This instrument has been designed to be used with an OLYMPUS video system center, light source, documentation equipment, display monitor, strobe light source specified by OLYMPUS, and ancillary equipment for endoscopic diagnosis within the nasal lumens and airway anatomy.
Device Story
Flexible rhino-laryngovideoscope; used for endoscopic diagnosis of nasal lumens and airway anatomy. Device connects to Olympus video system center, display monitor, and specified strobe light sources (e.g., Kay Elemetrics 9100 B/C). Operated by clinicians in clinical settings. Provides visual feedback to the physician to assist in diagnostic procedures. No changes to design, materials, or method of operation compared to predicate; expansion of intended use to include compatibility with specific strobe light sources.
Clinical Evidence
Bench testing only. No clinical data required as the device design, materials, and method of operation are identical to the predicate device.
Technological Characteristics
Flexible rhino-laryngovideoscope. Patient-contacting materials confirmed per ISO 10993-1. Meets IEC60601-1, IEC60601-1-1, IEC60601-1-2, and IEC60601-2-18 safety standards. Compatible with Olympus video system centers and specified strobe light sources.
Indications for Use
Indicated for endoscopic diagnosis within the nasal lumens and airway anatomy in patients requiring visualization of these areas.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
VISERA Rhino-Laryngovideoscope Olympus ENF type V (K021073)
EVIS EXERA Bronchovideoscope Olympus BF type 3C160 (K023984)
Submission Summary (Full Text)
{0}------------------------------------------------
# JUL 2 4 2003
# SMDA 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.
## A. GENERAL INFORMATION
### 1. Applicant
Olympus Optical Co., Ltd. 34-3 Hirai Hinode-machi, Nishitama-gun, Tokyo, 190-0182 Japan (Registration Number: 3003637092)
### 2. Submission Correspondence
Olympus Optical Co., Ltd. 2951 Ishikawa-cho Hachioji-shi, Tokyo 192-8507, JAPAN TEL: +81-426-42-5129 FAX: +81-426-42-9979 (Registration Number: 8010047)
#### 3. Official Correspondent
Tina Steffanie-Oak, Senior R.A. Analyst Olympus America Inc. Two Corporate Center Drive, Melville, NY 11747-3157 TEL: 631-844-5477 FAX: 631-844-5416 (Registration Number: 2429304)
#### B. DEVICE IDENTIFICATION
#### 1. Common/Usual Name
RHINO-LARYNGOVIDEOSCOPE, its accessories and ancillary equipment
#### 2. Device Name
VISERA RHINO-LARYNGOVIDEOSCOPE OLYMPUS ENF TYPE V, its accessories and ancillary equipment
#### 3. Classification Name
| CFR Number | Classification Name | Class | Product Code |
|------------|-------------------------------------------------|-------|--------------|
| 874.4760 | Nasopharyngoscope (flexible) and<br>accessories | II | EOB |
{1}------------------------------------------------
# C. IDENTIFICATION OF LEGALLY MARKETED DEVICES WHICH WE CLAIM SUBSTANTIAL EQUIVALENCE
| Model | 510(k)# | Manufacturer | Class | Product<br>Code |
|-------------------------------------------------------|----------|---------------------------|-------|-----------------|
| VISERA Rhino-Laryngovideoscope<br>Olympus ENF type V | #K021073 | Olympus Optical Co., Ltd. | II | EOB |
| EVIS EXERA Bronchovideoscope<br>Olympus BF type 3C160 | #K023984 | Olympus Optical Co., Ltd. | II | EOQ |
### D. DEVICE DESCRIPTION
#### 1. Summarv
The subject device, the VISERA Rhino-Larvngovideoscope Olympus ENF type V is identical to the predicate device cleared in #K021073. This premarket notification is submitted to expand the intended use to add compatibility with strobe light sources specified by Olympus and change the applications from the nasal and nasopharyngeal lumens and airway and airway anatomy. Olympus specifies the Light Source Model 9100 B and 9100 C (#K921184) manufactured by Kay Elemetrics corp., for endoscopic diagnosis using the strobe light source.
### 2. Design
The VISERA Rhino-Larvngovideoscope Olympus ENF type V has been designed, manufactured and tested in compliance with voluntary safety standards. It meets the requirement of IEC60601-1. IEC60601-1-1, IEC60601-1-2 and IEC60601-2-18.
#### 3. Materials
All the patient contacting materials used in this endoscope and ancillary equipment are identical to those used in the devices cleared in the past 510(k) submissions. All materials have been confirmed with ISO 10993-1.
#### 4. Intended Use of the device
This instrument has been designed to be used with an OLYMPUS video system center, light source, documentation equipment, display monitor, strobe light source specified by OLYMPUS, and ancillary equipment for endoscopic diagnosis within the nasal lumens and airway anatomy.
#### 5. Summary including conclusion drawn form Non-clinical Tests
When compared to the predicate device, the ENF-V does not incorporate any significant changes in the intended use, method of operation, material, or design that could affect safety and effectiveness. Therefore clinical data is not necessary for evaluation of safety and efficacy.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them.
JUL 2 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Olympus Optical Co., Ltd. c/o Tina Steffanie-Oak, Senior R.A. Analyst Olympus America Inc. Two Corporate Center Dr. Melville, NY 11747
Re: K031648
Trade/Device Name: Visera Rhino-Laryngovideoscope Olympus ENF Type V Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: May 23, 2003 Received: May 28, 2003
Dear Ms. Steffanie-Oak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Tina Steffanie-Oak, Senior R.A. Analyst
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A halpi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): Not assigned yet_K 0 3 1 k 4 8 Device Name:
Indications for Use:
This instrument has been designed to be used with an OLYMPUS video system center, light source, documentation equipment, display monitor, strobe light source specified by OLYMPUS, and ancillary equipment for endoscopic diagnosis within the nasal lumens and airway anatomy.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ (Per 21 CFR 801.109) ·
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Keun A. Boles
(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devi
510(k) Number K031658
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.