/AST) Classification: 21 CFR 862.1100, Class II, CIS (75) ### Identification of Legally Marketed Predicate Device(s) (21 CFR 807.92 (a)(3)) LDX AST is substantially equivalent to the Synchron CX® AST system (Synchron CX, Beckman Coulter, Inc., Fullerton, CA). The LDX AST method is identical or similar to its predicate in terms of: intended use, measurement principle (enzymatic reactions), measurement (assay) range, specimen type, and the requirement for an analyzer. {1}------------------------------------------------ ### Description of Device (21 CFR 807.92 (a)(4)) The Cholestech LDX System combines enzymatic methodology and solid-phase technology to measure AST. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin coated capillary tube), venous whole blood or serum. The sample is applied to a Cholestech LDX AST cassette. The cassette is then placed into the Cholestech LDX Analyzer where a unique system on the cassette separates the plasma from the blood cells. The plasma flows to both sides of the cassette and is transferred to the AST reaction pad. The Cholestech LDX Analyzer measures Aspartate aminotransferase by an enzymatic method based on the method formulation of Katsuyama et al. 12- Aspartic acid aminotransferase catalyzes the transfer of amino groups from L-Aspartic acid to alpha-Ketoglutarate producing oxaloacetate and glutamate. Oxaloacetate Decarboxylase converts the Oxaloacetate to Pyruvate by the removal of CO2. Pyruvate oxidase, in the presence of oxygen, oxidizes the pyruvate to acetylphosphate and hydrogen peroxide. In a reaction catalyzed by horseradish peroxidase, the peroxide reacts with an indicator dye to form a blue color at a rate proportional to the AST concentration of the sample. The resultant color in the reaction is measured by reflectance photometry. Image /page/1/Figure/5 description: This image shows a series of four chemical reactions. In the first reaction, L-Aspartic acid and alpha-Ketoglutarate react with Aspartate Aminotransferase to produce Oxaloacetate and Glutamate. In the second reaction, Oxaloacetate reacts with Oxaloacetate Decarboxylase to produce Pyruvate and CO2. In the third reaction, Pyruvate, Phosphate, O2, and H2O react with Pyruvate Oxidase to produce Acetylphosphate, CO2, and H2O2. In the fourth reaction, Indicator and H2O2 react with Peroxidase to produce Colored Blue Dye and H2O. A brown magnetic stripe on each cassette contains the calibration required for the Cholestech LDX Analyzer to convert the reflectance reading to the AST concentration in U/L. 37°C. ## Intended Use (21 CFR 807.92 (a)(5)) The Cholestech LDX aspartate aminotransferase (AST) test is for the in vitro quantitative determination of aspartate aminotransferase (AST) in whole blood or serum on the Cholestech LDX Analyzer. AST measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis), and heart diseases. {2}------------------------------------------------ : # Similarities to the Predicate(s) (21 CFR 807.92 (a)(6)) A summary table of the similarities and differences between the LDX AST and the predicate device follows. | Device Name | LDX AST<br>(new device) | Synchron CX® AST<br>(K952427) | |-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications<br>for use | The Cholestech LDX aspartate<br>aminotransferase (AST) test is for<br>the in vitro quantitative<br>determination of AST in whole<br>blood or serum on the Cholestech<br>LDX Analyzer. | The Synchron CX reagent, in conjunction with<br>the Synchron® Enzyme Validator Set, is<br>intended for the quantitative determination of<br>AST activity in serum or plasma on Synchron<br>CX Clinical Systems. | | | AST measurements are used in the<br>diagnosis and treatment of certain<br>liver diseases (e.g., viral hepatitis<br>and cirrhosis), and heart diseases. | AST measurements are used in the diagnosis<br>and treatment of certain types of liver and<br>heart diseases. | | Instrument<br>Required | LDX Analyzer | Synchron CX Clinical System | | Technology | Enzymatic methodology and a<br>solid-phase technology | Enzymatic rate method; appropriate sample<br>and reagent volumes are introduced to the<br>system's cuvette. | | Assay Range | 10 to 400 U/L | 5 