SWIFT 2.0, MODEL 132.001

K031158 · Nucletron Corp. · JAQ · Aug 7, 2003 · Radiology

Device Facts

Record IDK031158
Device NameSWIFT 2.0, MODEL 132.001
ApplicantNucletron Corp.
Product CodeJAQ · Radiology
Decision DateAug 7, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5700
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Swift is a software application intended for use with Brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving afterloading radioactive sources.

Device Story

Swift 2.0 is a real-time brachytherapy treatment planning system used primarily for prostate cancer. It accepts 3D ultrasound imaging inputs to visualize anatomy and catheter placement. The system enables 3D VOI contouring, 3D catheter recognition, and inverse planning to optimize source dwell positions. It provides physicians with anatomical and dosimetric data, including dose volume histograms, dose verification points, and dose profiles. The software calculates treatment times and dose distributions, which are then used to guide the afterloader. Used in clinical settings by physicians to determine radioactive source loading and positioning prior to treatment, the system aims to improve dose distribution accuracy and patient outcomes.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Software-based treatment planning system running on Windows XP. Features include 3D ultrasound acquisition, 3D VOI contouring, 3D catheter recognition, and inverse planning algorithms. Operates as an accessory to a remote-controlled radionuclide afterloader.

Indications for Use

Indicated for brachytherapy treatment planning for cancer patients requiring intercavitary, interstitial, or intraluminal radiation therapy involving radioactive sources.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 36 Special 510(k) Nucletron Swift 2.0 April 2003 ﺮ K031158 Page 1 of 2 etron NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 Fax +31 318 550485 Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Special 510(k) section ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by section 807.92(c) ## Submitter of 510(k): Nucletron Corporation Company name: 1121753 Registration number: 7080 Columbia Gateway Drive Address: Columbia, MD 21046-2133 410-312-4100 Phone: 410-312-4197 Fax: Lisa Dimmick Correspondent: Director Assurance & Regulatory Affairs ## Modified Device Name: | Trade/Proprietary Name: | Swift 2.0 | |-------------------------|-----------------------------------| | Common/Usual Name: | Radiation Therapy Planning System | | Classification Name: | Accessory to Radiotherapy Device | | Classification: | 21Cfr892.5050 Class II | ## Legally Marketed Device(s) Our device is substantially equivalent to the legally marketed predicate device cited in the table below: | Manufacturer | Device | 510(k) # | |--------------|---------------|----------| | Nucletron BV | Real Time HDR | K022739 | ## Description: Swift is a "real time" treatment planning system for brachytherapy especially meant for treatment of cancer in the prostate. Direct 3D Ultrasound imaging of the treatment gives the physician the possibility to update the planning of the dwell positions of the source in the prostate of the patient. {1}------------------------------------------------ ## Special 510(k) Nucletron Swift 2.0 April 2003 The program provides a variety of plan evaluation tools to assist in generating the most optimal dose distribution, i.e. dose volume histograms, dose verification points and dose profiles. The software program provides the physician with anatomical and dosimetric information, to determine the positioning and loading of the radioactive sources, prior to radiation treatment. The program also provides the treatment time and dose distribution for the specified loading. From this the information the patient can be treated with the dwell positions of the source. Modifications to the previously cleared device K022739, have been made to add functionality for: - Encoder ● - 3D Ultrasound acquisition . - 3D VOI contouring ● - 3D Catheter recognition ● - Inverse Plan ● The modified device is an accessory to an afterloader. The program consists of two modules: - Treatment Planning Software: Swift ● - Database: Smoothbase . The software runs on a Windows XP platform. #### Intended use: The modified device has the same intended use as the legally marketed predicate device cited: Swift is a software application intended for use with Brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving afterloading radioactive sources. #### Summary of technological considerations: Swift is substantially equivalent to the cleared predicate device. Real Time HDR. 510(k)#: K022739. Krieken 2-4-2003 --- Date Name: Frits van Krieken Title: Business Segment Manager Nucletron B.V. Veenendaal, The Netherlands Page 37 KOB1158? Page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo. The logo is circular and contains an image of a bird in the center. The bird is stylized with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the top and left side of the circle. ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG - 7 2003 Ms. Lisa Cole Dimmick Director of Regulatory Affairs Nucletron Corporation 8671 Robert Fulton Drive COLUMBIA MD 21046 Re: K031158 Trade/Device Name: Swift 2.0 Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: 90 JAQ Dated: July 14, 2003 Received: July 16, 2003 Dear Ms. Dimmick: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Special 510(k) Nucletron Swift 2.0 April 2003 # Indications for Use Statement 510(k) Number Device Name K031158 Swift 2.0 Indications for Use Swift is a software application for Brachytherapy Treatment Planning, for the treatment of cancer, i.e. intercavitary, interstitial, intraluminal, involving radioactive sources. ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) \ OR Over-The-Counter Use_ Daniel A. Lezerm (Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...