SMITHS MEDICAL INSULIN INFUSION PUMP WITH COZMONITOR GLUCOSE METER
K031738 · Smiths Medical MD, Inc. · LZG · May 27, 2004 · General Hospital
Device Facts
Record ID
K031738
Device Name
SMITHS MEDICAL INSULIN INFUSION PUMP WITH COZMONITOR GLUCOSE METER
Applicant
Smiths Medical MD, Inc.
Product Code
LZG · General Hospital
Decision Date
May 27, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Deltcc Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes. The Deltec CoZmonitor™ Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyle™, blood sampling will be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results. The Deltec CoZmonitor™ Blood Glucose Monitoring System is designed to be used only with the Deltec Cozmo® Insulin Pump to help you and your healthcare provider monitor your diabetes management program. When used properly, the monitor provides a quantitative measurement of glucose (or.sugar) in whole blood. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyle™ Test Strip and FreeStyle™Cry (100 Control Solution, as other products can give inaccurate results.
Device Story
System integrates CoZmonitor™ glucose meter module with Deltec Cozmo® insulin pump; enables automated blood glucose data transfer via infrared port. User performs finger/alternate site blood sampling using FreeStyle™ test strips; monitor measures glucose; data transmits to pump. Pump stores glucose results with insulin delivery history; allows direct access to bolus menu for correction based on glucose readings. Used by patients for diabetes management; facilitates proactive insulin dosing; reduces manual transcription errors. Healthcare providers review stored data to monitor therapy. Benefits include improved consistency in insulin delivery and streamlined diabetes management.
Clinical Evidence
No clinical efficacy studies performed. Usability study conducted with 57 participants (ages 9-76) using the integrated system. 100% of participants successfully obtained glucose readings within two attempts; 96.5% (55/57) on first attempt. 91% (52/57) obtained correct readings for low, middle, and high control samples. 96-98% of participants reported ease of attachment, ease of use, and ease of interpreting screen messages for corrective actions.
Technological Characteristics
Syringe infusion pump with integrated glucose monitor module. Features infrared communication port, mechanical attachment interface, and CR2032 battery. Software-controlled programming for insulin delivery and glucose data logging. In-vitro diagnostic sensing using FreeStyle™ test strips.
Indications for Use
Indicated for patients requiring Continuous Subcutaneous Insulin Infusion (CSII) for diabetes control and blood glucose monitoring. Suitable for use on finger, forearm, upper arm, thigh, calf, or hand. In-vitro diagnostic use only.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Deltec Cozmo® Insulin Infusion Pump
TheraSense FreeStyle Tracker™ Diabetes Management System
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 27 2004
K031738
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## Deltec Cozmo™ Insulin Infusion Pump with CoZmonitor™ Glucose Monitor
### April 9, 2004
#### I. GENERAL INFORMATION
| Applicant's Name and Address: | Smiths Medical MD, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
|--------------------------------|--------------------------------------------------------------------------------------------------|
| Contact Person: | David H. Short<br>Director Regulatory and Clinical Affairs |
| Common/Usual Name: | Insulin Infusion Pump and Glucose Monitor |
| Proprietary Name: | Deltec Cozmo® Insulin Infusion Pump with CoZmonitor™ Glucose Monitor |
| Equivalence Device Comparison: | Deltec Cozmo® Insulin Infusion Pump and TheraSense FreeStyle Tracker™ Diabetes Management System |
#### II. DEVICE DESCRIPTION
A CoZmonitor™ Glucose Monitor is a modified FreeStyle Tracker™ meter module designed to be coupled with the Deltec Cozmo® Insulin Pump. The Deltec Cozmo® Insulin Infusion Pump with CoZmonitor™ Glucose Monitor is lightweight and easy to use. Users follow their usual practice of blood sampling. The FreeStyle™ test strip is inserted into the CoZmonitor™ Glucose Monitor, which is attached to the Deltec Cozmo® Insulin Pump.
The CoZmonitor™ Glucose Monitor will automatically transfer the blood glucose information to the Deltec Cozmo® Pump through an infrared port. If the user's blood glucose level is higher than normal, he or she can go directly to the bolus menu of the pump to correct it without having to manually re-enter the blood glucose value, ensuring more consistent insulin delivery.
In addition, blood glucose results are combined with insulin delivery data and stored in the Deltec Cozmo® Pump's history log to provide a complete picture of the user's diabetes management, allowing them to be more proactive in their insulin decisions. The system's central data storage capabilities also allow users to go through fewer steps and therapy calculations – reducing the risk of data or transcription errors.
{1}------------------------------------------------
### III. INTENDED USE OF THE DEVICE
The Deltcc Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.
The Deltec CoZmonitor™ Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyle™, blood sampling will be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results.
