← Product Code [LPS](/productcode/LPS) · K032012

# HOMOCYSTEINE MICROTITER PLATE ASSAY (K032012)

_Diazyme Laboratories · LPS · Sep 3, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K032012

## Device Facts

- **Applicant:** Diazyme Laboratories
- **Product Code:** [LPS](/productcode/LPS.md)
- **Decision Date:** Sep 3, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1377
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Homocysteine Microtiter Plate Assay is intended for the quantitative determination of total L-homocysteine in human serum or plasma. The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.

## Device Story

The Homocysteine Microplate HPB Assay is an EIA-like diagnostic test for measuring total L-homocysteine (tHcy) in human serum or plasma. The process begins with the pretreatment of plasma samples using a reducing agent, TCEP, to convert protein-bound homocysteine into free homocysteine. Subsequently, SAH hydrolase converts the free homocysteine into S-adenosyl-L-homocysteine (SAH). The assay utilizes a genetically engineered Homocysteine Binding Protein (HBP) as a capturing reagent in a competition assay between the sample's SAH and a tracer SAH-HRP conjugate. The resulting signal is quantitated to determine tHcy levels. The device is intended for clinical laboratory use to assist healthcare providers in diagnosing and managing patients with suspected hyperhomocysteinemia and homocystinuria.

## Clinical Evidence

No clinical studies performed. Performance established via bench testing: precision (intra-assay CV 3.8-4.6%, inter-assay CV 6.7-9.2%), linearity (r²=0.99, range 1.5-60 μmol/L), and method comparison against predicate (n=107, Y=0.95X+0.72, r=0.96). Analytical specificity and interference testing conducted.

## Technological Characteristics

EIA-like competitive assay; utilizes genetically engineered Homocysteine Binding Protein (HBP) as a capturing reagent; TCEP reducing agent; SAH hydrolase enzyme; microtiter plate format.

## Regulatory Identification

A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.

## Predicate Devices

- Axis Homocysteine EIA (k980907)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k032012

B. Analyte:
Homocysteine

C. Type of Test:
Quantitative

D. Applicant:
Diazyme Laboratories

E. Proprietary and Established Names:
Homocysteine Microplate HPB Assay; Homocysteine Controls (Lyophilized Form)

F. Regulatory Information:
1. Regulation section:
21 CFR 862.1377; 21 CFR 862.1660
2. Classification:
Class II; Class I
3. Product Code:
LPS; JJX
4. Panel:
75

G. Intended Use
1. Intended use(s):
See indication for use below.
2. Indication(s) for use:
The Homocysteine Microplate HPB Assay is intended for the quantitative determination of total L-homocysteine in human serum or plasma.

The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.

The Homocysteine Controls (Lyophilized Form) are intended for use as an assayed quality control serum to monitor the precision of the laboratory testing procedures for homocysteine.

3. Special condition for use statement(s):
NA

{1}

Page 2 of 4

4. Special instrument Requirements:
Microtiter plate reader capable of reading at 450nm

H. Device Description:
The in vitro diagnostic reagent kit contains avidin coated microtiter stripwells (96), buffers, hydrolase, substrate, conjugate, enzyme inhibitor, adenosine in tris buffer, adenosine deaminase, stopping solution, Hcy-binding protein, and calibration materials (6). The two control materials containing lyophilized human plasma are sold separately.

I. Substantial Equivalence Information:
1. Predicate device name(s):
Axis Homocysteine EIA
2. Predicate K number(s):
k980907
3. Comparison with predicate:

|  Item | Homocysteine Microplate HPB | Axis Homocysteine EIA  |
| --- | --- | --- |
|  Intended use | For the quantitative determination of total L-homocysteine in human serum or plasma. | Same  |
|  Type of assay | enzyme immunoassay-like assay | antibody based enzyme immunoassay  |
|  Calibrator and control materials | 6 liquid calibrator materials included with kit; 2 lyophilized control materials sold separately | 6 liquid calibrator materials included with kit; 3 liquid control materials sold separately  |
|  Analytical range | 1.5 to 60 μmol/L | 1.0 to 50 μmol/L  |
|  Type of samples to be assayed | serum or EDTA plasma | Same  |
|  Pretreatment of samples | Add 20μL of sample and 150 μL of pretreatment reagent | Add 25 μL of sample and 500 μL of pretreatment reagent  |
|  Calculation of results and detection instrument | Four parameter logistic curve using a microplate reader at 450 nm | Same  |

J. Standard/Guidance Document Referenced (if applicable):
NCCLS Guideline EP5-A, (Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (1999))

K. Test Principle:
This is an EIA-like assay using a genetically engineered homocysteine binding protein as the capture reagent. After pretreatment of samples with a reducing agent, a

{2}

competition between free homocysteine from the samples and the tracer conjugate for binding sites on the binding protein occurs.

L. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:
Three specimens with homocysteine concentrations of 7.0 μmol/L, 10.5 μmol/L and 22.0 μmol/L were assayed in quadruplicate, for twenty days. The intra-assay %CVs were 4.1%, 4.6%, 3.8%, respectively and the inter-assay %CVs were 9.0%, 6.7% and 9.2%, respectively.

b. Linearity/assay reportable range:
The linearity was evaluated by diluting four samples, homocysteine concentrations between 15 and 60 μmol/L, with assay buffer. The assayed results were plotted on the Y axis and the theoretical results were plotted on the X axis. The resulting regression equation was Y = 0.98X – 0.57 umol/L, r² = 0.99. The dynamic range of the assay is stated as 1.5 – 60 μmol/L.

c. Traceability (controls, calibrators, or method):
The traceability of the control and calibrator materials is stated as using a known reference standard that is characterized by an HPLC reference method.

d. Detection limit (analytical sensitivity):
The 4 μmol/L SAH-calibrator was diluted with solution A to obtain sample concentrations of 3, 2.5, 2.0, 1.5 and 1.0 μmol/L. The samples were analyzed in replicates of six. The limit of quantification is defined as the lowest concentration having a CV &lt;20%. The value obtained is 1.5 μmol/L.

e. Analytical specificity:
The interference was determined for bilirubin, hemoglobin, lipids (triglycerides), red blood cells, protein and sodium fluoride by spiking them into plasma samples. The analysis showed that all interference values were less than 10% except for the following: bilirubin (0.2 mg/mL) showed 12.6% interference for homocysteine at 6.2 μmol/L; protein (60 mg/mL) showed -13.3% interference for homocysteine at 21 μmol/L; and sodium fluoride (5 mg/mL) showed -11.3% interference for homocysteine at 6.2 μmol/L.

The cross-reactivity was determined for adenosine (5 mmol/L), adenosyl-L-methionine (0.5 mmol/L), cystathionine (0.5 mmol/L), L-cysteine (100 mmol/L), glutathione (5 mmol/L), thiolactone (0.5 mmol/L) by spiking them into plasma samples. The analysis showed that all cross-reactivity values were less than 2.5%, except, L-cysteine which showed 2.7% cross-reactivity for homocysteine at 19.7 μmol/L.

f. Assay cut-off:
Not determined

{3}

Page 4 of 4

2. Comparison studies:

a. Method comparison with predicate device:

A study using 101 patient samples, ranging in concentration from 6.41 to 19.19 µmol/L and 6 spiked samples, ranging in concentration from 7.49 to 45.53 µmol/L, was performed. The samples were assayed by this device (Y) and the predicate device (X). The results of the linear regression analysis yielded the equation, Y = 0.95X +0.72, r = 0.96 for the 107 samples tested.

b. Matrix comparison:

No separate matrix comparison studies were performed. Published literature is provided in the labeling as a reference for the use of both serum and plasma samples.

3. Clinical studies:

None performed

4. Clinical cut-off:

Not determined

5. Expected values/Reference range:

Studies published in the literature are provided in the labeling for expected values for males and females. In particular, the U.S. National Health and Nutrition Examination Survey (NHANES), published in Annals of Internal Medicine (1999) 131, 331-339, is cited as representing the US population. A chart is provided in the labeling as follows:

|   | 12-19 years | ≥ 60 years of age | Cut-off for “high” levels  |
| --- | --- | --- | --- |
|  Male | 4.3-9.9 µmol/L | 5.9 -15.3 µmol/L | ≥ 11.4 µmol/L  |
|  Female | 3.3-7.2 µmol/L | 4.7-11.6 µmol/L | ≥10.4 µmol/L  |

Clinical limitations:

The labeling states that persons taking methotrexate, carbamazepine, phenytoin, nitrous oxide, anticonvulsants or 6-azauridine triacetate may have higher homocysteine values due to metabolic interference with homocysteine metabolism.

M. Conclusion:

Based upon the information provided for the file, I recommend that this device is substantially equivalent to the predicate device, regulated by 21 CFR 862.1377.

---

**Source:** [https://fda.innolitics.com/device/K032012](https://fda.innolitics.com/device/K032012)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
