The PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula is indicated for use in providing percutaneous temporary tracheal access for airway management.
Device Story
Kit facilitates percutaneous insertion of tracheostomy tube into stoma created by serial dilation. Components include specialty tracheostomy tube, disposable inner cannula, and angled obturator/introducer. Tube features radiopaque blue line, biocompatible polyvinylchloride construction, tapered/beveled distal tip, neck flange, 15 mm connector, and cuff with self-sealing Luer valve. Used in clinical settings by physicians for airway management. Modified design includes increased bevel, anterior positioning, altered tube/cannula arc of curvature, reduced straight portion length, and dilator stop markings. Output is established tracheal access; benefits include reduced incidence of tube occlusion and safe, trauma-free insertion.
Clinical Evidence
Clinical user evaluation conducted at St. John's Mercy Medical Center with 54 total subjects (28 control, 26 study). Primary endpoints included incidence of tube occlusion and ease of insertion. Results showed a statistically highly significant reduction in the incidence of tube occlusion with the modified device. No trauma or difficulty was associated with insertion using the modified bevel and angled obturator.
Technological Characteristics
Biocompatible polyvinylchloride construction; radiopaque blue line; tapered/beveled distal tip; 15 mm connector; cuff with self-sealing Luer valve. Modifications include increased bevel, anterior positioning, altered arc of curvature, reduced straight portion length, angled obturator/introducer, and dilator stop markings. Manual insertion via percutaneous serial dilation.
Indications for Use
Indicated for patients requiring percutaneous temporary tracheal access for airway management.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
Predicate Devices
PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula (K980466/K960429)
{0}------------------------------------------------
K031057
JUL 1 8 2003
Image /page/0/Picture/2 description: The image shows a logo with the word "PORTEX" in white letters. The word is set against a black background. The background is a square with a curved bottom edge.
Portex, Inc.
* Bourna: Drive Kerse NH-13431-1772-05-A let o63 357 3812 www.puttershall.com
## K: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
#### 510(K) SUMMARY:
#### COMPANY INFORMATION
Portex, Inc. 10 Bowman Drive Keene, NH 03431 (603) 352-3812 Contact: Timothy J. Talcott Director, Regulatory Affairs
#### PREPARATION DATE OF SUMMARY
April 2, 2003
#### TRADE NAME
PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula
#### COMMON NAME
Tracheostomy Tube with Disposable Inner Cannula
#### PRODUCT CLASS/CLASSIFICATION
Class II, 73 JOH, 21 CFR 868.5800.
### PREDICATE DEVICES
Predicate 1: Our current PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula, K980466[K960429]. This is the main predicate device, cited for materials/biocompatibility and for most design characteristics.
{1}------------------------------------------------
Predicate 2: Our DIC (Disposable Inner Cannula) Tracheostomy Tube, K934465[K903730]. This predicate device is cited, in conjunction with Predicate 1 above, for the tube/inner cannula shape and distal tip geometry.
Predicate 3: The Rusch 'PercuQuick® Dilatator Set', K011210. This predicate device is cited for the obturator angled tip.
### DESCRIPTION
The PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula is designed to permit percutaneous insertion of a specifically designed tracheostomy tube. This single cannula Percutaneous Tracheostomy Tube has a radiopaque blue line and is constructed of biocompatible polyvinylchloride material and incorporates a tapered and beveled distal tube tip to facilitate insertion through the percutaneously dilated stoma site. The tracheostomy tube has a neck flange, an integral 15 mm connector, and a cuff which deflates to a low profile on the tube for a smooth transition during insertion. The cuff inflation line has a self sealing Luer valve. The tracheostomy tube is supplied with a disposable inner cannula of the appropriate size. The Percutaneous Tracheostomy Kit comes with the necessary components for percutaneous tracheostomy tube insertion into a stoma created by serial dilation.
### INDICATIONS FOR USE
The PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula is indicated for use in providing percutaneous temporary tracheal access for airway management.
### TECHNICAL CHARACTERISTICS
The PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula has the same technical characteristics as is currently marketed with our existing PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula, K980466[K960429]. The device is being modified to increase the bevel and change to the anterior position, slightly modify the shape of the tube and inner cannula [increased arc of curvature, reduced length of straight portion resulting in an overall decrease in nominal length], provide an angled obturator/introducer, and add marking on the dilator stop on the guiding catheter. The Instructions for Use Physician's Clinical Guide are also revised to reflect these changes.
{2}------------------------------------------------
### SUMMARY OF NONCLINICAL AND CLINICAL TESTS
A side-by-side comparison of the proposed device to three legally marketed devices has been conducted, covering intended use, materials and design characteristics. A clinical user evaluation has been conducted at St. John's Mercy Medical Center with 28 subjects in the control group and 26 subjects in the two study groups.
### CONCLUSION OF NONCLINICAL AND CLINICAL TESTS
The side-by-side comparison concluded that the proposed device is substantially equivalent to the predicate devices cited.
The clinical user evaluation concluded that the reduction in the incidence of occlusion of the tracheostomy tube with the proposed device was statistically highly significant. It was concluded that the proposed device is at least equivalent to the current device in these respects.
The evaluation also found that there was no trauma or difficulty associated with insertion of the tube featuring the modified bevel and using the angled obturator. This provided useful support of the conclusion that the proposed device is equivalent to the predicate device(s) in these respects.
### CONCLUSION
The proposed device is safe and effective and is substantially equivalent to the predicate devices.
Portex, Inc.
Timothy Dalton
Timothy J. Paleott Director, Regulatory Affairs
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular logo. The logo features a stylized bird with three curved lines representing its body and wings. The bird is facing right. Encircling the bird is text that reads "HEALTH & HUMAN SERVICES USA". The text is arranged along the perimeter of the circle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 8 2003
Mr. Timothy J. Talcott Portex, Incorporated 10 Bowman Drive Keene, New Hampshire 03431
Re: K031057
Trade/Device Name: PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula Regulation Number: 868.5800 Regulation Name: Tracheostomy Tube and Tube Cuff Regulatory Class: II Product Code: JOH Dated: April 2, 2003 Received: April 3, 2003
Dear Mr. Talcott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Timothy J. Talcott
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Kein Muly
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# B: INTENDED USE OF DEVICE
Page 1_of_1
510(k) Number (if known): KO3105 F
Device Name: PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula
Indications For Use:
The PER-FIT™ Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy Tube and Disposable Inner Cannula is indicated for use in providing percutaneous temporary tracheal access for airway management.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use
OR
Over-The-Counter Use
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K031057
Concurrence of CDRH, Office of Device Evaluation (ODE)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.