to 400 U/L | | Sample Type | Whole blood (capillary and venous)<br>and serum | Serum or plasma | | Calibration<br>Requirements | No calibration performed by the<br>user; test information is encoded on<br>the magnetic stripe of the cassette,<br>and the stripe is read by the LDX<br>Analyzer each time a cassette is<br>run. | Calibration required via the use of the<br>Synchron Enzyme Validator Set; under typical<br>operating conditions, the AST reagent<br>cartridge must be calibrated every 5 days, and<br>also with certain parts replacement or<br>maintenance procedures. | | Testing | Professional-Use, point-of-care | Professional-Use, conventional laboratory | # Comparison Table: Cholestech LDX AST vs Beckman Coulter Synchron CX AST {3}------------------------------------------------ | Device Name | LDX AST<br>(new device) | Synchron CX® AST<br>(K952427) | |-------------|-------------------------|-------------------------------| | Environment | | | Brief Discussion of Nonclinical and Clinical Performance Data (21 CFR 807.92(b)(1,2, 3)} - 10 -- 400 U/L Assay range: . - up to 50% Hematocrit tolerance: . - less than 10% interference when challenged by evaluated levels of . Interference testing: endogenous and exogenous (therapeutic) substances - Precision: 2 levels of Controls (Low- ~31 U/L AST, and High- ~106 U/L AST) were tested . in duplicate, twice a day, over a 20 day period for a total of 80 replicates per level. The percent coefficient of variation (%CV) from the testing of the Low Control was 8.8%, and 4.4% from the testing of the High Control. When the same testing protocol was performed with a whole blood sample at 58 U/L, the %CV was 4.8%. - Accuracy: The LDX AST test was compared to the Synchron CX AST test with 109 . matched serum samples. The comparative data appear below (Synchron CX on x-axis). | LDX AST vs Synchron CX AST | | | | | |----------------------------|-------|-------------|-------|-----------------| | n | slope | y-intercept | “r” | Range of Values | | 109 | 0.97 | 1.6 | 0.983 | 12 – 396 U/L | A study was performed where results obtained from testing whole blood samples (both venous and fingerstick) on the LDX were compared to results obtained from testing matched serum samples on the LDX. The comparative data appear below (serum results on x-axis). | LDX AST: venous whole blood samples vs serum samples | | | | | |------------------------------------------------------|--|--|--|--| | | | | | | | n | slope | y-intercept | "r" | Range of Values | |----|-------|-------------|-------|-----------------| | 46 | 1.08 | 0.3 | 0.998 | 13 - 343 U/L | #### LDX AST: fingerstick whole blood samples vs serum samples | n | slope | y-intercept | "r" | Range of Values | |----|-------|-------------|-------|-----------------| | 21 | 0.86 | 4.4 | 0.934 | 13 - 65 U/L | {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the figure. SEP - 5 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Thomas E. Worthy, Ph.D. Vice President, Research and Regulatory Affairs Cholestech Corporation 3347 Investment Boulevard Hayward, CA 94545 k032027 -Re: Trade/Device Name: Cholestech LDX aspartate aminotransferase (AST) Test Regulation Number: 21 CFR § 862.1100 Regulation Name: Aspartate amino transferase (AST/SGOT) Test System Regulatory Class: II Product Code: CIS Dated: June 30, 2003 Received: July 1, 2003 Dear Dr. Worthy We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). {5}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use Statement 510(K) Number (if known): Device Name: Cholestech LDX aspartate aminotransferase (AST) Test #### Indications for Use: The Cholestech LDX aspartate aminotransferase (AST) test is for the in vitro quantitative determination of AST in whole blood or serum on the Cholestech LDX Analyzer. AST measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis), and heart diseases. Carol C Benson for Jean Cooper, DVM Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K032027 (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE AS NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use * (Per 21 CFR 801.109) OR Over -the-Counter Use ________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...