← Product Code [LGZ](/productcode/LGZ) · K031390

# MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODEL MMT-712 (K031390)

_Minimed, Inc. · LGZ · Jul 23, 2003 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K031390

## Device Facts

- **Applicant:** Minimed, Inc.
- **Product Code:** [LGZ](/productcode/LGZ.md)
- **Decision Date:** Jul 23, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Medtronic MiniMed Paradigm Model 712 Insulin Pump is indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

## Device Story

External, portable insulin pump for continuous subcutaneous insulin infusion. Device delivers U100 insulin at basal rates (0.00-35.00 units/hr) and meal boluses (up to 25.00 units). User inputs include insulin sensitivity, insulin-to-carbohydrate ratio, and target blood glucose (BG) to calculate bolus doses via integrated Bolus Wizard. Features include BG reminders for daily or post-bolus checks. Programming allows 0.1 unit bolus increments and 0.05 unit basal increments. Uses proprietary 3.0 ml reservoir compatible with Paradigm infusion sets. Operated by patient to manage glycemic control; healthcare provider programs settings. Benefits include precise, automated insulin delivery tailored to individual metabolic needs.

## Technological Characteristics

External, portable insulin pump. Accommodates proprietary 3.0 ml reservoir. Compatible with Paradigm infusion sets. Basal delivery increments: 0.05 units. Bolus delivery increments: 0.1 units. Software-controlled delivery rates.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Medtronic MiniMed Paradigm Model 511 Insulin Pump

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ms. Mirielle Mengotto Senior Regulatory Affaires Specialist Medtronic MiniMed 18000 Devonshire Street Northridge, California 91325

JAN 1 0 2017

Re: K031390

Trade/Device Name: Medtronic MiniMed Paradigm Model 712 Insulin Pump Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: April 30, 2003 Received: May 2, 2003

Dear Ms. Mirielle Mengotto:

This letter corrects our substantially equivalent letter of Jul 23, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809 ): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

## Page 2 - Ms. Mirielle Mengotto

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809] ), please contact the Division of Industry and Consumer Education at its toll-free no mber (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

## Erin I. Keith -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Medronic MiniMed Premarket Notification - 510(k) Paradigm 712m Insulin Pump

## INDICATIONS FOR USE .

14031390

510(k) Number:

Device Name:

Medtronic MiniMed Paradigm Model 712 Insulin Pump

Indications for Use:

The Medtronic MiniMed Paradigm Model 712 Insulin Pump is indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

Pastatia Cuscenite

sion Sinn. Off ision of Anesthesio Infection Control, Denta

**510(k) Number:** K031390

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _ V (Per 21 CFR 801.109)

or

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

Confidential

ix

{3}------------------------------------------------

Medronic MiniMed Inc. Premarket Notification-510 (k) Paradigm Model 712 Insulin Pump

JUL 2 3 2003

## SECTION D. 510(k) Summary For KO31390

In accordance with the requirements of SMDA 1990, and 21 CFR 807.92, this 510(k) Summary is provided:

Submitter: Medtronic MiniMed, 18000 Devonshire St., Northridge, CA 91325

Contact: Mirielle Mengotto (818) 576-4112

Name of Device: Medtronic MiniMed Paradigm Model 712 Insulin Pump

Predicate Device: Medtronic MiniMed Paradigm Model 511 Insulin Pump

Description of the Device: The Paradigm Model 712 is an external, portable insulin pump, designed for continuous deliveryof insulin. It is designed to deliver 0.00 to 35.00 units of U100 insulin per hour in basal rates and up to 25.00 units of U100 insulin per meal bolus. This pump will include a bolus estimator feature that can be tailored to an individual user's insulin sensitivity, insulin to carbohydrate ratio, and target BG (Blood Glucose). Also, there is a BG reminder feature, allowing the user to set daily and post-bolus reminders to check The insulin programming step size will allow users to program insulin BG readings. delivery in steps of 0.1 units for boluses and 0.05 units for basal delivery.

The Model 712 accommodates a proprietary 3.0 ml reservoir. This new reservoir mates with the existing and future Paradigm infusion sets.

Intended Use of the Device: The Medtronic MiniMed Paradigm Model 712 Insulin Pump is intended for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

Comparison of the Technological Features of the New Device and Predicate Device: The new and predicate devices have similar materials and basic design. The new device will have a larger case than Paradigm Model 511 and contains software modifications including the Bolus Wizard, as compared to the predicate, which does not.

of the

Gerda Resch Manager, Regulatory Affairs Medtronic MiniMed

4/30/03

Date

Confidential

TM Paradigm Model 712is a Trademark of Medtronic MiniMed

---

**Source:** [https://fda.innolitics.com/device/K031390](https://fda.innolitics.com/device/K031390)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
