← Product Code [IYN](/productcode/IYN) · K031663

# GE VIVID 7, MODEL FC0003XX (K031663)

_General Electric Co. · IYN · Jun 9, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K031663

## Device Facts

- **Applicant:** General Electric Co.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jun 9, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic: Adult Cephalic: Cardiac (adult and pediatric): Peripheral Vascular (PV): Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).

## Device Story

GE Vivid 7 Diagnostic Ultrasound System is a mobile echocardiography imaging and analysis system; console includes digital acquisition, processing, and display hardware; user interface features keyboard, specialized controls, and color CRT display. Device processes ultrasonic pulsed echo and Doppler signals to generate diagnostic images. Used by qualified physicians in clinical settings for evaluation of anatomy and fluid flow. Modification adds Tissue Synchronization Imaging (TSI) to enhance wall motion evaluation during echocardiography. Output displayed on screen for physician interpretation to support clinical decision-making and patient diagnosis.

## Clinical Evidence

Bench testing only. Device evaluated for acoustic output, biocompatibility, cleaning/disinfection, electromagnetic compatibility, and thermal/electrical/mechanical safety. Conforms to applicable medical device safety standards. No clinical data required.

## Technological Characteristics

Mobile console (65cm deep, 139cm high); digital acquisition/processing/display; CRT monitor. Supports multiple transducers (e.g., 3.5C, 8C, M7C, 7L, 10L, 12L, MI2L, E8C, 6T, 7T, 8T, 9T, P2D, P6D, 18L, i8L, i13L). Modes: B, M, PW/CW Doppler, Color Doppler, Color M-Doppler, Power Doppler, Combined modes, Harmonic Imaging, Coded Pulse. Quality systems: 21 CFR 820, ISO 9001:2000, ISO 13485.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE Vivid 7 Diagnostic Ultrasound System ([K003931](/device/K003931.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KQBI663

**JUN - 9 2003**

# Attachment B:

Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).

Image /page/0/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular shape. The letters and the circle are black, creating a strong contrast. The logo is a well-known symbol associated with the General Electric company.

GE Medical Systems

General Electric Company P.O. Box 414, Milwaukee, WI 53201

#### Section a):

| 1. | Submitter:       | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201                                                                                                                                 |
|----|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Contact Person:  | Allen Schuh,<br>Manager, Safety and Regulatory Engineering<br>Telephone: 414-647-4385; Fax: 414-647-4090                                                                                |
|    | Date Prepared:   | May 27, 2003                                                                                                                                                                            |
| 2. | Device Name:     | GE Vivid 7 Diagnostic Ultrasound System with TSI<br>Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN |
| 3. | Marketed Device: | GE Vivid 7 Diagnostic Ultrasound System K003931 (90-IYO/IYN)                                                                                                                            |

A device currently in commercial distribution.

4. Device Description: The GE Vivid 7 Diagnostic Ultrasound is a full featured echocardiography imaging and analysis system. It consists of a mobile console approximately 65 cm deep and 139 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls and a color video CRT display. This modification will provide users with enhanced ability to evaluate wall motion during echocardiography.

5. Indications for Use: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic: Adult Cephalic: Cardiac (adult and pediatric): Peripheral Vascular (PV): Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).

6. Comparison with Predicate Device: The GE Vivid 7 with TSI is of a comparable type and substantially equivalent to the current GE Vivid 7. It has the same technological characteristics, key safety and effectiveness features, physical design, construction, and has the same intended uses and basic operating modes as the predicate device.

#### Section b):

1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, electromagnetic compatibility, as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.

