K031419 · Laboratorios Grifols, S.A. · KPE · Nov 24, 2003 · General Hospital
Device Facts
Record ID
K031419
Device Name
GRI-FLEX, MODELS 1000ML, 500ML, 250ML, 100ML
Applicant
Laboratorios Grifols, S.A.
Product Code
KPE · General Hospital
Decision Date
Nov 24, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5025
Device Class
Class 2
Attributes
3rd-Party Reviewed
Intended Use
GRI-FLEX is a bag with an incorporated 0.2 µm filter for the removal of undesired particulate or microbial matter intended for use as a container in the preparation of drug solutions with the GRI- FILL system.
Device Story
GRI-FLEX is a single-use, sterile, flexible I.V. container with an integrated 0.2 µm polyethersulphone filter. Used in pharmacy settings with the GRI-FILL system for drug solution preparation; the device acts as a container for drug admixtures. The integrated filter removes undesired particulate or microbial matter from the solution. After preparation, the bag is connected to an I.V. administration set for patient infusion. Operated by trained healthcare personnel. The device benefits patients by reducing the risk of administering inadvertent contaminants, including bacteria and particulates, during intravenous therapy.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 and ISO DIS 15747. Functional testing conducted per ISO DIS 15747, USP <661>, ISO 8536-4, and ASTM F838-83 to verify correct operation of the bag and filter components.
Technological Characteristics
Flexible I.V. container with integrated 0.2 µm polyethersulphone filter membrane in a polypropylene housing. Available in 100 ml, 250 ml, 500 ml, and 1000 ml volumes. Single-use; sterile (SAL 10⁻⁶) via ETO. Complies with ISO 10993, ISO DIS 15747, ISO 8536-4, ASTM F838-83, and USP <661>.
Indications for Use
Indicated for use by trained healthcare personnel as a flexible I.V. container with an integrated 0.2 µm filter to remove particulate or microbial matter during the preparation of drug solutions using the GRI-FILL system. Restricted to prescription use only.
Regulatory Classification
Identification
An I.V. container is a container made of plastic or glass used to hold a fluid mixture to be administered to a patient through an intravascular administration set.
Predicate Devices
VIAFLEX PLASTIC CONTAINER (BAXTER)
PALL SUPOR AEF
Related Devices
K033916 — GRI-BAG, GRI-BAG AP · Laboratorios Grifols, S.A. · Dec 30, 2003
K964850 — SOLUTION ADMINISTRATION SET WITH 0.22 MICRON HIGH PRESSURE EXTENDED LIFE FILTER · Baxter Healthcare Corp · Feb 25, 1997
K030385 — IV ADD-ON FILTER ADMINISTRATION SET WITH AND WITHOUT Y CONNECT OR NEEDLELESS ACCESS CONNECTOR · Kawasumi Laboratories America, Inc. · Apr 15, 2003
K143583 — Cathivex ¿GV filter units · Merck Millipore , Ltd. · Aug 11, 2015
K972464 — VITALMIX PLUS EMPTY I.V. CONTAINER · Pacific Device, Inc. · Nov 5, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
K031419
න් පිහිටු පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පි
NOV 2 4 2003
:
| GRIFOLS | | TECHNICAL EVALUATION DOCUMENTATION | # Document: | TED-GRIFLEX-01 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-------------|----------------|
| | | SECTION 1 - GRI-FLEX: 510(k) SUMMARY | | |
| DATE OF SUBMISSION: | 2003-02-03 | | | |
| SUBMITTER NAME: | Laboratorios Grifols, S.A. | | | |
| SUBMITTER ADDRESS: | C/ Can Guasch, 2 08150 PARETS DEL VALLES BARCELONA SPAIN | | | |
| TELEPHONE: | + 34 93 571 01 00 | | | |
| FAX: | + 34 93 573 09 12 | | | |
| e-mail: | sebastian.gascon@grifols.com | | | |
| CONTACT: | Sebastián Gascón Technical Director | | | |
| DEVICE TRADE NAME: | GRI-FLEX | | | |
| COMMON NAME: | I.V. BAG with in-line 0.2µm filter | | | |
