← Product Code [LLZ](/productcode/LLZ) · K031778

# PRACTICEBUILDER 1-2-3 (K031778)

_Erad Imagemedical Corp. · LLZ · Aug 7, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K031778

## Device Facts

- **Applicant:** Erad Imagemedical Corp.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 7, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

PracticeBuilder 1-2-3 is a PACS and teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. PracticeBuilder 1-2-3 is for use in hospitals, imaging centers, radiologist reading practices and any user who requires and is granted access to patient image, demographic and report information.

## Device Story

PracticeBuilder 1-2-3 is a PACS and teleradiology system; components include GatewayServer, SendServer, SmartServer, ArchiveServer, and diagnostic workstations. System acquires DICOM images and scheduling info; retrieves prior studies from archive; registers data for operator worklist access. Radiologists use diagnostic workstations to view images; system supports optional voice dictation and transcription. Reports are registered and made available to referring physicians or exported to information systems. Used in hospitals, imaging centers, and reading practices; facilitates clinical decision-making by providing diagnostic-quality images and reports to authorized users via web or custom interfaces.

## Clinical Evidence

Bench testing only. Software package validated through testing by programmers, non-programmers, quality control staff, and potential customers to ensure diagnostic quality of images and associated information.

## Technological Characteristics

PACS software system; utilizes off-the-shelf computer hardware components. Connectivity via network (web and customized interfaces). Supports DICOM standard for image acquisition and transmission. Operates as a client-server architecture with central system management, acquisition servers, and diagnostic workstations.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Stentor iSite ([K013630](/device/K013630.md))
- Stentor iVault (Class I, Exempt)
- Agfa IMPAX ([K922292](/device/K922292.md))
- Ultravisual Visual PACS (Vortex) ([K012097](/device/K012097.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K031778

## AUG - 7 2003

### 510(k) SUMMARY

In accordance with the provisions of the Safe Medical Device Act of 1990, eRAD ImageMedical Corp., is providing a summary of safety and effectiveness information regarding the Practice Builder 1-2-3, Picture Archiving and Communications System.

- 1.1 Company Identification
ERAD/ ImageMedical Corp.. 824 San Antonio Road Palo Alto CA 94303 Establishment Registration: 2954766 Owner Operator Number: 9042966 Contact: Jim Connors Tel: (650) 813-0400 Fax: (650) 813-0110

- 1.2 Official Correspondent
Gary J. Allsebrook Regulatory Management Services 16303 Panoramic Way San Leandro CA USA 94578-1116 Tel: (510) 276-2648 Fax: (510) 276-3559 Email: regman10@attbi.com

- 1.3 Date of Submission June 1, 2003
- 1.4 Device Name

Classification Name: Common/Usual Name: Proprietary Name:

PACS Soft-copy reading system PracticeBuilder 1-2-3

- ાં ર Substantial Equivalence
PracticeBuilder 1-2-3 software is substantially equivalent to the Stentor, iSite (K013630), Stentor iVault (Class I, Exempt), Agfa IMPAX, Diagnostic Display Station (K922292) and the Ultravisual Visual PACS (Vortex) (K012097) .

{1}------------------------------------------------

#### 1.6 Device Description and Intended Use

PracticeBuilder 1-2-3 is a PACS system, comprised of acquisition components (GatewayServer and SendServer), a central system manager component (SmartServer), a diagnostic workstation component (Workstation and Viewer), and an archiving component (ArchiveServer). The data flow is such that patient and procedure information is optionally delivered to the central system manager, followed by the acquisition of the image objects directly from the image sources or by one of the acquisition components. After receiving the procedure information or after receiving image objects, the central system manager searches for and retrieves relevant prior procedure data from the archive component. When the central system manager registers the acquired image objects and the retrieved prior procedure data, a user can access the information by selecting the item from the operator worklist. The image data is transmitted to and rendered on the user's workstation using the diagnostic workstation components. After using the workstation to view the images, the user optionally dictates a report into the system, after which, a user can play back the dictation and transcribe it to text. Once PracticeBuilder 1-2-3's central system manager registers a report, the report is available for access by the referring physician, or it can be exported into an information system. At some configured point in time, the image data and the report information is delivered to the archiving component for backup and long-term storage.

PracticeBuilder 1-2-3 is also a teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. Practice Builder 1-2-3 is for hospitals, imaging centers, radiologist reading practices and any user who requires and is granted access to patient image, demographic and report information.

#### 1.7 Software Development

ERAD/ImageMedical Corp., certifies that the PracticeBuilder 1-2-3 software is designed, developed, tested and validated according to written procedures. These procedures identify individuals within the organization responsible for developing and approving product specifications, coding and testing, validation testing and field maintenance. The software developed for this product is used to provide diagnostic quality images and associated information to the indented users.

{2}------------------------------------------------

#### 1.8 Safety and Effectiveness

#### General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and indications for use.

The hardware components specified (but not supplied) are all "off the shelf" computer components.

### Validation and Effectiveness:

Extensive testing of the software package has been performed by programmers, by non-programmers, quality control staff, and by potential customers.

### Substantial Equivalence:

PracticeBuilder 1-2-3 has Indications for Use and a Target Population similar to other medical image devices, including Stentor's iSite Radiology (K013630) and Stentor's iVault (Class I Exempt), Agfa's IMPAX (K022292), and Ultravisual's Vortex (K012097). All of the functions PracticeBuilder 1-2-3 performs are available in at least one of the listed substantially equivalent devices. In most cases, the function is available in all of them. There are no significant differences between PracticeBuilder 1-2-3 and the collective functions of all the predicate devices. See section E for additional information.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. Inside the circle is a stylized image of an eagle with three lines extending from its head, resembling feathers or wings. The seal appears to be a logo or emblem, possibly representing a government agency related to health and human services in the United States.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 7 2003

eRad/ImageMedical Corp. % Mr. Gary J. Allsebrook Official Correspondent Regulatory Management Services 16303 Panoramic Way SAN LEANDRO CA 94578

Re: K031778

Trade/Device Name: PracticeBuilder 1-2-3 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: June 1, 2003

Received: June 11, 2003

Dear Mr. Allsebrook:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Ko 3 1778

Page 1 of 1

510(k) Number (if known): K031778

eRAD/ImageMedical Corp., PracticeBuilder 1-2-3, Device Name: Picture Archiving and Communications System

Indications For Use:

PracticeBuilder 1-2-3 is a PACS and teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces.

PracticeBuilder 1-2-3 is for use in hospitals, imaging centers, radiologist reading practices and any user who requires and is granted access to patient image, demographic and report information.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 901.109)

OR

Over-the-Counter Use

Daniel R. Lyson
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices

510(k) Number K031778

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K031778](https://fda.innolitics.com/device/K031778)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
