VITEK 2 GRAM POSITIVE LINEZOLID FOR STREPTOCOCCUS PNEUMONIAE

K031410 · bioMerieux, Inc. · LON · Jul 10, 2003 · Microbiology

Device Facts

Record IDK031410
Device NameVITEK 2 GRAM POSITIVE LINEZOLID FOR STREPTOCOCCUS PNEUMONIAE
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateJul 10, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK 2® Gram Positive Linezolid for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. It is intended for use with the VITEK 2® System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

Device Story

VITEK 2 AST-GP Linezolid card performs in vitro antimicrobial susceptibility testing for Streptococcus pneumoniae. Bacterial isolate is diluted in 0.45% saline, used to rehydrate antimicrobial medium in card. VITEK 2 system automatically fills, seals, and incubates card. System monitors growth in each well over 18 hours. Output includes MIC value and interpretive category result for antibiotic. Used in clinical laboratories by technicians/microbiologists to aid clinical decision-making regarding antibiotic therapy.

Clinical Evidence

External evaluation using fresh and stock clinical isolates and stock challenge strains. Compared performance against NCCLS microbroth dilution reference method. Results showed 100.0% overall Category Agreement. Reproducibility and Quality Control testing demonstrated acceptable results.

Technological Characteristics

Automated antimicrobial susceptibility testing system. Uses miniaturized doubling dilution technique in AST cards. Incubator/reader monitors bacterial growth via optical sensing over 18 hours. Standalone system component. Software-controlled automated filling, sealing, and incubation.

Indications for Use

Indicated for antimicrobial susceptibility testing of Streptococcus pneumoniae in clinical laboratory settings. For prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a date and a company name. The date is "JUL 10 2003", indicating July 10th, 2003. Below the date, the company name "BIOM ÉRIE U X" is printed in a stylized font. There is also a black circle between the date and the company name. K0314/10 ## 510(k) SUMMARY ### VITEK 2® Gram Positive Linezolid for Streptococcus pneumoniae #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |----------------------|--------------------------------------------------------------------------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Nancy Weaver, Staff Regulatory Specialist | | Phone Number: | 314-731-8695 | | Fax Number: | 314-731-8689 | | Date of Preparation: | April 25, 2003 | | B. Device Name: | | | Formal/Trade Name: | VITEK 2® Gram Positive Linezolid for<br>Streptococcus pneumoniae | | Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device,<br>21 CFR 866.1645 | | Common Name: | VITEK 2 AST-GP Linezolid for Streptococcus<br>pneumoniae | | C. Predicate Device: | VITEK 2® Streptococcus pneumoniae<br>Susceptibility Test for Ceftriaxone (N50510/S135) | ### D. 510(k) Summary: The VITEK 2® Gram Positive Linezolid for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. It is intended for use with the VITEK 2® System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) my microdilution methodology. The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 automatically fills, seals and places the card into the incubator/reader. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). {1}------------------------------------------------ At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card. VITEK 2® Gram Positive Linezolid for Streptococcus pneumoniae demonstrated substantially equivalent performance when compared with the NCCLS reference microbroth dilution method, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial susceptibility Devices", dated March 8, 2000 and Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Feb.5 2003. The Premarket Notification (510[k]) presents data in support of VITEK 2® Gram Positive Linezolid for Streptococcus pneumoniae. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Positive Linezolid for Streptococcus pneumoniae by comparing its performance with the NCCLS microbroth dilution reference method. VITEK 2 Gram Positive Linezolid for Streptococcus pneumoniae demonstrated acceptable performance of 100.0% overall Category Agreement when compared to the microbroth dilution reference method. Reproducibility and Quality Control demonstrated acceptable results. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows a logo with a circular border containing text, surrounding a stylized depiction of an eagle. The eagle is composed of three curved lines that suggest the wings and body of the bird. The text around the border is small and difficult to read, but it appears to be part of an organization's name. The logo is simple and uses a monochrome color scheme. JUL 1 0 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Nancy Weaver Staff Regulatory/Clinical Affairs Specialist BioMerieux, Inc. 595 Anglum Road Hazelwood, MO 63042-2320 Re: k031410 Trade/Device Name: VITEK 2® Gram Positive Linezolid (1 ug/ml) for Streptococcus pneumoniae Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: April 25, 2003 Received: May 13, 2003 Dear Ms. Weaver: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 # Indications for Use Statement **510(k) Number (if known):** K034110 Device Name: VITEK 2® Gram Positive Linezolid (1ug/ml) for Streptococcus pneumoniae Indications for Use: VITEK 2® Gram Positive Linezolid for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. It is intended for use with the VITEK 2® System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Freddie L. Poole Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) KO3 1410 for Prescription Use only
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...