K031693 · Biomet, Inc. · LPH · Aug 6, 2003 · Orthopedic
Device Facts
Record ID
K031693
Device Name
MALLORY/HEAD MODULAR CALCAR WITH AND WITHOUT HA
Applicant
Biomet, Inc.
Product Code
LPH · Orthopedic
Decision Date
Aug 6, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cemented and non-cemented total joint replacement in cases of: 1) Noninflamatory degenerative joint disease including osteoarthritis and avascular necrosis 2) Rheumatoid arthritis 3) Correction of functional deformity 4) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5) Revision of previously failed total hip arthroplasty
Device Story
Modular femoral hip prosthesis; consists of proximal metaphyseal segment and distal stem. Surgeon selects components intraoperatively to construct appropriate size; components joined by Morse locking taper and secured with locking screw. Proximal segment replaces proximal femur in severe bone loss cases; features porous coating for fixation; includes options for trochanteric plate/bolt or claw for greater trochanter compression. Distal stems available in various lengths and porous-coated or splined configurations. Used in orthopedic surgery for total hip replacement; provides structural support and fixation for femoral component. Benefits patient by restoring hip function and stability in cases of joint disease or fracture.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including engineering analysis, finite element analysis, and mechanical testing.
Technological Characteristics
Modular femoral hip prosthesis; proximal metaphyseal segment and distal stem; Morse locking taper connection with secondary locking screw; porous-coated and splined configurations; various lengths and widths; non-cemented or cemented application.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, femoral neck, trochanteric) with head involvement, or revision of failed total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
AUG - 6 2003
Image /page/0/Picture/1 description: The image shows the logo for Biomet Orthopedics, Inc. The word "BIOMET" is in large, bold, block letters on the top line. The words "ORTHOPEDICS, INC" are in smaller letters on the second line, centered below the first word.
page 1 of 2
K031693
ORTHOPEDICS, INC.
## Summary of Safety and Effectiveness
Applicarit/Sponsor:Biomet Orthopedics, Inc.
Contact Person: Patricia Sandborn Beres Senior Regulatory Specialist
Proprietary Name: Mallory-Head® Modular Calcar Femoral Component with and without HF
Cc mmon Name: Total hip femoral component
Classification Name: Hip joint metal/polymer/metal semi-constrained, porous coated, uncemented prosthesis (21 CFR 888.3358)
Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Miallory-Head Modular Calcar Replacement Components, 510(k) K945115 Mallory-Head Modular Calcar Total Hip with Roller- Hardened/Machined Tapers, 510(k) KU00335
Tr e Reach® Femoral Component, 510(k) K000760
Mallory-Head Modular Calcar (100% porous stems), 510(k) K001660 The HA Modular Reach® Femoral Component, 510(k) K022463
Device Description: : Each stem consists of two separate parts, a proximal metaphyseal segment and a distal stem. A variety of stem lengths and proximal segment sizes allow the surgeon to "construct" the proper sized implant while in surgery. All pieces are in erchangeable. The device utilizes a modular femoral head taper fit on to the stem at the time of surgery.
T re proximal segment is designed to replace the proximal portion of the femur in cases of suvere bone loss. The device is porous coated. The proximal components are available ir 5 resection levels and 7 widths.
A hole through the proximal component gives the surgeon the option of using a trochanter p ate and trochanter bolt or a trochanter claw. This assembly will allow the greater tr ochanter to be compressed against the prosthesis for enhanced fixation and proximal s :ability.
> MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587
> > 1
SHIPPING ADDRESS Airport Industrial Park Warsaw, IN 46580
1
OFFICE 219.267.6639
I·AX 219.267.8137
E-MAIL וואסט.comet@biomet.con
{1}------------------------------------------------
page 2 of 2
K031693
Dis al stems are available in six lengths, 165mm, 200mm, 240mm, 250mm and 30( mm. Stems are available with 40%, 80% or 100% of the stem porous coated, as well as a non-coated, splined configuration, Since the stems are modular, left and right cor ponents are not necessary.
Th‹› proximal and distal segments are joined by means of a Morse locking taper similar to tha: used to attach the modular head components to most hip stems. Additional fixation is ach ieved through a locking screw inserted through the driving platform on the metaphyseal cornponent and engaging with the stem taper.
Int anded Use: Non-cemented application in cases of:
- Noninflamatory degenerative joint disease including osteoarthritis and avascular 1) necrosis
- 2) Fheumatoid arthritis
- Correction of functional deformity 3)
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the 4) proximal femur with head involvement, unmanageable by other techniques.
- Flevision of previously failed total hip arthroplasty 5)
Su mmary of Technologies: The materials, design and processing of the devices are identical to or similar to the predicates.
N‹››Clinical Testing: Engineering analysis, finite Element Analysis and mechanical te: ting were provided.
Clinical Testing: None provided .
All Tractemarks are property of Biomet, Inc.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/12 description: The image is a black and white seal. The seal is circular and contains text around the perimeter. The text reads "Department of Health & Human Services USA". In the center of the seal is a stylized symbol that resembles three abstract human profiles or faces.
AUG - 6 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Patricia Sandborn Beres Senior Regulatory Specialist Biomet Orthopedics, Inc. P.O. Box 587 Warsaw, IN 46581-0587
Re: K031693
Trade/Device Name: Mallory-Head Modular Calcar with and without HA Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint/metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH Dated: July 18, 2003 Received: July 22, 2003
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Patricia Sandborn Beres
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510[k) Number (if known): K031693
Device Name: Mallory/Head Modular Calcar with and without HA
Indications For Use:
Cemented and non-cemented total joint replacement in cases of:
- 1) Noninflamatory degenerative joint disease including osteoarthritis and avascular necrosis
- 2) Rheumatoid arthritis
- 3) Correction of functional deformity
- 4) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- 5) Revision of previously failed total hip arthroplasty
iriam C. Provost
on of General. Restorative and Neurological Devices
510(k) Number K031693
(I'LEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
f 'rescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Formal 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.