The subject devices are intended for single use only, and may be used with or without bone cement. The subject devices may be used with an Howmedica Osteonics Endo Head and Neck Extension for hemi-hip arthroplasty, a UHR® Universal Head for bipolar hip replacement, or a Howmedica Osteonics C-Taper cobalt chromium head and Howmedica Osteonics acetabular component for conventional total hip arthroplasty.
Device Story
Omnifit® HFx Hip Stem Series are femoral hip stems for total or bipolar hip arthroplasty. Devices are forged from cobalt chromium alloy; intended for single use; compatible with bone cement or cementless fixation. Used in conjunction with modular components including Endo Head and Neck Extensions, UHR® Universal Heads, or C-Taper heads and acetabular components. Operated by orthopedic surgeons in clinical/hospital settings. Device provides structural replacement for diseased or fractured femoral head/neck; restores joint function; alleviates pain. Strength of neck and body regions validated via Finite Element Analysis.
Clinical Evidence
Bench testing only. Finite Element Analysis was performed to evaluate the strength of the neck and body regions of the hip stems.
Technological Characteristics
Materials: ASTM F-799 cobalt chromium alloy. Design: Femoral hip stem for cemented or uncemented use. Connectivity: None (mechanical implant). Sterilization: Not specified. Software: None.
Indications for Use
Indicated for bipolar hip replacement in patients with femoral head/neck fractures or non-union, aseptic necrosis, or arthritis (osteo-, rheumatoid, post-traumatic) with minimal acetabular involvement; also for patients requiring conservative acetabular procedures or salvage of failed total hip arthroplasty. Indicated for total hip replacement in patients with painful, disabling joint disease (degenerative, rheumatoid, post-traumatic arthritis, or late-stage avascular necrosis), revision of failed previous procedures, clinical management problems where arthrodesis is less satisfactory, or clinical circumstances requiring altered femoral resection due to proximal fracture, bone loss, or calcar lysis.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
AUG - 8 2003
Omnifit® HFx Hip Stcms
י
Image /page/0/Picture/2 description: The image shows handwritten text on a white background. The text includes the alphanumeric string "K031744" on the first line. The second line contains the words "page 10 of 2" with "s10(k)" written below the number 2. The handwriting appears to be somewhat rough and uneven.
# 510(k) Summary: Omnifit® HFx Hip Stem Series
| Proprietary Name: | Omnifit® HFx Hip Stem Series |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Femoral Hip Stems |
| Classification Name and Reference: | Hip joint, metal/polymer semi-constrained,<br>cemented prosthesis<br>21 CFR §888.3350<br>Hip joint femoral (hemi-hip) metallic cemented or<br>uncemented prosthesis<br>21 CFR §888.3360 |
| Proposed Regulatory Class: | Class II |
| Device Product Code: | 87 – JDI: Prosthesis, hip, semi-constrained,<br>cemented<br>87 – LWJ: Prosthesis, hip, semi-constrained,<br>metal/polymer, uncemented |
| Predicate Proprietary Name(s): | Osteonics ODC/ODC Plus Hip Stems, Omnifit X<br>(Eon) Hip Stems |
| Predicate Regulatory Class: | Class II |
| Predicate Product Code(s): | 87 JDI, KWL |
| Submitted By: | Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, New Jersey 07401-1677<br>(201) 825-4900 |
| Contact Information: | Debra Bing<br>Phone: (201) 831-5413<br>Fax: (201) 831-6038 |
| Date Summary Prepared: | July 23, 2003 |
{1}------------------------------------------------
#### Omnifit® HFx Hip Stems
## Description/Technological Comparison
031744
The subject and predicate hip stems are both hip fracture stems intended for cemented use. The subject devices and the predicate 6076 series hip stems (K954598) are forged from ASTM F-799 cobalt chromium alloy.
## Intended Use
The subject devices are intended for single use only, and may be used with or without bone cement. The subject devices may be used with an Howmedica Osteonics Endo Head and Neck Extension for hemi-hip arthroplasty, a UHR® Universal Head for bipolar hip replacement, or a Howmedica Osteonics C-Taper cobalt chromium head and Howmedica Osteonics acetabular component for conventional total hip arthroplasty.
## Indications:
For use as a bipolar hip replacement
- Femoral head/neck fractures or non-union, .
- Asentic necrosis of the femoral head, ●
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acctabular . involvement or distortion,
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum,
- Salvage of failed total hip arthroplasty .
- Femoral Neck Fractures .
### For use as a total hip replacement
- Painful, disabling joint disease resulting from: degenerative arthritis, rheumatoid arthritis, ● post-traumatic arthritis or late stage avascular necrosis,
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty, or other . procedure,
- Clinical management problems where arthrodesis or alternative reconstructive techniques are . less likely to achieve satisfactory results,
- Clinical circumstances that require an altered femoral resection level due to proximal ● fracture, bone loss, or calcar lysis.
## Testing Summary
Finite Element Analysis was used to evaluate the strength of the neck and body regions of the hip stems.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized horizontal lines above a wavy line, resembling a stylized human figure.
Public Health Service
AUG - 8 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Debra Bing Regulatory Affairs Manager Howmedica Osteonics Corp. 59 Route 17 Allendale, New Jersey 07401
Re: K031744 Trade/Device Name: Omnifit® HFx Hip Stem Series Regulation Number: 21 CFR 888.3350 and 21 CFR 888.3360 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis and Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Regulatory Class: II Product Code: JDI, LWJ, and KWL Dated: June 4, 2003 Received: June 5, 2003
Dear Ms. Bing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Debra Bing
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Device Name: Omnifit® HFx Hip Stem Series
The subject devices are intended for single use only, and may be used with or without bone cement. The subject devices may be used with an Howmedica Osteonics Endo Head and Neck Extension for hemi-hip arthroplasty, a UHR® Universal Head for bipolar hip replacement, or a Howmedica Osteonics C-Taper cobalt chromium head and Howmedica Osteonics acetabular component for conventional total hip arthroplasty.
### Indications:
### For use as a bipolar hip replacement
- Femoral head/neck fractures or non-union, .
- Aseptic necrosis of the femoral head, ●
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular ● involvement or distortion,
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum,
- . Salvage of failed total hip arthroplasty
- Femoral Neck Fractures .
### For use as a total hip replacement
- Painful, disabling joint disease resulting from: degenerative arthritis, theumatoid arthritis, . post-traumatic arthritis or late stage avascular necrosis,
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty, or other . procedure,
- Clinical management problems where arthrodesis or alternative reconstructive techniques are ● less likely to achieve satisfactory results.
- Clinical circumstances that require an altered femoral resection level due to proximal fracture, . bone loss, or calcar lysis.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|------------------|--|
|------------------|--|
OR
| Over-The-Counter Use | | (Per 21 CFR 801.109) |
|----------------------|--|----------------------|
|----------------------|--|----------------------|
(Optional Format 1-2-96)
*Miriam C. Provost*
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K031744 |
|---------------|---------|
|---------------|---------|
TOTAL P.21
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.