INTERGRO DBM

K031399 · Interpore Cross Intl. · MQV · Feb 18, 2005 · Orthopedic

Device Facts

Record IDK031399
Device NameINTERGRO DBM
ApplicantInterpore Cross Intl.
Product CodeMQV · Orthopedic
Decision DateFeb 18, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Indications for Use

InterGro DBM products are to be used for filling bony voids or gaps in the extremities, and pelvis that are not intrinsic to the bony stability of the structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to bone. The device may also be used for filling craniofacial defects and craniotomies that are no larger than 25cm². The amount of InterGro DBM products to be used should be based on the type of procedure and size of the graft site.

Device Story

InterGro DBM is a bone graft substitute composed of human allograft bone in a lipid carrier; some versions include porous ceramic granules (calcium carbonate/calcium phosphate). Device is ready-to-use, requiring no rehydration. Used by surgeons to fill non-structural bony voids in extremities, pelvis, and craniofacial sites. Provides a scaffold for bone healing. Clinical benefit derived from filling osseous defects to support bone repair. Aseptically processed for single-patient use.

Clinical Evidence

No clinical trials provided. Evidence includes viral inactivation validation, in vitro SAOS human osteosarcoma cell proliferation assays for osteoinductive potential, and animal model performance testing (rabbit and sheep) using radiographic and histological methods.

Technological Characteristics

Composition: Human allograft bone, lipid carrier, and optional ceramic granules (calcium carbonate with calcium phosphate coating). Form: Granulated, ready-to-use. Sterilization: Aseptic processing. No electronic or software components.

Indications for Use

Indicated for filling non-structural bony voids or gaps in extremities and pelvis, including surgically created or traumatic osseous defects, and craniofacial defects/craniotomies ≤ 25cm².

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY K031399 This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR \$807.92. | Submitted By: | Interpore Cross International<br>181 Technology Drive<br>Irvine, California 92618 | |-----------------|-----------------------------------------------------------------------------------| | Date: | December 16, 2003 | | Contact Person: | Mark Loar<br>Manager, Regulatory and Clinical Affairs | ## CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME Classification Name: Human Tissue Common/Usual Name: DBM Bone Graft Substitute Product Classification: Unclassified Proprietary Name: InterGro® DBM ### PREDICATE DEVICE Predicate device information is provided in this premarket notification. #### DEVICE DESCRIPTION InterGro® DBM products contain human tissue (allograft bone) and are intended for transplantation. The allograft bone has been granulated, demineralized and provided in a lipid carrier. Some versions contain porous ceramic granules that are a composite of highly resorbable calcium carbonate with a slower resorbing 2 to 10 um outer layer of calcium phosphate. InterGro DBM products have been processed aseptically and are ready to use. They do not require rehydration or any special preparation. InterGro DBM is intended for single patient use only. #### INDICATIONS-FOR-USE InterGro DBM products are to be used for filling bony voids or gaps in the extremities, and pelvis that are not intrinsic to the bony stability of the structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to bone. The device may also be used for filling craniofacial defects and craniotomies that are no larger than 25cm". The amount of InterGro DBM products to be used should be based on the type of procedure and size of the graft site. {1}------------------------------------------------ ## SUBSTANTIAL EQUIVALENCE INFORMATION InterGro® DBM was found to be substantially equivalent to the predicate devices. The safety and effectiveness of InterGro DBM is adequately supported by the substantial equivalence information, materials data, and testing results provided within this Premarket Notification. . Viral Inactivation Validation The methods for processing the DBM contained in InterGro® DBM were evaluated for their viral inactivation potential. A select panel of viruses representing various virus types, sizes, shapes, and genomes were evaluated. The viral inactivation testing demonstrated suitable viral inactivation potential of the processing methods for a wide range of potential human viruses. - Osteoinductive Potential . Each lot of DBM incorporated into InterGro® DBM is assayed for its osteoinductive potential. The assay measures the proliferation of SAOS human osteosarcoma cells in the presence of human DBM compared to positive and negative controls (osteoinductive index). Results of the assay have been correlated with results from implantation of DBM into athymic rat muscle, which demonstrated a correlation coefficient of 0.850 (p<0.0005) and accurately predicted the in vivo osteoinductivity in 25 donor lots. Additionally, clinical results using DBM with >0.20 and ≤0.20 (osteoinductive index) demonstrated a significant difference in healing as evaluated by radiography, 92% and 33% healing, respectively.2 The combination of DBM, the carrier and, in some formulations, ceramic granules has not been evaluated for osteoinductivity; therefore it is unknown to what extent the formulation components may alter the osteoinductive character of the DBM. Additionally, it is unknown how osteoinductivity of the DBM component. measured via the in-vitro "SAOS" bioassay will correlate with human clinical performance of InterGro DBM products. - . Product Performance Testing Performance of InterGro® DBM was evaluated in rabbit and sheep models by radiographic and histological methods. l Adkisson HD, Strauss-Schoenberger J, Gillis M, Wilkins R, Jackson M, and Hruska KA. Rapid Quantitiative Bioassay of Osteoinduction. J Ortho Res, 2000, 18:503-511. י נ Wilkins RM, Clinical Effectiveness of Demineralized Bone Matrix Assayed in Human Cell Culture, Advances in Tissue Banking. 1999 3:113-124 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures. FEB 1 8 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Mark Loar Manager, Regulatory and Clinical Affairs INTERPORE CROSS International 181 Technology Drive Irvine, California 92618 Re: K031399 Trade/Device Name: InterGro® DBM Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler Regulatory Class: II Product Code: MQV,GXP Dated: December 16, 2004 Received: December 21, 2004 Dear Mr. Loar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assession to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Mr. Mark Loar This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally premaince nedicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, O. Mark N. Millikan Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K031399 InterGro® DBM Device Name: #### Indications-For-Use: InterGro DBM products are to be used for filling bony voids or gaps in the extremities, and pelvis that are not intrinsic to the bony stability of the structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to bone. The device may also be used for filling craniofacial defects and craniotomies that are no larger than 25cm². The amount of InterGro DBM products to be used should be based on the type of procedure and size of the graft site. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |------------------------------------------------------------|--| | Division of General, Restorative, and Neurological Devices | | | 510(k) Number | K031399 | |---------------|---------| |---------------|---------| Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...