EMIT II PLUS COCAINE METABOLITE ASSAY
K031512 · Dade Behring, Inc. · DIO · Jul 23, 2003 · Clinical Toxicology
Device Facts
| Record ID | K031512 |
| Device Name | EMIT II PLUS COCAINE METABOLITE ASSAY |
| Applicant | Dade Behring, Inc. |
| Product Code | DIO · Clinical Toxicology |
| Decision Date | Jul 23, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3250 |
| Device Class | Class 2 |
Intended Use
The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semi-quantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
Device Story
Homogeneous enzyme immunoassay for detection of benzoylecgonine in human urine; utilizes chemistry analyzers for processing; provides qualitative and semi-quantitative results; intended for preliminary screening; requires confirmatory testing via GC/MS; used in clinical or laboratory settings by trained personnel; assists healthcare providers in identifying cocaine metabolite presence; supports clinical decision-making regarding drug-of-abuse testing.
Clinical Evidence
Method comparison study performed against GC/MS reference method. At 150 ng/mL cutoff, 93% agreement observed. At 300 ng/mL cutoff, 94% agreement observed. Data demonstrates performance consistency with predicate.
Technological Characteristics
Homogeneous enzyme immunoassay; utilizes chemistry analyzers; qualitative and semi-quantitative detection of benzoylecgonine; reagent-based system.
Indications for Use
Indicated for qualitative and semi-quantitative analysis of benzoylecgonine (cocaine metabolite) in human urine. Intended for use with chemistry analyzers as a preliminary analytical test. Requires confirmatory testing via GC/MS or other chemical methods.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Emit® II Plus Cocaine Metabolite Assay (K993988)
Related Devices
- K062929 — QUEST DIAGNOSTICS URINE COCAINE METABOLITE EIA · Quest Diagnostics, Inc. · Dec 18, 2006
- K020441 — EMIT II PLUS MONOCLONAL COCAINE METABOLITE ASSAY · Dade Behring, Inc. · Mar 28, 2002
- K013004 — COCAINE · Abbott Laboratories · Mar 13, 2002
- K020763 — COCAINE METABOLITE ENZYME IMMUNOASSAY MODELS #0030 (500 TESTS KIT), #0031 (5000 TESTS KIT) · Lin-Zhi International, Inc. · May 10, 2002
- K191638 — Pointe Scientific Cocaine Metabolite Enzyme Immunoassay · Medtest DX · Mar 12, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Syva Company - Dade Behring Inc. 510(K) Notification - K031512 Additional Information
JUL 2 8 2003
### 510(k) Summary for Emit® Il Plus Cocaine Metabolite Assay
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
KO31512 The assigned 510(k) number is:
- 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation:
Manufacturer:
Contact Information:
Syva Company - Dade Behring Inc. 20400 Mariani Ave Cupertino, CA 95014
Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Kathleen Dray-Lyons Tel: 781-826-4551 Fax: 781-826-2497
Preparation date:
July 14, 2003
- 2. Device Name/ Classification:
Emit® Il Plus Cocaine Metabolite Assay:
Cocaine and cocaine metabolite test system Class II (862.3250)
Product Code:
- 3. Identification of the Legally Marketed Device:
Emit® II Plus Cocaine Metabolite Assay (K993988)
র্ব . Device Description:
> Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay for qualitative and semi-quantitative analysis of cocaine metabolite (benzoylecgonine) in human urine.
91 DIO
{1}------------------------------------------------
Syva Company - Dade Behring Inc. 510(k) Notification - K031512 Additional Information
#### ર. Device Intended Use:
The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semi-quantitative analyses of benzoylecgonine (cocaine metabolite) in hurnan urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.
The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
- Medical device to which equivalence is claimed and comparison information: 6. The modified Emit® II Plus Cocaine Metabolite Assay is substantially equivalent in intended use to the Emil® II Plus Metabolite Assay currently marketed. The modified Emit® II Plus Metabolite Assay, like the current Emit® II Cocaine Metabolite Assay is intended to be used for the qualitative and semi-quantitative analyses of cocaine metabolite (benzoylecgonine) in human urine.
#### Device Performance Characteristics: 7.
### Method Comparison:
### Qualitative Results
### 150 ng/mL CUTOFF
| | 1<br>to compara al affered i<br>Career States of Children Comments of the Children<br>1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | |
| Partis of Program Career Property of<br>States States of States<br>No Valley Controller<br>199 1991<br>14.0<br>. | | |
| ಿ ಸಾಮಾನ್ಯ<br>11122222222222222<br><br>And Charles of Children Comments of the Children<br>100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | 大会 | |
Percent agreement: 93%
· GC/MS results were 123 and 140 ng/mL
** GC/MS results were 150, 151, 157, 162, 164, 166, and 190 ng/mL
### 300 ng/mL CUTOFF
| | <br><br>Childer The State 2<br>ええているという<br>A B<br><br><br> | |
|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|----|
| | | |
| Free Bronnell Comments of the commended to the comments of<br><br>.<br>19.55 55 55 55 | 1 6 | |
| <br><br><br>and the country of the first of the first of the first of the first of the first of the first and<br><br><br><br>. | A th | 84 |
#### Percent agreement: 94%
* GC/MS results were 218, 241, 261, 294, 295, and 298 ng/mL
** GC/MS result was 354 ng/mL
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird with outstretched wings, depicted in a stylized, minimalist design.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 2 3 2003
Ms. Kathleen Dray-Lyons Manager Regulatory Affairs and Compliance Syva Co., Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, DE 19714
Re: k031512
Trade/Device Name: Emit® II Plus Cocaine Metabolite Assay Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and Cocaine Metabolite test system Regulatory Class: Class II Product Code: DIO Dated: May 12, 2003 Received: May 14, 2003
Dear Ms. Dray-Lyons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Syva Company - Dade Behring Inc. 510(k) Notification - K031512 Additional Information
# Indications Statement
Emit® II Plus Cocaine Metabolite Assay Device Name:
## Indications for Use:
The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semi-quantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.
The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-ofabuse test result, particularly when preliminary positive results are used.