MAYO CLINIC BC-10 1.5T

K031119 · Mayo Medical Ventures · MOS · Jul 15, 2003 · Radiology

Device Facts

Record IDK031119
Device NameMAYO CLINIC BC-10 1.5T
ApplicantMayo Medical Ventures
Product CodeMOS · Radiology
Decision DateJul 15, 2003
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The Mayo Clinic BC-10 1.5T wrist coil will be used in conjunction with several commercially available MRI systems to more accurately image a variety of maladies associated with the hand, wrist, forearm, and elbow. Maladies include, but are not limited to, carpal tunnel syndrome, nerve compression, ligament injuries, tendon abnormalities, and scarring. Its 10cm diameter design governs the specific area of the body which can be imaged.

Device Story

The Mayo Clinic BC-10 is a transmit/receive (T/R) magnetic resonance imaging (MRI) coil. It utilizes a high-pass, quadrature birdcage circuit design to capture radiofrequency signals from small extremities (hand, wrist, forearm, elbow). The device is used in clinical settings in conjunction with 1.5T MRI systems. It provides high signal-to-noise ratio (SNR) imaging within a 10cm field of view. The coil is positioned overhead or at the patient's side during the scan. Radiologists use the resulting images to identify anatomical abnormalities and pathologies, aiding in clinical diagnosis and treatment planning.

Clinical Evidence

Bench testing only. Testing demonstrated that the device performed in accordance with predetermined acceptance criteria and met or exceeded safety and reliability standards.

Technological Characteristics

High-pass, quadrature, transmit/receive (T/R) birdcage coil. 10cm diameter. Optimized for small extremity anatomy. No external tuning required. Compatible with Signa 1.5T MR systems.

Indications for Use

Indicated for patients requiring MRI imaging of hand, wrist, forearm, and elbow to diagnose conditions including carpal tunnel syndrome, nerve compression, ligament injuries, tendon abnormalities, and scarring.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko3 11/19 1 of 2 # JUL 15 2003 ## 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c). | Submitter: | Mayo Medical Ventures<br>200 First St. SW<br>Rochester, MN 55905 | |-----------------|------------------------------------------------------------------| | Contact Person: | Jade Sadosty<br>Program Manager | | Telephone: | (507) 538-1352 | | Fax: | (507) 284-5410 | | Date Prepared: | June 18, 2003 | # Device Name: Mayo Clinic BC-10 MR coil # Device Description: The Mayo Clinic BC-10 magnetic resonance coil is a transmit/receive, high-pass, quadrature coil using the birdcage circuit design. # Indications for Use: The Mayo Clinic BC-10 1.5T wrist coil will be used in conjunction with several commercially available MRI systems to more accurately image a variety of maladies associated with the hand, wrist, forearm, and elbow. Maladies include, but are not limited to, carpal tunnel syndrome, nerve compression, ligament injuries, tendon abnormalities, and scarring. Its 10cm diameter design governs the specific area of the body which can be imaged. #### Predicate Device: The predicate devices for this coil include both the MRI Devices 1.5T phased array wrist coil and the GE Signa 1.5T birdcage, high-pass, quadrature T/R head coil. | Feature | Mayo Clinic BC-10 | MRI Devices Signa 1.5T | |--------------------|------------------------------------------------------------------------------------|--------------------------------| | Coil Type | High-pass, quadrature, T/R | Phased array, receive only | | Region of Interest | Small extremities < 10cm | Hand and wrist | | Compatibility | All Signa 1.5T MR systems | Signa 1.5T with PA option only | | Tuning | No external tuning necessary.<br>Coil is optimized for small<br>extremity anatomy. | No external tuning necessary. | {1}------------------------------------------------ | Patient Positioning | Overhead or at patient's side | Overhead or at patient's side | |-----------------------|--------------------------------------------------------|------------------------------------------------------| | Imaging Configuration | High SNR for small extremity<br>imaging with 10cm FOV. | High SNR for hand and wrist<br>imaging with 6cm FOV. | ## Summary of Studies: Testing was performed to demonstrate the Mayo Clinic BC-10 performed in accordance with predetermined acceptance criteria and that the safety and reliability of the coil meets/exceeds standards. Details of this testing are included in the abbreviated 510(k) submission. # Conclusions: The results of the testing and evaluations referenced above show that the Mayo Clinic BC-10 is substantially equivalent to already approved extremity coils. It is a proven design with well documented clinical, safety, and reliability data from years of use at the Mayo Clinic in Rochester, MN. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem of an eagle with its wings spread, stylized with three curved lines representing the feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 15 2003 Mr. Jade Sadosty Program Manager Mayo Medical Ventures 200 First Street SW ROCHESTER MN 55905 Re: K031119 Trade/Device Name: Mayo Clinic BC-10 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: April 7, 2003 Received: April 29, 2003 Dear Mr. Sadosty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ Page 1 of 1 510(k) Number (if known): Mayo Clinic BC-10 1.5T Device Name: Indications For Use: The Mayo Clinic BC-10 1.5T wrist coil will be used in conjunction with several commercially available MRI systems to more accurately image a variety of maladies associated with the hand, wrist, forearm, and elbow. Maladies include, but are not limited to, carpal tunnel syndrome, nerve compression, ligament injuries, tendon abnormalities, and scarring. Its 10cm diameter design governs the specific area of the body which can be imaged. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) *Prescription Use* David R. Segerson (Division Sign-Of and Radiological 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...