| CLASSIFICATION NAME: | I.V. CONTAINER (880.5025 KPE and 880.5440, NEP) with in line filter (880.5440, FPB) | | | |
| PREDICATE DEVICE: | (BAG) VIAFLEX PLASTIC CONTAINER (BAXTER) (FILTER) PALL SUPOR AEF | | | |
| DEVICE DESCRIPTION: | GRI-FLEX is a single-use, non-pyrogenic flexible empty container with incorporated 0.2 µm filter. It is supplied sterile in sealed peel-pack pouches and is available in volume capacities of 100 ml, 250 ml, 500 ml and 1000 rıl. | | | |
| INTENDED USE: | GRI-FLEX is a bag with an incorporated 0.2 µm filter for the removal of undesired particulate or microbial matter intended for use as a container in the preparation of drug solutions with the GRI- FILL system. | | | |
Page 18 of 32
{1}------------------------------------------------
KO3/419
| | On the super of the state of the state of the state of the state of the county of<br>RIVIUAL LIT<br><br><br>CARTER LAND STATE CONTRACT CARACTER CARAL CARAL CARAL CARAL CARA CONTRACT CARAL CONTRACT CARACT CARACT CARACT CONTRACT CONTRACT CARACT CARACT CARACT CARACT CARACT CARACT CARA<br> | Company Company Comments of Children Comments of<br>The Research Party of the Property of<br>12 - 11 - 12 - 12 - 12 - 1 20<br>DODITITUT TEAST<br>Comments of the management of the more of the minimal comments of the mannel of the mannel of the mannel | because . BEREST LAUNES LEVENT TO ICCULERATIONAL Top<br> # Document: # Document: # Document<br>A . A B . F . L<br>and the country of the county of the county of the county of the county of the county of the county of the county of the county of the county of the county of the county of t<br>SCORDELE FERRENT FREERS SERVER FLE 2000 CONTRACT FOR ARRETT FOR ARRETT<br>CARDER CONSULTION COLLECTION COLLECTION CONSULTER CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CON<br>Children and the country of the country of the county of the county of the county of the county of the county of<br>Comment Carlos Comers<br><br>ERTER ! ! " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "<br>Comments of the first of the first of the contribution and consisted to comments of the comments of the more of the more of<br>Charaka TAATEELanda A La LEWILLE FE' L<br>LES BERTEPS LE LES LEBERES P F L L C C C C C C C C C C C C C C C C C C C C C C C C C C . C . C<br><br>Comments of the consisted on the contribution is a consideration of the comments of the comments of the comments of<br>Comments of Concession Compression Comprehension Career Comments of Canada Compression<br>Comments of the commended to the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of t<br>Carles Concerner Consistent Consistence of Children Comments of Children Comments of Children Comments of Children<br>Comments of the consisted on the commend of the commend of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the<br> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CALL POTTER PERSONAL COLLECT LEAR CARRELLAR CARRELLA COLLEGIAL<br>MALL NATION CONSULTION COLOGICOLOGICOLOGICAL COLLULARIO LOCK<br>.<br>.<br>Canada Children Color Station Commended Commender Comment Comment<br>FAR (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1<br><br>. | 03 800 65.05 00 1 00<br>1 20 1 888 0 000 -<br>CLEEPPERESSES<br>COLLECT COLLECT TO COLLECT TO COLLECT COLLEGION COLLECTION COLLECTION COLLECTION<br>Company of the first of the program and the many of the many of the comments of the<br>Comments of the Partifical Property of September 1999 for the Comments of Concession of Concession of Concession of Concession Comparison<br>Montana Marcel Market Pro analy Property Program 2000 MP Parties Program Program Program Program Program Province Province Province Province Province Province Province Provin<br> | 