The Deltec CoZmonitor™ Blood Glucose Monitoring System is designed to be used only with the Deltec Cozmo® Insulin Pump to help you and your healthcare provider monitor your diabetes management program. When used properly, the monitor provides a quantitative measurement of glucose (or.sugar) in whole blood. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyle™ Test Strip and FreeStyle™Cry (100 Control Solution, as other products can give inaccurate results.
#### DEVICE COMPARISON IV.
The Deltec Cozmo® Insulin Pump is very similar to the currently distributed pump, with the exception of updated software to allow for its use with the CoZmonitor™ Glucosc Monitor.
The CoZmonitor™ Glucose Monitor is a modified FreeStyle Tracker™ meter module designed to be coupled with the Deltec Cozmo® Insulin Pump.
The FreeStyle Tracker™ meter modified to include the mechanical connections to allow for the connection to a Deltec Cozmo® Insulin Pump along with designing the module to fit on the back of the pump, updated software to allow for use with the pump, infrared ports to allow for commis cations with the pump, and the use of a CR2032 battery
### V. SUMMARY OF STUDIES
#### A. Functional Testing
Software validation, verification of software controlled programming functions, and software related to proper pump operation were completed for the Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor.
#### B. Clinical Studies
Clinical studies were not deemed necessary regarding the Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor due to their similarity in design and function to the current Deltec Cozmo™ Insulin Pump and TheraSense FreeStyle Tracker™ monitor.
Usability studies of the Deltec CoZmonitor™ Blood Glucose Monitoring System were conducted with 57 participants (23 females and 32 males) ranging in age from 9 years to 76 years. Two participants did not provide their age. All participants were Deltec Cozmo® Insulin Pump users
{2}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Deltec Cozmo™ Insulin Infusion Pump with CoZmonitor™ Grucose Monitor April 9, 2004
and ranged from 1 month to >12 months. Five participants did not provide the length of time they had used the Deltec Cozmo® Insulin Pump.
Participants were asked to obtain their blood glucose reading and 57 of 57 participants were able to successfully obtain their blood glucose reading within two attempts, 55 of 57 partics partics parts successfully obtain blood glucose readings with their first attempt.
Participants were also provided three control samples with low, middle, and high glucose concentrations and were asked to obtain glucose readings from each sample and 52 of 57 participants were able to obtain correct blood glucose readings using the low, middle and high control samples. Of the 5 participants that did not obtain correct glucose readings, I participant obtained an out of range reading for the high control on test and retest; one participant abtained an out of range reading for a middle control, but did not retest; two participants did not sample 1 of the 3 control samples, but were able to obtain correct readings for the other two control samples,
Participants were asked to attach the CoZmonitor™ Glucose Monitor to the Deltec Cozmo® Insulin Pump and 55 of 57 participants indicated the CoZmonitor™ Glucose Monitor was easy to attach to the Deltec Cozmo® Insulin Pump. 55 of 57 participants indicated the CoZmonitori was Casy Glucose Monitor was easy to use. 55 of 57 participants indicated the CoZmonitor™ Glucose Monitor System was easy for them to know what to do based on the messages displayed on the Deltec Cozmo® Insulin Pump screen. 56 of 57 participants indicated that it was easy for them to select the appropriate choice from Deltec Cozmo® Insulin Pump menu options to bring their blood glucose reading back on target. 56 of 57 participants indicated that they successfully used the CoZmonitor™ Glucose Monitor.
#### C. Conclusion Drawn from the Studies
Based upon the information provided above, the Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor is substantially equivalent to other commercially available devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three wavy lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Public Health Service
## MAY 2 7 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David H. Short Director Regulatory and Clinical Affairs Smiths Medical MD, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112
Re: K031738
Trade/Device Name: Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: March 16, 2004 Received: March 17, 2004
Dear Mr. Short:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Mr. Short
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
### 510(k) Number (if known): K031738
Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor Device Name:
### Indications for Use:
## Deltec Cozmo® Insulin Pump
The Deltec Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.
| Prescription Use X | OR | Over-The Counter Use |
|----------------------|----|----------------------|
| (Per 21 CFR 801.109) | | |
### Indications for Use:
### CoZmonitor™ Glucose Monitor
The Deltec CoZmonitor™ Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results.
Prescription Use (Per 21 CFR 801.109) ાર
Over-The Counter Use X
### Indications for Use:
### Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor
The Deltec CoZmonitor™ Blood Glucose Monitoring System is designed to be used only with the Deltec Cozmo® Insulin Pump to help you and your healthcare provider monitor your diabetes management program. When used properly, the monitor provides a quantitative measurement of glucose (or sugar) in whole blood. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results.
Prescription Use X (Per 21 CFR 801.109)
()R
Over-The Counter Use ----------
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
.
Concurrence of CDRH, Office of Device Evaluation (ODE)
29. May
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number:_KG31738
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.