← Product Code [DQA](/productcode/DQA) · K031740

# VSM - VITAL SIGNS MONITOR, MODEL 53000 SERIES (K031740)

_Welch Allyn Protocol, Inc. · DQA · Aug 26, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K031740

## Device Facts

- **Applicant:** Welch Allyn Protocol, Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Aug 26, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The model 53000 Series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients. The most likely locations for patients to be monitored are general med/surg. floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.

## Device Story

Vital Signs Monitor 53000 Series provides multi-parameter patient monitoring; inputs include noninvasive blood pressure (NIBP), pulse rate, SpO2, and body temperature. Device integrates established measurement modules; temperature module is identical to SureTemp Plus. Used by clinicians/medically qualified personnel in general hospital/med/surg floors. Output displayed for real-time clinical assessment; assists providers in patient status evaluation. Benefits include consolidated vital sign tracking for neonatal, pediatric, and adult populations.

## Clinical Evidence

Bench testing only. Performance verified per Test Plan/Report #831-0719-00 and Clinical Validation Report #831-0752-00 using production-equivalent units. Risk analysis and mitigation documented per product development procedures. Quality system conforms to 21CFR820, ISO 9001, EN 46001, and ISO 13485.

## Technological Characteristics

Multi-parameter monitor for NIBP, SpO2, pulse, and temperature. Integrates SureTemp Plus temperature module. Connectivity and specific software architecture not detailed. Quality system certified to ISO 9001, EN 46001, ISO 13485.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Welch Allyn CVSM vital signs monitors model 52STP ([K951193](/device/K951193.md))
- Protocol Systems Propaq Encore 200 Series ([K951246](/device/K951246.md))
- Nellcor Puritan Bennett Pulse Oximeter module MP506 ([K021090](/device/K021090.md))
- Welch Allyn SureTemp Plus ([K030580](/device/K030580.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 26 2003

K031740

i i

## 510(k) Summary of Safety and Effectiveness

May 30, 2003

#### Submitter

Welch Allyn Protocol. Inc. 8500 S.W. Creekside Place Beaverton, OR 97008-7107 USA

Telephone: (503) 530-7500 Fax: (503) 526-4901

Contact: Mr. James W. Sandberg, Director of Regulatory Affairs & Compliance Engineering

#### Device Name

Vital Signs Monitor, Model 53000 Series Trade Name: Common Name: Vital Signs Measurement Device

Classification Name: Noninvasive Blood Pressure Measurement System, 21CFR870.1130, April 1.2002. The VSM model 53000 Series also offers a Pulse Oximetry channel (SpO2) - reference 21CFR870-2700, April 1, 2002, and a patient temperature channel, Clinical electronic thermometer, reference 21CFR880.2910, April 1, 2002.

Classification: Class II

#### Predicate Devices

The predicate devices for the model 53000 Series monitors are the

- Welch Allyn CVSM vital signs monitors model 52STP, K951193, and .
- Protocol Systems Propaq Encore 200 Series, K951246, and .
- Nellcor Puritan Bennett Pulse Oximeter module MP506 cleared with Pulse Oximeter . Model N-550 K021090, and
- Welch Allyn SureTemp™ Plus, K030580.

#### Device Description

The model 53000 Series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpOz), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.

The most likely locations for patients to be monitored are general med/surg. floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.

{1}------------------------------------------------

#### Indications for Use

The model 53000 Series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO₂), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.

The most likely locations for patients to be monitored are general med/surg. floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.

#### Technological Comparison to the Predicate Devices

It is Welch Allyn Protocol's conclusion that the vital signs monitors model 53000 Series are substantially equivalent to the Welch Allyn Tycos vital signs monitor CVSM Model 52SP-E1, the Protocol Systems Propaq Encore 200 Series monitors, the Nellcor Puritan Bennett Pulse Oximeter, Model OxiMAX N-550, and the Welch Allyn Clinical electronic thermometer, Model SureTemp™ Plus.

- . The noninvasive blood pressure measurement specifications and performance are equivalent to the Welch Allyn Tycos CVSM monitor and the Protocol Systems Propaq Model 200 Series monitors, and
- The pulse rate specifications and performance derived from either noninvasive blood . pressure or SpO2 are equivalent to the Welch Allyn Tycos CVSM monitor, the Protocol Systems Propaq, Model 200 Series monitors, and the Nellcor Puritan Bennett Pulse Oximeter Model N-550, and
- . The Sp0, specifications and performance are equivalent to the Nellcor Puritan Bennett Model N-550 Pulse Oximeter, and
- . The Welch Allyn Clinical temperature meter SureTemp™ Plus is the same temperature module used in the Welch Allyn model 53000 Series monitors.

#### Summary of Performance Testing

The Welch Allyn vital signs monitors, model 53000 Series will be tested in accordance with the Test Plan / Report, #831-0719-00, and Clinical Validation Report #831-0752-00 included with the submission using production equivalent units prior to release to market.

A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Welch Allyn Protocol's product development procedure. Welch Allyn Protocol's Quality System conforms to 21CFR820 and is certified by TÜV Product Service to ISO 9001, EN 46001 and ISO 13485.

#### Conclusions

As stated above, Welch Allyn Protocol's conclusion is that the vital signs monitor Model 53000 Series are safe and effective and comply with the appropriate medical device standards and are substantially equivalent to the earlier identified predicate devices.

#### This 510(k) Summary of Safety and Effectiveness may be copied and submitted to interested parties as required by 21CFR807.92.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with text around the perimeter. Inside the circle is a stylized image of an eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 6 2003

Welch Allyn Protocol, Inc. c/o Mr. James W. Sandberg Director of Regulatory Affairs & Compliance Engineering 8500 S.W. Creekside Place Beaverton, OR 97008-7107

Re: K031740

Trade Name: VSM Vital Signs Monitor, Model 53000 Series Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (two) Product Code: DOA Dated: May 30, 2003 Received: June 4, 2003

Dear Mr. Sandberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. James W. Sandberg

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bram D. Luckerman, M.D. Director Division of Cardiovascular Device Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 3/740

# INDICATIONS FOR USE

### Applicant:

Welch Allyn Protocol, Inc. 8500 S.W. Creekside Place Beaverton, OR 97008-7107 USA

Telephone: (503) 530-7500 Fax: (503) 526-4901

510(k) Number:

Device Name: Vital signs monitors, model 53000 Series

#### Indications for Use:

The VSM series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature. noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.

The most likely locations for patients to be monitored are general med/surg, floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | ✓ |
|----------------------|---|
| OR Over-The-Counter  |   |
| (Per 21 CFR 801.109) |   |

(Division Sign-Off)
Division of Cardiovascul.

| 510(k) Number | K031740 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/device/K031740](https://fda.innolitics.com/device/K031740)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
