SynVitro® Hyadase is for the removal of the cumulus complex and corona radiata surrounding the oocyte in preparation for ICSI.
Device Story
SynVitro® Hyadase is a ready-to-use hyaluronidase solution derived from a non-bovine source; used in clinical laboratory settings by embryologists to denude oocytes prior to ICSI. The solution facilitates the removal of the cumulus complex and corona radiata, allowing access to the oocyte for injection. It contains no human serum albumin (HSA) or antibiotics. The product is supplied as a liquid reagent; it does not have direct patient contact. Clinical benefit includes effective preparation of oocytes for fertilization procedures without significant differences in embryo quality, fertilization rates, or cleavage rates compared to HSA-containing alternatives.
Clinical Evidence
Clinical evidence includes two studies. First, a study comparing hyaluronidase from two different sources showed no significant difference in embryo morphological quality, fertilization rate, or cleavage rate. Second, a randomized clinical study compared SynVitro® Hyadase (no HSA) to the predicate Medi-Cult Hyaluronidase (contains HSA). Results confirmed the subject device is equally effective regarding embryo quality, fertilization rate, and cleavage rate. No serious adverse events or complaints were reported.
Technological Characteristics
Ready-to-use hyaluronidase solution (80 IU/ml) from non-bovine source. Ingredients: E ART supplement, glucose, sodium pyruvate, calcium chloride, magnesium sulphate, sodium chloride, sodium bicarbonate, HEPES. Does not contain HSA or antibiotics. Quality controls include sterility, pH, osmolality, and Mouse Embryo Assay (Blastocyst rate > 70%).
Indications for Use
Indicated for the removal of the cumulus complex and corona radiata surrounding the oocyte in preparation for intracytoplasmic sperm injection (ICSI).
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{1}------------------------------------------------
K031228
# JUN 2 4 2003
# 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination
#### Intended use
SynVitro® Hyadase is for the removal of the cumulus complex and corona radiata surrounding the oocyte in preparation for ICSI.
#### Composition
- Hyaluronidase 80 IU/ ml (non-bovine source) 내
- E ART supplement
- 트 Glucose
- Sodium pyruvate I
- 트 Calcium chloride
- 트 Magnesium sulphate
- 트 Sodium chloride
- 피 Sodium bicarbonate
- 비 HEPES
### Biocompatibility testing
When following the procedure described in our package insert Syn Vitro@Hyadase will not have patient contact, thus biocompatibility testing has not been performed.
## Product testing controls
Sterility tested pH tested Osmolality tested Mouse Embryo Assay, (one cell assay, Blastocyst rate > 70 %) For each batch a Certificate of Analysis with the results of the above tests is available.
### Clinical Documentation:
Prior to the ICSI procedure, the cells of the cumulus and corona radiata must be removed in order to facilitate access to the oocyte and minimize contamination of the injection needle.
SynVitro® Hyadase is a ready-to-use hyaluronidase product designed for denudation of the oocyte and is based on raw material from a non-bovine source. SynVitro®Hyadase does not contain human serum albumin (HSA) or antibiotics.
{2}------------------------------------------------
Hyaluronidase originating from two different sources were compared in a clinical study. Two comparable groups of oocytes were inseminated after the cumulus had been removed using one of the two sources of hyaluronidase. The result of the study showed that there was no significant difference between the two groups regarding the morphological quality of the embryos, the fertilisation rate and the clevage rate.
Furthermore a randomised clinical study has been performed comparing SynVitro® Hyadase, which does not contain human serum albumin (HSA) to Medi-Cult's Hyaluronidase product that contains HSA. The results of this study is that SynVitro® Hyadase and Medi-Cult Hyaluronidase are equally effective regarding the morphological quality of the embryos, the fertilisation rate and the cleavage rate.
Based on these studies it is concluded that SynVitro® Hyadase is substantial equivalent to Medi-Cult Hyaluronidase ( K 991334) and is effective for denuding oocytes prior to ICSI.
During our studies there has been no registered complaints and no evidence that the product has been the cause of any serious adverse events in connection with its intended use.
Thus based on the clinical data presented and our experience with the SynVitro®Hyadase product we feel that the safety and effectiveness of the product for its intended use is shown in the present submission and the product is substantially equivalent to the predicated device Medi-Cult Hyaluronidase ( K 991334).
Prepared and Submitted by:
Ronald G. Leonardi, Ph.D.
Ronald G. Leonardi. Ph.D. President R & R REGISTRATIONS P.O. Box 262069 San Diego, Ca 92131 619-586-0751
April 17, 2003
Date
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 4 2003
Medi-Cult a/s % Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 SAN DIEGO CA 92196-2069 Re: K031228 Trade/Device Name: SynVitro® Hyadase Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: 85 MQL Dated: April 18, 2003 Received: April 24, 2003
Dear Dr. Leonardi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{5}------------------------------------------------
#### k031228 510(k) Number (if known)
SynVitro®Hyadase Device Name:
# INDICATIONS FOR USE:
For the removal of the cumulus complex and corona radiata surrounding the oocyte in preparation for ICSI.
(PLEASE DO NO WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Jancy C. Hayden
(Division Sign-Off) / Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.