← Product Code [MNI](/productcode/MNI) · K031354

# EBI OMEGA21 SPINAL FIXATION SYSTEM (K031354)

_Ebi, L.P. · MNI · Aug 6, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K031354

## Device Facts

- **Applicant:** Ebi, L.P.
- **Product Code:** [MNI](/productcode/MNI.md)
- **Decision Date:** Aug 6, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The EBI® SpineLink® II Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

## Device Story

Spinal fixation system consisting of rods, screws, couplers, and hooks; used for non-cervical stabilization. Implanted by surgeons in clinical settings to address spinal pathologies including degenerative disc disease, trauma, and deformities. Provides mechanical support to the spine to facilitate fusion or stabilization. System components are assembled intraoperatively to match patient anatomy.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Components manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. System includes rods, screws, couplers, and hooks. Mechanical fixation device.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Spinal Concepts BacFix Spinal Fixation System ([K013887](/device/K013887.md))
- Sofamor Danek CD Spinal System ([K964159](/device/K964159.md))
- Sofamor Danek Townley Pedicle Screw Plating System ([K983706](/device/K983706.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# AUG - 6 2003

# K031354 Page 1 of 2

## 510(k) Summary

This 510(k) Summary for the EBI® Omega21™ Spinal Fixation System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.

Contact Person: Frederic Testa, RAC 1. Submitter: EBI, L.P. Telephone: (973) 299-9300, ext.2208 100 Interpace Parkway Parsippany, NJ 07054

Date prepared: July 14, 2003

| 2. Proprietary Name:  | EBI® Omega21™ Spinal Fixation System            |
|-----------------------|-------------------------------------------------|
| Common Name:          | Spinal Fixation Device                          |
| Classification Names: | Spondylolisthesis Spinal Fixation Device System |
|                       | Spinal Intervertebral Body Fixation Orthosis    |
|                       | Spinal Interlaminal Fixation Orthosis           |

# 3. Predicate or legally marketed devices that are substantially equivalent:

- Spinal Concepts BacFix Spinal Fixation System (K013887) .
- Sofamor Danek CD Spinal System (K964159) .
- Sofamor Danek Townley Pedicle Screw Plating System (K983706) ●
- 4. Description of the device: The EBI® Omega21 Spinal Fixation System is a spinal fixation device that utilizes rods, screws, couplers and hooks. This submission is for the modification to the Indications for Use.
- 5. Intended Use: The EBI® SpineLink® II Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system.

{1}------------------------------------------------

Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

- 6. Materials: The components of the System are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136.
- 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the proposed EBI® Omega21™ Spinal Fixation System and the currently marketed EBI Omega21 Spinal Fixation System. It is substantially equivalent* to the predicate devices in regards to intended use, materials, and function.

<sup>*</sup>Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or reclassification and is not interpreted as an admission or any other type of evidence in patent infringement litigation. JEstablishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" in a circular arrangement. The central symbol consists of three curved lines stacked on top of each other, resembling a stylized bird or abstract design. The logo appears to be a seal or emblem, possibly representing a government agency or organization related to health and human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 6 2003

Mr. Frederic Testa, RAC Senior Regulatory Affairs Specialist EBI, L.P. 100 Interpace Parkway Parsippany, NJ 07054

Re: K031354

Trade/Device Name: EBI Omega21 Spinal Fixation System Regulation Number: 21 CFR 888.3050, 888.3060, 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation Orthosis, Pedicle screw spinal system Regulatory Class: III Product Code: KWP, KWQ, MNH, MNI, NKB Dated: July 14, 2003 Received: July 15, 2003

Dear Mr. Testa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

### Page 2 - Mr. Frederic Testa, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

forCelia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Statement of Indications for Use

Page 1 of of los

510(k) Number (if known): K031354

Device Name: EBI® Omega21™ Spinal Fixation System

Indications For Use:

The EBI Omega21 Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use (Optional Format 1-2-96)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K031354

---

**Source:** [https://fda.innolitics.com/device/K031354](https://fda.innolitics.com/device/K031354)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com