AQUILION SUPER 4 MULTISLICE CT SCANNER, MODEL TSX-101A/7
K031469 · Toshiba America Medical Systems, In.C · JAK · May 22, 2003 · Radiology
Device Facts
| Record ID | K031469 |
| Device Name | AQUILION SUPER 4 MULTISLICE CT SCANNER, MODEL TSX-101A/7 |
| Applicant | Toshiba America Medical Systems, In.C |
| Product Code | JAK · Radiology |
| Decision Date | May 22, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1750 |
| Device Class | Class 2 |
| Attributes | 3rd-Party Reviewed |
Intended Use
This device is designed to produce cross-sectional images of a human body by reconstruction of x-ray transmission data from the same axial plane taken at different angles. These images have been proven to be clinically useful in the diagnosis of spine and head injuries, intracranial tumors, blood clots in the brain, eye trauma, soft tissue lesions in the extremities, gastrointestinal lesions, abdominal and pelvic malignancies, and hepatic metastases. CT is also used to evaluate intestinal obstructions, assess intra-abdominal abnormalities and to examine musculoskeletal degeneration. This device employs no intended uses that are not in cleared devices already found in the marketplace.
Device Story
Aquilion Super 4 CT Scanner produces cross-sectional human body images; utilizes x-ray transmission data from multiple angles; reconstructs data into digital images for clinical diagnosis of injuries, tumors, lesions, and malignancies. Operated by clinical staff in medical settings; images stored electronically or filmed for physician review. Device features improved patient couch and updated computational subsystem hardware compared to predicate; retains same scanning hardware and software. Benefits include diagnostic utility for head, spine, abdominal, and musculoskeletal conditions.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and compliance with established performance standards for diagnostic X-ray equipment.
Technological Characteristics
High-frequency x-ray controllers; x-ray tube radiation source; x-ray detector array; computer-based image reconstruction. Digital image output. Conforms to 21 CFR 1020.30/1020.33 and IEC-60601. Hardware includes updated computational subsystem and improved patient couch.
Indications for Use
Indicated for X-ray imaging of the whole body using Computerized Tomography, including Axial, Volumetric (Helical), and CT Fluoroscopy.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- TSX-101A Aquilion CT w/ CGS-22A
Related Devices
- K984156 — SHIMADZU WHOLE BODY X-RAY COMPUTED TOMOGRAPHY SCANNER SCT-7000TX W/FLURO OPTION · Shimadzu Medical Systems · Apr 21, 1999
- K982583 — SHIMADZU WHOLE BODY X-RAY COMPUTED TOMOGRAPHY SCANNER SCT-6800TXL · Shimadzu Corp. · Oct 5, 1998
- K020913 — CT-C3000 SPIRAL CT SCANNER SYSTEM · Neusoft Digital Medical Systems Co., Ltd. · Apr 3, 2002
- K110066 — AQUILION PRIME · Toshiba America Medical Systems, In.C · Jan 26, 2012
- K013561 — LIGHTSPEED 4.0 CT SCANNER SYSTEM · GE Medical Systems · Nov 8, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
| Date: | April 8, 2003<br>MAY 22 2003 |
|----------------------------------------|--------------------------------------------------------------------------------------|
| Submitter's Name: | Toshiba America Medical Systems, Inc. |
| Submitter's Address: | P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068 |
| Submitter's Contact:<br>(714) 730-5000 | Paul Biggins, Regulatory Affairs Manager, |
| Establishment Registration<br>Number: | 2020563 |
| Device Proprietary Name: | TSX-101A/7, Aquilion Super 4 Multislice<br>CT Scanner |
| Common Name: | Scanner, Computed Tomography, X-Ray<br>[Fed. Reg. No. 892.1750, Pro. Code:<br>90JAK] |
| Regulatory Class: | II (per 21 CFR 892.1750) |
| Performance Standard:<br>Standard | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment |
| Predicate Device(s):<br>Multislice kit | TSX-101A Aquilion CT w/ CGS-22A |
| Reason For Submission | Modification of cleared device |
#### Description of this Device:
The Aquilion Super 4 CT Scanner is similar to the Aquilion Multislice CT scanner. The major difference between the two devices is a change of hardware in the computational subsystem and an improved patient couch. This device uses the same software and scanning hardware as the predicate device.
### Summary of Intended Uses:
This device is designed to produce cross-sectional images of a human body by reconstruction of x-ray transmission data from the same axial plane taken at different angles. These images have been proven to be clinically useful in the diagnosis of spine and head injuries, intracranial tumors, blood clots in the brain, eye trauma, soft tissue lesions in the extremities, gastrointestinal lesions, abdominal and pelvic malignancies, and hepatic metastases. CT is also used to evaluate intestinal obstructions, assess intra-abdominal abnormalities and to examine musculoskeletal degeneration. This device employs no intended uses that are not in cleared devices already found in the marketplace.
{1}------------------------------------------------
#### Technological Characteristics:
This device employs the same technological characteristics as the predicate device, differing only in the specifics of subassembly composition. Both of these systems employ the use of high frequency x-ray controllers to generate x-radiation from the x-ray tube. The x-ray transmission data is detected by the x-ray detector and is reconstructed by the computer. This device produces digital images that can be filmed or electronically stored for future review.
#### Safety and Effectiveness Concerns:
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR § 1020.30 and 1020.33, which apply to this upgrade, will be met and reported via a supplement to the initial report for the predicate device. Additionally this system is in conformance with the applicable parts of the IEC-60601 - Medical Device Safety standards.
#### Substantial Equivalence:
Based upon the above considerations TAMS believes that the Aquilion Super 4 CT System is substantially equivalent to the TSX-101A, Aquilion CT scanner. The major difference is a change in the configuration of the computational subsystem.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ----
MAY 22 2003
Toshiba America Medical Systems, Inc. c/o Ms. Laura Danielson Responsible Third Party Official TÜV America, Inc. 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891
Re: K031469
Trade/Device Name: Aquilion Super 4 CT Systems, TSX-101A/7 Regulation Number: 21 CFR §892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: May 8, 2003 Received: May 9, 2003
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
Aquilion Super 4 CT System; TSX-101 A/7 Device Name: Indications for Use:
> X-ray imaging of whole body - Computerized Tomography Including: Axial Volumetric (Helical) CT Fluoroscopy
Daniel A. Lyons
(Division Sign-Off Division of Reproductive, Ab and Radiological Devices 510(k) Number
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
1 Prescription Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________