#### 2. Clinical Tests: None required.

3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001:2000 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE Vivid 7 Diagnostic Ultrasound with TSI is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 9 2003

Mr. Allen Schuh Manager, GE Ultrasound Safety and Regulatory Engineering GE Medical Systems P.O. Box 414 MILWAUKEE WI 53201

Re: K031663

Trade Name: GE Vivid 7 with TSI Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasound pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: May 27, 2003 Received: May 29, 2003

#### Dear Mr. Schuh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the GE Vivid 7 with TSI, as described in your premarket notification:

#### Transducer Model Number

3.5C રુ 8C

{2}------------------------------------------------

M7C 7L 10L 12L MI2L 35 રકે 78 102 M38 E8C 6T 7T 8T 9T P2D P6D 18L ! 131

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information. including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

{3}------------------------------------------------

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

Carolyn 4. Newland

ಕ್ಕೆ Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE Vivid 7 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                 | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Small Organ[2]                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |  |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Transrectal                                        | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |  |
| Transvaginal                                       | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative[5]                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] Other use includes Urology/Prostate

[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CORH, Office of Device Evaluation (ODE)

David A. Seaman
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Device

510(k) Number

{5}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE Vivid 7 with 3.5C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |  |
| Abdominal(1)                                       | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |  |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other[4]                                           | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Ingram

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Det

Prescription User (Per 21 CFR 801.109)

510(k) Number.

{6}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form

## GE Vivid 7 with 5C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |       |  |
| Abdominal[1]                                       | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |       |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |       |  |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |       |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic, Renal and Aorta-iliac artery;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of GORH, Office of Device Evaluation (ODE)

David A. Slayton

(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devi

s10(k) Number K031663

{7}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE Vivid 7 with 8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal [1]                                      | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Pediatric                                          | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Small Organ (specify)                              | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Neonatal Cephalic                                  | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other [4]                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic, Renal and Aorta-iliac artery;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Logman

(Division Sign-Off and Radi

510(k) Number

{8}------------------------------------------------

# GE Vivid 7 with M7C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Abdominal[1]                                       | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Pediatric                                          | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Small Organ (specify)                              | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic, Renal and Aorta-iliac artery;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Sycamore

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device

510(k) Number_

{9}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE Vivid 7 with 7L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Lyman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices

510(k) Number K031663

{10}------------------------------------------------

## GE Vivid 7 with 10L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical<br>Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                             | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Small Organ[2]                                        | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Neonatal Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                   | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Musculo-skeletal Conventional                         | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Musculo-skeletal Superficial                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other (specify)                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative[5]                                     | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Intraoperative Neurological                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Lymm

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devic

510(k) Number KU314

{11}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE Vivid 7 with 12L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                 |                     | P              |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other (specify)                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONȚINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CORH, Office of Device Evaluation (ODE)

David R. Liggerm

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi

510(k) Number K031663

{12}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

# GE Vivid 7 with M12L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                          | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |
| Small Organ [2]                                    | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |
| Musculo-skeletal Conventional                      | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other (specify)                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                           | E                 | E | E             |               | E                | E                  | E                | E                 |                     | E              |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CPRH, Office of Device Evaluation (ODE)

David A. Heymann

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{13}------------------------------------------------

,

#### Diagnostic Ultrasound Indications for Use Form

# GE Vivid 7 with 3S Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    |   | Mode of Operation |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|---|-------------------|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M                 | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Abdominal[1]                                       | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                                          | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Small Organ (specify)                              |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                  |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Cardiac[3]                                         | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Peripheral Vascular                                |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                      |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                       |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                           | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Exam Type, Means of Access                         |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                           |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                        |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |   |                   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |   |                   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[3] Cardiac is Adult and Pediatric;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

![Signature](signature.png)

| (Division Sign-Off)                  |
|--------------------------------------|
| Division of Reproductive, Abdominal, |
| and Radiological Devices             |

510(k) Number_________________________________________________________________________________________________________________________________________________________________

{14}------------------------------------------------

## GE Vivid 7 with 5S Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[3] Cardiac is Adult and Pediatric;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|                                      | <img alt="signature" src=""/> |
|--------------------------------------|-------------------------------|
| (Division Sign-Off)                  |                               |
| Division of Reproductive,…

---

**Source:** [https://fda.innolitics.com/device/K031663](https://fda.innolitics.com/device/K031663)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
