K031979 · Conmed Integrated Systems (Cmis) · FSY · Jul 25, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031979
Device Name
CONMED INTEGRATED SYSTEMS NURSES ASSISTANT
Applicant
Conmed Integrated Systems (Cmis)
Product Code
FSY · General, Plastic Surgery
Decision Date
Jul 25, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Nurse's Assistant® is intended to be used to turn on and off, and adjust certain settings of, endoscopic and surgical cameras, electrosurgical generators such as the ConMed System 5000 ESU, laparoscopic insufflators such as the Linvatec GS1002 Insufflator, surgical lamps, operating room ("O.R.") lights, and operate digital documentation products, VCR's, video monitors, video printers, radios and CD players.
Device Story
Nurse's Assistant is a programmable controller with touch screen GUI; provides centralized remote interface for O.R. staff to control and monitor settings of diverse surgical equipment; eliminates need for individual device control panels. Inputs include user commands via touch screen; outputs include control signals to connected devices (cameras, electrosurgical generators, insufflators, surgical lights, documentation/media equipment). Used in O.R. by professional staff; simplifies workflow by consolidating equipment management; improves efficiency by allowing remote adjustment of surgical environment settings.
Clinical Evidence
Bench testing only. Device performance verified against international safety and electromagnetic compatibility standards including IEC 601-1, EN 60601-1-2, and UL 2601-1.
Technological Characteristics
Programmable controller with touch screen GUI. Connectivity to surgical equipment via remote interface. Tested to IEC 601-1, IEC 601-2-18, EN 60601-1-2, UL 2601-1, and various IEC 61000-4-x EMC standards.
Indications for Use
Indicated for use in general, cardiovascular, ENT, gastroenterology, urology, plastic, obstetrics, gynecology, and orthopedic surgery, and general thorascoscopy, general cardiothoracic surgery, general laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy. Applicable to procedures including laparoscopic cholecystectomy, hernia repair, appendectomy, pelvic lymph node dissection, hysterectomy, spinal fusion, lung/pleural biopsy, sympathectomy, pleurodesis, coronary artery bypass, valve replacement, and arthroscopic repairs.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
ConMed Integrated Systems' Nurse's Assistant® O.R. Control System (K010754)
Olympus' Integrated Endosurgery System EndoALPHA (K022270)
Computer Motion's HERMES Operating Room Control Center (K030240)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird in flight, with its wings and tail feathers represented by thick, curved lines. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ConMed Intergrated Systems Mr. C. Jeff Lipps Director of Regulatory Affairs 1815 NW 169th Place, Suite 4020 Beaverton, OR 97006
JUL 2 7 2015
Re: K031979
> Trade/Device Name: ConMed Integrated Systems' Nurse's Assistant® O.R. Control System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ, FET, FSY, HET, HIF, GEI and LMC Dated (Date on orig SE ltr): June 25, 2003 Received (Date on orig SE ltr): June 26, 2003
Dear Mr. Lipps,
This letter corrects our substantially equivalent letter of July 25, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{2}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# XIV. INDICATIONS FOR USE STATEMENT
510(k) Number: K031979
Device Name:
#### Modification to ConMed Integrated Systems' Nurse's Assistant® O.R. Control System
Indications for Use:
The Nurse's Assistant® is a programmable controller operated through a touch screen GUI that offers professional O.R. staff a simplified remote user interface and a real time display of settings for the devices attached to the system thereby eliminating the necessity for using the various control panels on diverse surgical equipment.
The Nurse's Assistant® is intended to be used to turn on and off, and adjust certain settings of, endoscopic and surgical cameras, electrosurgical generators such as the ConMed System 5000 ESU, laparoscopic insufflators such as the Linvatec GS1002 Insufflator, surgical lamps, operating room ("O.R.") lights, and operate digital documentation products, VCR's, video monitors, video printers, radios and CD players.
