SURGRX ENSEAL VESSEL SEALING AND HEMOSTASIS SYSTEM INCLUDING THE ENSEAL CONTROLLER ADAPTER, ENSEAL OPEN AND LAPAROSCOPIS
K031133 · Surgrx, Inc. · GEI · Jul 3, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031133
Device Name
SURGRX ENSEAL VESSEL SEALING AND HEMOSTASIS SYSTEM INCLUDING THE ENSEAL CONTROLLER ADAPTER, ENSEAL OPEN AND LAPAROSCOPIS
Applicant
Surgrx, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 3, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The SurgRx EnSeal system includes bipolar electrosurgical instruments and an accessory adapter for use with standard electrosurgical generators. It is intended for use during open or laparoscopic, general and general and general to cut and seal vesses, cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic general and gynecological surgical procedures (including unlogic, thoracic, plastic and reconstructive, bowel resections, hysterectornies, cholecystectornies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cuting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mm and bundles as large as will fit in the jaws of the Instruments. The SurgHx EnSeal Vessel Sealing & Hemostasis System has not been shown to be effective for tubal sterilization or tubal coagulation for sterlization procedures. Do not use this system for these procedures.
Device Story
EnSeal Vessel Sealing & Hemostasis System consists of bipolar electrosurgical instruments (laparoscopic tubular instruments with grasping jaws; open reusable forceps) and an EnSeal Controller Adapter (ECA). System connects to standard OR electrosurgical generators. Distal jaws contain bipolar electrodes for RF-powered vessel sealing and a mechanical blade for tissue transection. Operated by surgeons in open or laparoscopic surgical settings. Device enables vessel ligation, tissue grasping, and dissection; provides hemostasis during surgery. Benefits include effective vessel sealing up to 7 mm. Safety precautions required for patients with pacemakers or ICDs; not for use in pregnant patients or for tubal sterilization.
Clinical Evidence
No clinical data. Evidence consists of preclinical laboratory bench testing and performance evaluations conducted to ensure the device functioned as intended and met design specifications. Testing included human factors evaluation by physicians.
Technological Characteristics
Bipolar electrosurgical instrument; RF energy source. Materials: surgical stainless steel, carbon-filled polyethylene, polyetherimide (Ultem), silicone, ABS. Laparoscopic devices: radiation sterilized. Packaging conforms to ISTA 1A/2A and ASTM 4169-99 standards. Connectivity: connects to standard OR electrosurgical generators via EnSeal Controller Adapter (ECA).
Indications for Use
Indicated for open and laparoscopic general and gynecological surgical procedures (including urologic, thoracic, plastic/reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies) requiring vessel ligation, tissue grasping, and dissection. Suitable for vessels up to 7 mm. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
ValleyLab LigaSure Vessel Sealing System (K981916)
Submission Summary (Full Text)
{0}------------------------------------------------
# lot 2
### 510(k) SUMMARY
#### 1. Submitter Name and Address
SurgRx, Inc. 380 Portage Avenue Palo Alto, CA 94306 Contact: Linda Oleson Phone: (650) 739-0920 xt 107 Date: 6/10/03
#### 2. Device Name
Trade name: EnSeal Vessel Sealing & Hemostasis System
Common name: Electrosurgical open and laparoscopic instruments and accessories
Classification name: Electrosurgical Cutting and Coaqulation Device and Accessories (per 21 CFR section 878.4400) and Gynecologic Electrocautery and Accessories (per 21 CFR 884.4120).
#### 3. Predicate Device
ValleyLab LigaSure Vessel Sealing System (K981916).
#### 4. Device Description
The EnSeal Laparoscopic devices are tubular instruments with grasping jaws at the distal end, which are actuated by a handle at the proximal end of the device. The distal end contains bipolar electrodes for sealing vessels, and a mechanical cutting blade for transecting vessels. The EnSeal Open devices are reusable forceps-type devices with one snap-in single use, disposable bipolar electrode, and one permanent, reusable electrode in the jaws of the device. Connectors attached to an electrical cable connect the device to the ECA (EnSeal Controller Adapter) accessory which connects the device to a standard O.R. electrosurgical generator.
#### 5. Intended Use
The SurgRx EnSeal devices are intended for use during open or laparoscopic general and gynecologic surgery to cut and seal vessels, cut, grasp and dissect tissue and/or seal vessels during surgery. The SurgRx EnSeal Vessel Sealing & Hemostasis System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.
#### 6. Technological Characteristics
The EnSeal devices are similar to the predicate devices in that they are all bipolar instruments used to cut and seal vessels, grasp and dissect tissue, utilizing RF powered distal ends. They vary in technological characteristics by jaw design.
#### 7. Performance Data
Preclinical laboratory (bench studies) and performance tests were executed to ensure the devices functioned as intended and met design specifications.
