← Product Code [KWP](/productcode/KWP) · K031175

# CLICK'X MONOAXIAL SCREWS AND HOOKS (K031175)

_Synthes (Usa) · KWP · Nov 14, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K031175

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Nov 14, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial, Dual Opening and the Small Stature USS (which includes small stature and pediatric patients) are noncervical spinal fixation devices intended for use as a posterior pedicle screw fixation systems (T1-S2), a posterior hook fixation systems (TI-L5), or as anterolateral fixation systems (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion. When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct. When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial and Dual Opening USS can be linked to the Cervifix system. In addition, when used with 3,5/5.0 mm parallel connectors, the Synthes Small Stature USS, can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including the Click'X and USS VAS variable axis components), the Click'X Monoaxial and Dual Opening USS Systems. In addition, Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial and the Dual Opening USS can be interchanged with all USS 6.0mm rods and transconnectors.

## Device Story

Synthes Click'X Monoaxial System is a spinal fixation implant set; includes rods, hooks, screws, and transconnectors. Used by surgeons in clinical/OR settings to stabilize the spine. Implants are surgically placed to provide structural support for spinal fusion. System components are interchangeable with existing Universal Spinal System (USS) 6.0mm rods and can link to the CerviFix system via parallel connectors. Benefits include stabilization of spinal segments to treat degenerative conditions, deformities, and trauma. No software or electronic components involved.

## Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and design comparisons.

## Technological Characteristics

Materials: Commercially pure titanium alloy conforming to ASTM F-67 or ASTM F-1295. Components: Rods, hooks, screws, transconnectors. Connectivity: Mechanical linkage via parallel connectors to USS and CerviFix systems. Sterilization: Not specified.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Reference Devices

- Cervifix system
- Universal Spinal System (USS)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 1 4 2003

Page 1 of 1

# 510(k) Summary of Safety and Effectiveness

Company:

Synthes Spine Co., L.P. 1230 Wilson Drive West Chester, PA 19380 (610) 647-9700

### Proposed Proprietary Trade Name:

Synthes Click'X Monoaxial System

Regulation Numbers: 888.3050, 888.3060 and 888.3070

Regulation Names: Spinal Interlaminal Fixation Orthosis, Spinal Intervertebral Body Fixation Orthosis, Spondylolisthesis Spinal Fixation Device System, and Pedicle Screw Spinal System

Codes: NKB, KWP, MNH and MNI

#### Description:

The purpose of this submission is to incorporate additional Click'X Monaxial components into the Universal Spinal system. The Synthes Click'X Monoaxial system consists of a variety of shapes and sizes of rods, hooks, screws, transconnectors and connecting components. The Click'X Monoaxial implant materials are fabricated from commercially pure titanium alloy, conforming to ASTM F-67 or ASTM F-1295. respectively.

### Indications:

The Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial, Dual Opening and the Small Stature USS (which includes small stature and pediatric patients) are noncervical spinal fixation devices intended for use as a posterior pedicle screw fixation systems (T1-S2), a posterior hook fixation systems (TI-L5), or as anterolateral fixation systems (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial and Dual Opening USS can be linked to the Cervifix system. In addition, when used with 3,5/5.0 mm parallel connectors, the Synthes Small Stature USS, can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including the Click'X and USS VAS variable axis components), the Click'X Monoaxial and Dual Opening USS Systems.

In addition, Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial and the Dual Opening USS can be interchanged with all USS 6.0mm rods and transconnectors .

#### Performance and SE Determination:

Comparisons of device performance data, materials, indications and design/function to predicate devices were provided in making a determination of substantial equivalence.

Synthes Spine Click'X Monoaxial 510(k)/A1

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines above them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 4 2003

Mr. John Walsh Director, Regulatory/Clinical Affairs Synthes Spine 1380 Enterprise Drive West Chester, PA 19380

Re: K031175

Trade/Device Name: Synthes Click'X Monoaxial System Regulation Number: 21 CFR 888.3050, 21 CFR 888.3060, 21 CFR 888.3070 Regulation Name: Spinal interlaminal fixation orthosis; Spinal intervertebral body fixation orthosis: Pedicle screw spinal system Regulatory Class: Class II and Class III Product Code: KWP, KWQ, MNH, MNI, NKB Dated: August 20, 2003 Received: August 22, 2003

Dear Mr. Walsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. John Walsh

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

sincerely, yours,

Mark N Mellman

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## K031175 - Click'X Monoaxial System - Additional Information

# INDICATIONS FOR USE STATEMENT

Page 1 of 1

510(k) Number (if known): K031175

Device Name: Synthes Click'X Monoaxial System

INDICATIONS:

The Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial, Dual Opening and the Small Stature USS (which includes small stature and pediatric patients) are noncervical spinal fixation devices intended for use as a posterior pedicle screw fixation systems (T1-S2), a posterior hook fixation systems (TI-L5), or as anterolateral fixation systems (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0mm parallel connectors, the Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial and Dual Opening USS can be linked to the Cervifix system. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS, can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including the Click'X and USS VAS variable axis components), the Click'X Monoaxial and Dual Opening USS Systems.

In addition, Synthes USS (including the Click'X and USS VAS variable axis components), Click'X Monoaxial and the Dual Opening USS can be interchanged with all USS 6.0mm rods and transconnectors.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

lo. Munk-n-Mulluson

110(k) Number

Synthes Spine Click'X Monoaxial 510(k)/A1

---

**Source:** [https://fda.innolitics.com/device/K031175](https://fda.innolitics.com/device/K031175)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