The Nurse's Assistant® is indicated for use in general, cardiovascular, ENT, gastroenterology, urology, plastic, obstetrics, gynecology, and orthopedic surgery, and general thorascoscopy, general cardiothoracic surgery, general laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy. A few examples of the more common surgical procedures where this system could be used are laparoscopic cholecystectorny, laparoscopic hernia repair, laparoscopic appendectorny, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectorny, laparoscopic and thorascopic anteriorspinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated, examination of the evacuated cardiac chamber during performance of valve replacement, arthroscopic meniscus repair, anterior cruciate ligament repair and associated procedures.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
oncurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use_
(Per 21 C.F.R. 801.109)
(Optional Formal 1-2-96)
ConMed Integrated Systems Nurse's Assistant Special 510/k) Notification
{4}------------------------------------------------
03/979
Confidential
#### JUL 25 2003 XIII. 510(K) SUMMARY
# Modified Nurse's Assistant® O.R. Control System
In accordance with 21 CFR section 807.92, ConMed Integrated Systems is submitting the following 510(k) summary:
# 1) Date Submitted
June 25, 2003
## 2) Submitter Information
ConMed Integrated Systems 1815 NW 169th Place, Suite 4020 Beaverton, OR 97006 Contact: C. Jeff Lipps, BS/MBA, Dir. of RA (503) 614-1106 ext. 1206 Phone Fax (503) 614-1109
### 3) Name of Device
| Proprietary Name: | Nurse's Assistant® O.R. Control System |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Surgical Control Center |
| Classification Names: | Surgical Lamp, 878.4580<br>Endoscope and accessories, 876.1500<br>Gynecologic laparoscope and accessories, 884.1720<br>Laparoscopic insufflator, 884.1730<br>Electrosurgical cutting and coagulation device<br>and accessories, 878.4400<br>Medical image hardcopy device, 892.2040 |
| Product Codes: | FSY, GCJ, FET, KOG, HET, HIF, GEI, LMC |
| Classification Panel: | General & Plastic Surgery |
| Device Class: | Class II |
1 of 3
ConMed Integrated Systems Nurse's Assistant Special 510(k) Notification
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "Ko3/1979 Confidential". The text is written in black ink on a white background. The word "Confidential" is written in a smaller font size than the other text.
510(k) Summary - Modified Nurse's Assistant® O.R. Control System
Page 2
#### 4) Predicate Devices
This device is substantially equivalent to the ConMed Integrated Systems' Nurse's Assistant®O.R. Control System (K010754), Modification to Olympus' Integrated Endosurgery System EndoALPHA (K022270), and Modification to Computer Motion's HERMES Operatir g Room Control Center (K030240).
### 5) Device Description
The Nurse's Assistant is a programmable controller operated through a touch screen GUI that offers professional O.R. staff a simplified remote user interface and a real time display of settings for the devices attached to the system thereby eliminating the necessity for using the various control panels on diverse surgical equipment.
#### 6) Intended Use
The Nurse's Assistant is intended to be used to turn on and off, and adjust certain settings of, endoscopic and surgical cameras, electrosurgical generators such as the ConMed System 5000 ESU, laparoscopic insufflators such as the Linvatec GS1002 Insufflator, surgical lamps, operating room ("O.H.") lights, and operate digital documentation products, VCR's, video monitors, video printers, radics and CD players.
#### 7) Indications for Use:
The Nurse's Assistant® is indicated for use in general, cardiovascular, ENT, gastroenterology, urology, plastic, obstetrics, gynecology, and orthopedic surgery, and general thorascoscopy, general cardiothoracic surgery, general laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy. A few examples of the more common surgical procedures where this system could be used are laparoscopic cholecystectorny, laparoscopic hernia repair, laparoscopic appendectorny, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic and thorascopic anteriorspinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectorny, pleurodesis, internal mammary artery dissection for coronary antery bypass, coronary artery bypass grafting where endoscopic visualization Is indicated, examination of the evacuated cardiac chamber during performance of valve replacement, arthroscopic meniscus repair, anterior cruciate ligament repair and associated procedures.
ConMed Integrated Systems Nurse's Assistant Special 510{k} Notification 27
2 of 3
{6}------------------------------------------------
Ko31979
Confidential
510(k) Summary – Modified Nurse's Assistant® O.R. Control System
Page 3
# 8) Applied Safety Standards:
The Nurse's Assistant®O.R. Control System has been tested to the following standards:
| Test | Title |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| IEC 601-1 | International Standard for Medical Electrical<br>Equipment |
| IEC 601-1 Amendment 1 | International Standard for Medical Electrical<br>Equipment |
| IEC 601-2-18 | International Standard for Medical Electrical<br>Equipment |
| EN 60601-1 | International Standard for Medical Electrical<br>Equipment |
| EN 60601-1-1 | General Requirements for Safety - Collateral<br>Standard |
| EN 60601-1-2 (2001) | International Standard for Medical Electrical<br>Equipment, Electro Magnetic Compatibility<br>(EMC) |
| CAN/CSA C22.2 601.1-M90 | Medical Electrical Equipment - Part 1: General<br>Requirements for Safety, General Instruction<br>No. 1; Supplement 1; 1994 R(1997) |
| UL 2601-1, 2nd Ed. | UL Standard for Safety Medical Electrical<br>Equipment, Part 1: General Requirements for<br>Safety, 2nd Ed. |
| EN 55011 | Conducted Emissions |
| EN 55011 | Radiated Emissions |
| IEC 61000-4-2 | Electrostatic Discharge |
| IEC 61000-4-3 | Radio Frequency Electromagnetic Fields |
| IEC 61000-4-4 | Fast Transients Common Mode |
| IEC 61000-4-5 | High Energy Surge |
| IEC 61000-4-6 | Conducted Immunity |
28
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.