199188 88188<br>and the consistence of the manufacturer of the more of the management of the comments of the comments of the commended to the<br>Comments of the comments of the commend of the commended to the comments of the comments of the comments of the may<br>PAN TO T RAA SHO<br>Company of the comments on the many of the many of the many of the many of the many of the many of the many of the may<br>and and the can and the commend of the many of the comments of the comments of the comments of the more of the more of the more of the more of the more of the more of the mor<br><br>11 . BRISTER START FEE FEBRE KEEW STATUS AND PERSONAL<br><br> | <br>Portuges of the program and the program and the program and any and<br>Canada Call Company Call Proven Career Passe Class Class Canada C<br>CONTRACONIALI ANDE FIGULA QUIRIELOVE CONTRACT<br>Carlos Controllation Comments of Children<br><br>КЕРЕДЕЕМЕ. В Г. СПОССИОНИЯ В СОДЕЛИ В ОД ВОДОВОД В ОД ВОДОВОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В ОД В<br><br><br><br><br>LARE LARRET E ST LE B LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE<br>NETTI E FREELL BELLEAST LE FREE LEFER POST STORES<br>THE CELESS AND THE COLLECT COLLEGELE CONNECTION CONSULTION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLECTION COLL<br>.<br>THE TOUCH THE LE FELL TRE FEE LE FREE E<br>.<br><br><br>.<br><br><br><br>CONTRACTORAL COLLECTION COLLECTION CONSULTER CONSULTER CONSTITUTION CONTRACT AND CLAIM CAN<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
## SUMMARY OF COMPARISON WITH PREDICATE DEVICE:
In the establishment of substantial equivalence, the GRI-FLEX device is compared with 2 predicate devices (bag + filter) due to the fact that it is essentially a combination of these 2 devices marketed as a single unit.
The following table summarizes the similarities of the principal technological characteristics and features of both predicate and new devices.
| # | Characteristic / Feature | GRI-FLEX | PREDICATE | |
|----|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | BAG | FILTER |
| 1. | Filter membrane | Polyethersulphone | N/A | Polyethersulphone |
| 2. | Filter pore | 0.2μm | N/A | 0.2μm |
| 3. | Filter housing | Polypropylene | N/A | Polypropylene |
| 4. | Sterility | SAL 10⁻⁶ ETO | SAL<br>Gamma | SAL 10⁻⁶ ETO |
| 5. | Single-use | YES | YES | YES |
| 6. | Intended use | Bag, with an<br>incorporated 0.2 μm<br>filter for the removal<br>of undesired drug<br>particulate or<br>microbial matter,<br>intended for use as<br>a container in the<br>preparation of drug<br>solutions with the<br>GRI-FILL system. | Bag intended to<br>be used in the<br>preparation of<br>drug admixtures<br>available in sizes<br>from 50 ml - 400<br>ml. | Removal by in-line<br>filtration of<br>inadvertent<br>contaminants<br>(including bacteria,<br>particulates, and<br>entrained air) from<br>infused intravenous<br>fluids. |
The principal differences between the GRI-FLEX device and the predicate devices lie with the fact that the material used for the bag is different and also, the predicate devices are marketed separately whereas the GRI-FLEX is presented as an integral unit combining the two.
## SUMMARY DISCUSSION OF NON-CLINICAL DATA:
According to the biocompatibility test data available for all materials and the biological testing performed on the final product, we have established that the GRI-FLEX device fulfills the requirements set out in ISO 10993 and ISO DIS 15747.
Functional testing performed in accordance with applicable clauses of ISO DIS 15747 and USP <661> (for the bag component) and ISO 8536-4 and ASTM F838-83 (for the filter component) shows correct operation of the device as per its intended use.