#### 8. Conclusions
The EnSeal devices outperformed the predicate devices in comparative preclinical testing. Based on performance and functional similarities to the predicate devices, we believe the EnSeal devices are safe and effective and substantially equivalent to the predicate devices.
380 PORTAGE AVE • PALO ALTO • CALIFORNIA • 94306 PHONE: (650) 739-0920 • FAX: (650) 739-0929 E-10
{1}------------------------------------------------
K031133 2 of 2
- · Pacemakers and implanted cardioverter/defibrillators can be adversely affected by RF signals. Consult the pacemaker/cardioverter/defibrillator manufacturer for further information when use of electrosurgical equipment is planned in patients with these devices. If the patient has an internal cardiac defibrillator (ICD), contact the ICD manufacturer instructions before performing electrosurgery. This treatment may cause multiple activations of an ICD.
- · SurgRx recommends against the use of laparoscopic surgery on pregnant patients.
- · SurgRx recommends against the use of electrosurgery for circumcisions.
. . .
- The SurgRx EnSeal Vessel Sealing & Hemostasis System has not been shown to be effective . for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.
#### 3. Engineering Specifications
See Attachment 4 for engineering drawings/physical descriptions of the devices and complete materials lists.
#### a) Components
| Open Instruments | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Reusable Handle : | Surgical stainless steel |
| Jaws (disposable and<br>reusable portions): | Surgical stainless steel, carbon filled polyethylene,<br>polyetherimide (Ultem), silicone, silicone coating. |
| Laparoscopic Instruments | |
| Handle: | Acrylonitrile-butadiene-styrene Terpolymer (ABS) |
| Jaws: | Surgical stainless steel, carbon filled polyethylene,<br>silicone coating. |
| I-Beam: | Surgical stainless steel |
#### c) Human Factors
Shaft:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Instruments are designed similar to currently marketed devices. The device has been evaluated from the user perspective in pre-clinical and bench testing by physicians, and taken into consideration as design input. Devices should be easily adaptable to users.
polyolefin shrink tubing - black.
Surgical stainless steel, heat shrink insulation of
#### d) Packaging
All SurgRx EnSeal products packaging shall conform to the International Safe Transit Association (Project 1A and Project 2A) requirements for packaged products weighing less than 100 pounds. Customer over shippers shall meet the following ASTM 4169-99 standards for: Impact Test Criteria, Static Compression Test Criteria, Vibration Test Criteria.
Laparoscopic devices will be packaged on a chip board tray in a Tyvek pouch for radiation sterilization, and placed in a chip board box. Six unit boxes will be packaged in a corrugated over shipper.
The non-sterile Open device handles will be packaged in a bubble-wrap pocket and placed in a corrugated cardboard box for shipping.
Disposable electrodes will be packaged in a Tyvek pouch for radiation sterilization and 12 unit pouches will be placed in a chipboard box.
The ECA box will be placed in an anti-static bag, surrounded by protective foam cushioning and placed in a corrugated box. ECA Accessories (batteries in a ziplock bag, footswitch in a ziplock bag, Reusable Extension Cable in a ziplock bag) will be placed in a packaging insert corrugated box and placed in the same box as the ECA (see Attachment 4 for exploded view).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an emblem that features a stylized depiction of an eagle or other bird, with three curved lines representing its wings or feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
3 2003 JUL
Surgrx, Inc. c/o Mr. Morten Simon Christensen Underwriters Laboratories, Inc. 1655 Scott Boulevard Santa Clara, California 95050
Re: K031133
Trade/Device Name: SurgRx EnSeal Vessel Sealing & Hemostasis System Regulation Number: 21 CFR 884.4120, 21 CFR 878.4400 Regulation Name: Gynecologic electrocautery and accessories Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI, HGI Dated: June 17, 2003 Received: June 19, 2003
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Morten Simon Christensen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mark A. Wilkerson
Celia M. Witten, Ph.D., M.D. Director
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Applicant: SurgAx, Inc.
Kostla 3 510(k) number (if known):_
Device Name: SurgRx EnSeal Vessel Sealing & Hemostasis System
Indications for Use:
The SurgRx EnSeal system includes bipolar electrosurgical instruments and an accessory adapter for use with standard electrosurgical generators. It is intended for use during open or laparoscopic, general and general and general to cut and seal vesses, cut, grasp and dissect tissue during surgery.
Indications for use include open and laparoscopic general and gynecological surgical procedures (including unlogic, thoracic, plastic and reconstructive, bowel resections, hysterectornies, cholecystectornies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cuting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mm and bundles as large as will fit in the jaws of the Instruments.
The SurgHx EnSeal Vessel Sealing & Hemostasis System has not been shown to be effective for tubal sterilization or tubal coagulation for sterlization procedures. Do not use this system for these procedures.
## (Please DO NOT WRITE BELOW THIS LINE - Continue on another Page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use*_*
(Optional Format 1-2-96)
Page 1 of 1
for Mark N. Millen
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
510(k) Number K031133
D-8
06/27/2003 09:56AM
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.