PROCEED TRILAMINATE SURGICAL MESH

K031925 · ETHICON, Inc. · FTL · Sep 17, 2003 · General, Plastic Surgery

Device Facts

Record IDK031925
Device NamePROCEED TRILAMINATE SURGICAL MESH
ApplicantETHICON, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateSep 17, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

PROCEED Mesh may be used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Device Story

PROCEED Surgical Mesh is a sterile, trilaminate implant for hernia repair and fascial deficiency reinforcement. Device consists of three layers: PROLENE Soft Mesh (nonabsorbable polypropylene), oxidized regenerated cellulose (ORC) fabric, and a polydioxanone polymer film. Implanted by surgeons in clinical settings to provide structural support; ORC layer serves to minimize tissue attachment to the base mesh. Device functions as a physical reinforcement material; clinical benefit derived from hernia repair and fascial defect closure.

Clinical Evidence

Non-clinical laboratory testing and in-vivo testing provided. Results demonstrate the device is comparable to standard surgical mesh devices indicated for hernia repair and fascial deficiencies, performing as intended and claimed.

Technological Characteristics

Trilaminate surgical mesh; materials include polypropylene (PROLENE Soft Mesh), oxidized regenerated cellulose (ORC), and polydioxanone polymer film. Designed for hernia repair and fascial deficiency reinforcement. Functionally similar to composite meshes using ORC to minimize tissue attachment.

Indications for Use

Indicated for patients requiring hernia repair or fascial deficiency reinforcement/bridging.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KO31925 (1. 1982) ## SECTION 8 ## SUMMARY OF SAFETY AND EFFECTIVENESS | 510(k) Summary | Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is summarized below. For the convenience of the Reviewer, this summary is formatted in accordance with the Agency's final rule "...510(k) Summaries and 510(k) Statements..." (21 CFR 807) and can be used to provide a substantial equivalence summary to anyone requesting it from the Agency.<br>NEW DEVICE NAME: PROCEED* Surgical Mesh<br>PREDICATE DEVICES NAME: PROLENE Soft Polypropylene Mesh. | |----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Description | PROCEED* Surgical Mesh is a sterile mesh designed for the repair of hernias and other fascial deficiencies. The mesh product has separate layers comprising of PROLENE* Soft Mesh, a nonabsorbable polypropylene mesh, a fabric of oxidized regenerated cellulose (ORC) and a polydioxanone polymer film. | | Intended Use | PROCEED Mesh may be used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result. | | Indications Statement | PROCEED Mesh may be used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result. | | Technological<br>Characteristics | PROCEED has similar technological characteristics as the<br>predicate devices. The addition of another layer to the<br>standard base surgical mesh material is common to the<br>W.L. GORE & Associates, Inc. DualMesh** Biomaterial<br>(K963619), the Genzyme Corp. Sepramesh** Biosurgical<br>Composite (K994328) and the Bard Composix** Mesh<br>(510K K971745) with a performance characteristic of<br>minimizing tissue attachment to the base mesh material.<br>With the addition of the ORC layer to minimize tissue<br>attachment to the base PROLENE Soft Mesh, the<br>PROCEED Mesh continues to meet the functional<br>requirements of a surgical mesh | | Performance Data | Non-clinical laboratory testing was performed<br>demonstrating that the device is comparable to standard<br>surgical mesh devices that are indicated for hernia repair<br>and other fascial deficiencies that require the addition of a<br>reinforcing or bridging material to obtain that desired<br>surgical result. Additionally, in-vivo testing was provided<br>showing that the device performed as intended and as<br>claimed. | | Conclusions | Based on the 510(k) summaries and 510(k) statements (21<br>CFR 807) and the information provided herein, we<br>conclude that the modified device is substantially<br>equivalent to the Predicate Devices under the Federal Food,<br>Drug, and Cosmetic Act. | | Contact | Rey Librojo<br>Senior Project Manager, Regulatory Affairs<br>ETHICON Products<br>ETHICON, Inc.<br>Rt. #22, West<br>Somerville, NJ 08876-0151 | | Date | September 3, 2003 | {1}------------------------------------------------ ψ31925 (P. 20A2) {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a stylized drawing of three birds in flight, arranged closely together. The birds are depicted with simple, curved lines, giving them a sense of movement. A dotted circle partially surrounds the birds, adding a sense of enclosure or focus to the composition. The drawing is in black and white, with a minimalist aesthetic. food and Drug 200 Corporate Boute Rockville MD 20050 SEP 1 7 2003 Mr. Rev Librojo Senior Project Manager, Regulatory Affairs Ethicon, Inc. Rt. #22 West Somerville, New Jersey 08876-0151 Re: K031925 Trade/Device Name: PROCEED* Trilaminate Surgical Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Polymeric surgical mesh Regulatory Class: II Product Code: FTL Dated: June 20, 2003 Received: June 27, 2003 Dear Mr. Librojo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Rey Librojo This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE 510(k) Number (if known): K431925 Device Name: Indications for Use: PROCEED* Trilaminate Surgical Mesh PROCEED Mesh may be used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result. #### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _ (Per 21 CFR 801.109) OR , i Over-The Counter Use (Optional Format 1-2-9G) Miriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices K031925 510(k) Number_ Page iii
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...