2
{2}------------------------------------------------
| | And and the county of the program and the county of the county of<br><br> | A AR ES S E F F P S<br>I = Document: " = Document<br>CONTRACTOR COLLECT COLLECT CARDED CONSULT CREASE COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONS<br>Comments of the commend of 2012 22 21 31 31 31 31 31 31 31 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 1<br>AMARAGE CONSULTION ALLEA POSTER STEEL LETTER CONSTITUTION COLLECTE<br>beared to and and a sense to the country of the class .<br><br>SECOND STATE LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST<br><br><br><br>Callery of the first and and and and and and and the comment of the comments of<br><br><br>Comments of the state of the state and states and the comments of the comments of the comments of the first of the first of the first of the first of the first of the first o<br>MANAGER PARTIER FREE PARTY PRODUCT COLLECT COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION<br>SAMMANAGE LEE LEE LEE LEE LEE LEE LEE LEE LEE LEE SER LEE LEAST LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE<br><br>TO CONTINUE OF COLLECTION A A A CARLAN A A A CA A L L L C OCCULT OF C L C O C L C . C<br>CONSULTION COLLEGION COLLECTION COLLECTION CONSULTER CONSULTER CONSULTER COLLECTION CONSULTER COLLECTION CONSULTER COLLECTION CONSULTER COLLECTION CONSULTER COLLECTION CONSUL<br>POST COLLECT COLLECT CONSULTION CONSULTION CONSULTION CONSULTION CONTINUES OF THE COLLECTION CONTINUES OF THE COLLECTION CONTINUE<br>PERSONALERS PARECENTRA COLUMNIC LULA<br>and and the manufacturer and the many commend of the many of the many of the many of the<br>Comments of the contribution of the contribution of the comments of the comments of |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| .<br><br><br><br>.<br> | CORPORAL TO THE SECTION SECTION CALL CARAL<br>TEXTERESERANE<br>Comments of Acres Company College Life Species and Company of<br>CONSTITUTION COLUMNIA COLUMNICAL<br>and the comments of the first of the first to the first to the first to the first to the first to the first to the first to the first to the first to the first to the first t<br>FEREE E ME LAND<br>THE LED COLLECT COLLECT CONTRACT<br><br>CONNECES A CONTRACTOR COLLECT CALLERS CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT C<br><br><br>MACONS LEARLE LEAST LEAST REAL PRESELLER LETT COLLECT COLLECTION COLLECTION<br><br>Comments of the consisted on the consisted to the contribution of the comments of the comments of the comments of the comments of the comments of the comments of the contribu<br>CALLER P. P. C. CLELELAR STATE SEMARAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA | and and the management of the many of the many of the more of the comments of the comments of the many of the many of the many of the many of the many of the many of the many<br>The first and a security of the same to the can and the comments of the comments of the comments of<br><br>A B B P<br>67.00<br><br>.<br>********************* |
## CONCLUSIONS:
We believe the intended use, the indications for use and the design for both GRI-FLEX and the combination of predicate devices are essentially the same. Moreover, the construction materials, although different, are of comparable safety and, hence,
substantial equivalence of GRI-FLEX with the legally marketed devices may be established.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 4 2003
Laboratoris Grifols, S.A. C/O Ms. Susan A. Gill Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive, P.O. Box 13995 Research Triangle Park, North Carolina 27709-3995
Re: K031419
Trade/Device Name: GRI-FLEX Regulation Number: 880.5025, 880.5440 Regulation Name: I.V. Container, Intravascular Administration Set Regulatory Class: II Product Code: KPE, NEP, FPB Dated: November 12, 2003 Received: November 13, 2003
Dear Ms. Gill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Gill
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Qurfs
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## # Documint TECHNICAL EVALUATION GRIFOLS TED-GRIFLEX-09 DOCUMENTATION SECTION 09 - GRI-FLEX: INDICATIONS FOR USE STATEMENT
PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT (as required by ODE for all 510(k) received after Jan. 1, 1996)
510(k) Number:
Device Name:
GRI-FLEX
Indications for Use:
All models of GRI-FLEX are flexible I.V. bags, with incorporated 0.2 um filter for the removal of undesired particulate or microbial matter, for use with the GRI-FILL pharmacy compounding system as a container in the preparation of drug solutions. The drug solution is later administered to the patient by cor necting the bag to an I.V. administration set. This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physicis n.
(Do not write below this line. Continue on another page in needed) Concurrence od CDRH, Office of Device Eval ration (ODE)
OR
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use
Pulitzer Crescent
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device
510(k) Number: K031419
Page 17 of 32
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.